8 January 2015 Noninvasive high resolving power entangled photon quantum microscope
Author Affiliations +
Abstract
A noninvasive high resolving power quantum microscope would facilitate progress in the areas of research and development in biosciences as well as in the area of biomedical technology. Longer-wavelength microscopes, i.e., visible or near-infrared, can provide noninvasive features. On the other hand, shorter wavelengths, i.e., in the ultraviolet, can provide better resolving power. We propose the development of both a noninvasive and high resolving power quantum microscope by using two-color entangled photon ghost imaging technology.

1.

Introduction

Many key areas of research and development in the biosciences and biomedical technology areas are progressing rapidly. One of the more exciting areas is biomedical imaging. Unfortunately, the light from high-resolving microscopes tends to damage cells. A high resolving power noninvasive microscope, capable of imaging without damaging cells or tissues, would enable physicians and scientists to better see and understand tissue and organ function.

The resolving power of a microscope can be improved in several ways. One example was the development of confocal microscopes, which provided better resolving power as well as higher contrast with respect to standard wide-field microscopy. Two other approaches have been proposed and analyzed by Simon and Sergienko1,2 to improve the resolution of standard confocal microscopy. In one of these approaches, Simon and Sergienko used photon correlation, which improved resolution up to 50%.1 In the other approach, they used a degenerate entangled photon pair generated from spontaneous parametric down-conversion (SPDC) to improve the resolution up to 77%.2 Simon and Sergienko suggested the use of a nonlinear crystal after the object to illuminate the specimen with a high pump power to generate sufficient signal and idler photons. To obtain visible-light images with UV-level resolution, they also suggested the use of a UV pump to generate signal and idler photons in the visible range. However, because the specimen was illuminated by high-power UV light, there was a high probability of damaging the specimen, i.e., the cells and tissues studied in biomedical and biosciences.

In 1995 at UMBC, while investigating Einstein-Podolsky-Rosen phenomena, Pittman et al.3 discovered ghost imaging using entangled photons to transmissively image a stencil mask internal to their experimental setup. Meyers et al.,4,5 in 2007 and 2008, generated the first ghost image of a remote object by imaging a small toy soldier placed remotely from their setup at the Army Research Laboratory. This demonstration was the first practical application of ghost imaging, and it showed that ghost imaging could be applied to remote sensing from astrophysical to microscopic scales with the potential of increased resolution, increased contrast, and mitigation of adverse effects from distorting media. In 2010, Karmakar and Shih6,7 reported on an experimental demonstration of two-color ghost imaging with enhanced resolving power. The resolution of two-color ghost imaging was also studied theoretically by Rubin and Shih8 and Chan et al.9 Certainly, disturbances around the object can degrade the image contrast of a microscope. However, in 2011, Meyers et al.10,11 showed that disturbances, such as turbulence and distortions, do not affect multiphoton interference ghost imaging. While these studies used the quantum properties of thermal light, the Meyers et al. turbulence-free ghost imaging design also applies to entangled photons. In addition, absorption of photons by an object, though this leads to image information in conventional imaging, can potentially degrade the correlations in certain correlation-based imaging. An advantage of entangled photon imaging is that it eliminates most noise photons that can degrade correlation techniques and relatively few entangled photons are needed to accurately determine features.3 However, accidental coincidences will form a noise background. With good detection equipment, accidental coincidences will be minimized. Furthermore, positive-negative ghost imaging techniques have been developed and experimentally verified, which achieve near 100% contrast by eliminating the effect of the noise background,11,12 which improves images even using thermal light two-photon interference ghost imaging.

In this report, we propose a quantum microscope configuration capable of achieving high resolving power without any destruction of cells or tissues by using two-color ghost imaging technology. Such a quantum microscope configuration could provide improved image contrast and would not be affected by disturbances around the object.

2.

Proposed Configuration

In the proposed experimental configuration, two-color ghost imaging technology is applied to improve the resolving power of a microscope system. Figure 1 shows a schematic of the experimental setup. By recording the joint-detection counts between the bucket detector and the CCD as a function of the transverse plane coordinates of the CCD in the idler beam, the image of the object aperture is observed with a magnification factor that is the same as the degenerate one-color case3 as well as the classical case. In this experimental configuration, entangled photon pairs with λs>λi generated from SPDC can reproduce a ghost image with enhanced resolving power by a factor of λs/λi by means of higher imaging amplification, where λs is the wavelength of the radiation that illuminates the object and λi is the wavelength of the idler beam. In this case, image amplification is enhanced by a factor of λs/λi more than that of the one-color ghost imaging or that of a classical imaging setup. Since the lowest limit of transparency of the nonlinear crystals for SPDC is 190nm, we propose the use of a 197-nm excimer laser as a pump. A type I BBO crystal together with this pump can be used as an entangled two-photon source to generate a signal beam of 695 nm in the visible range and an idler beam of 275 nm in the UV range. This configuration would improve the magnification power of a quantum microscope by a factor of 2.5. If one could produce a two-photon source capable of generating λs500nm in the visible region and λi1nm in the x-ray region, then the magnification power could be improved by a factor of 500.

Fig. 1

Schematic of the experimental setup. DM is a dichroic mirror, which is used to separate two different colors of signal and idler beams.

JBO_20_1_016008_f001.png

3.

Discussion and Analysis

In the two-color case, the signal and the idler photon pair from the SPDC source are emitted at different angles and opposite directions. Furthermore, if one photon is emitted at a certain angle, the emission angle of its conjugate is determined by the certain angle of that photon with unit probability. This permits us to explain the experiment in terms of geometrical optics by unfolding the schematic in Fig. 1 into that shown in Fig. 2 by considering the SPDC source as a hinge. The geometrical rays in Fig. 2 represent the two-photon amplitudes, therefore, the image is formed in terms of coincidence counts when the object aperture, imaging lens, and CCD are located according to the modified Gaussian thin lens equation. The relationship among the object aperture’s optical distance from the lens, so, the image plane’s optical distance from the lens, si (which is actually the distance from the lens back to SPDC source and then straight forward to the image plane), and the focal length of the imaging lens, f, satisfies a modified Gaussian thin-lens equation for two colors:6,8

Eq. (1)

ωsso+ωisi=ωsfor1so+1sid1+d2ωsωisi(ghost)=1f,
where

Eq. (2)

si=d1ωiωs+d2andsi(ghost)=d1+d2,
and ωs and ωi are the angular frequencies of the signal and idler beam, respectively, and the Gaussian thin-lens equation in classical as well as in the degenerate case is

Eq. (3)

1soO+1siO=1f,
where soO and siO are the object distance and image distance of the standard degenerate ghost imaging and classical imaging experimental setup.

Fig. 2

Unfolded version of the experimental setup is shown to better understand the physics.6,13

JBO_20_1_016008_f002.png

By using the Glauber theory of photodetection, the joint detection probability between two detectors D1 and CCD located at (r1,t1) and (r2,t2), respectively, can be expressed as follows:14,15

Eq. (4)

G(2)(r1,t1;r2,t2)=|0|E^1()E^2()E^2(+)E^1(+)|Ψ|2=|Ψ(ρ1,z1,t1;ρ2,z2,t2)|2,
where E^1(), E^2(+) are the negative and positive frequency fields at space-time points (r1,t1) and (r2,t2), respectively, and Ψ(ρ1,z1,t1;ρ2,z2,t2) is the two-photon amplitude, also called the two-photon effective wavefunction. Since imaging takes place on the transverse plane, we can calculate the transverse part of the two-photon effective wavefunction at the point (z1=zo,ρ1=ρo;z2,ρ2):

Eq. (5)

Ψ(ρo;ρ2)=Ψodκsdκiδ(κs+κi)dωsdωiδ(ωs+ωiωp)×g(κs,ωs;ρo,zo)g(κi,ωi;ρ2,z2),
where g(κs,ωs;ρo,zo) and g(κi,ωi;ρ2,z2) are the Green’s functions evaluated by propagating the signal from the output plane of the two-photon source to the object plane and the idler from the output plane of the two-photon source to a CCD, where ρo defines the transverse coordinates of the object plane.16,17 Consider that an arbitrary phase of Δϕ(ρo) is introduced on the transverse coordinates of the object plane ρo by disturbances around the object. Then the two-photon wavefunction becomes:

Eq. (6)

Ψ(ρo;ρ2)=Ψodκsdκiδ(κs+κi)dωsdωiδ(ωs+ωiωp)×g(κs,ωs;ρo,zo)eiΔϕ(ρo)g(κi,ωi;ρ2,z2).

Since G(2)(ρo;ρ2)=|Ψ(ρo;ρ2)|2, the disturbances have no effect on the formed ghost image in terms of coincidences. After simplification and using the modified Gaussian thin-lens equation shown in Eq. (1) in Eq. (5), we obtain the generalized two-photon effective wavefunction for this case as:

Eq. (7)

Ψ(ρo;ρ2=ρi)somb[Rc(ωssoρo+ρid1ωs+d2ωi)],
where the point-spread function somb(x) is a result of the finite size of the lens (radius R) and ρi defines the transverse coordinates of the image plane.

A collection lens along with detector D1 can be treated as a bucket detector, which is shown in Figs. 1 and 2. The single photon detectors used in quantum imaging are usually very fast, have low dead time, high time resolution, and support high count rates (typically 32 ns dead time, 300 ps timing resolution, and also support up to 30millioncounts/s). These features of single-photon detector allow fast accumulation of necessary counts. These features of single-photon detectors allow fast accumulation of necessary counts. Fast single photon detectors allow an integration time of approximately a few minutes for the reported quantum microscopy. Generally, this integration time is less than the life time of the specimen. Faster detection methods currently in development, such as those described by Pittman,18 will further reduce the integration times needed for quantum microscopy. The bucket detector integrates all Ψ(ρo;ρ2=ρi) reflected and scattered from the object aperture A(ρo) as a joint-detection event. This process can be written as in the following convolution:

Eq. (8)

R12objectdρo|A(ρo)|2|Ψ(ρo;ρi)|2,
where the CCD is scanned in the image plane ρ2=ρi. The image is observed in terms of coincidence counts (R12) as a function of the transverse plane coordinates of the CCD (ρ2=ρi).

The finite size of the image spot defined by a point-spread function in Eq. (7) determines the angular resolution of the imaging. According to Rayleigh’s criterion, the images of two nearby point objects are said to be unresolvable when the center of one point-spread function falls on the first minimum of the point-spread function of the other. To clarify this situation, consider a point located at ρo=0 on the object plane. The corresponding unique spot in the image plane is determined by the point-spread function defined in Eq. (7). Rayleigh’s criterion defines the size of the point-spread function by considering its first minimum, i.e.,

Eq. (9)

Rc(d1ωs+d2ωi)Δ|ρi|3.83orΔ|ρi|3.83cR(d1ωs+d2ωi).

Now, we consider another nearby point on the object plane at ρo=a. In order to separate the image of ρo=a from that of ρo=0, the value of m|a|min cannot be smaller than Δ|ρi|, and hence

Eq. (10)

m|a|min3.83cR(d1ωs+d2ωi),
where m=|ρi|/|ρo|=[(d1/ωs)+(d2/ωi)]/(so/ωs), is the magnification factor of the image. Thus, we have the angular limit of spatial resolution for two color ghost imaging as

Eq. (11)

Δθmin=|a|minso=3.83cR1ωs.

Last, under the approximation d1d2, the two-photon effective wavefunction becomes

Eq. (12)

Ψ(ρo;ρ2=ρi)somb[Rc(ωssoρo+ωisiρi)],
with the modified image forming equation as

Eq. (13)

1so+1siωsωi=1f,
where sid2.

Compared to the degenerate ghost imaging system17 and classical imaging system, we found three interesting features for this quantum microscope system.

  • 1. An angular limit of resolution that is determined by the wavelength of the signal λs:

    Eq. (14)

    Δθmin=|a|minso=3.83cRωs1.22λsD,
    where D is the diameter of the imaging lens.

  • 2. A different image forming equation: As shown in Eq. (13), the image plane must be moved away from the imaging lens by a factor of ωi/ωs in order to form an image when the focal length of the imaging lens f and the object distance so=soO are kept the same as in the degenerate case:

    Eq. (15)

    si=ωiωssiO.

  • 3. The same imaging magnification factor as that of the degenerate case, with imaging amplification affected by a factor of ωs/ωi:

    Eq. (16)

    |ρi||ρo|=ωsωisiso=siOsoO.

Comparing the two-color case with the degenerate case and with the classical imaging system with a longer wavelength λ=λs, we achieve the same angular resolution limit Δθmin=1.22(λs/D). When the image plane is moved away from the imaging lens by a factor of ωi/ωs in order to form an image while keeping the above conditions steady, the imaging amplification increases by a factor of ωi/ωs. In this case, the improved amplification factor is 2.5. As stated above, greater imaging amplification under the enhanced resolving power of an imaging system is a useful and important feature in a microscope.

In case of the one-color ghost imaging as well as the classical case, we can resolve 1μm with visible light when the object distance (so) from the imaging lens is 4 cm. The size of human cells and tissues are of the order of 1μm. Again, this proposed entangled photon quantum microscope technique would improve the resolving power by a factor of 2.5 over that of a classical wide-field microscope. In this case, the image contrast would be unaffected by disturbances around the object. Furthermore, in this proposed configuration, cells and tissues would be illuminated by visible light instead of UV light, and this would minimize the potential for cell or tissue damage.

4.

Comparison of Features of Microscopy

In this section, we present a table to compare the resolving power of different kinds of microscopes in terms of the resolving power of a standard wide-field microscope (R). We also compare the image contrast and specimen safety. The invention of confocal microscopy has provided two important advantages over wide-field microscopes: (1) better resolving power and (2) very good contrast. Later on, the advent of a twin-photon confocal microscope reported improved resolving power significantly beyond the standard confocal microscope. Although twin-photon confocal microscopes can provide very good resolving power, there remains a high probability for cell or tissue damage. In this article, we have reported on a quantum noninvasive, high-resolution microscope that comprises all of the desired features, as shown in Table 1.

Table 1

Comparison of features of microscopes.

MicroscopeResolving powerContrastSafety of specimen
Wide-fieldRNoisyVery high
Confocal.73RVery goodVery high
Twin-photon confocal.17RVery goodVery low
Noninvasive high resolving power.4RVery goodVery high

5.

Conclusion

The main goal of a modern microscope is to achieve high resolving power, high contrast, and noninvasive imaging. Yet, it has been difficult to design a microscope to achieve all of these features together as shown in the comparison of current microscopes above. Nevertheless, the proposed quantum microscope reported here using two-color ghost imaging technology will be able to achieve all the desired features. Here, cells and tissues will be illuminated by one color in the visible range. The resolving power will be determined by the ratio of the signal and idler photon wavelengths, where, for this case, the ratio is 2.5. As a result, the proposed quantum microscope would be noninvasive with high resolving power. In addition, configurations involving virtual images are conceivable, which may increase magnification. Note that in alternate configurations, the resolving power could be improved further by increasing the wavelength ratio of the signal and idler photons since images are formed in terms of coincidence counts between a bucket detector and the CCD. Significantly, image disturbances, such as from motion of intervening fluid in biological samples or from media turbidity, will have virtually no effect on the contrast of the images in this system. We find that such a quantum microscope would be widely applicable in advancing the progress of biomedical research and medical technologies.

Acknowledgments

The authors thank K. Deacon, A. Tunick, P. Hemmer, and J. Sprigg for helpful discussions. S. Karmakar is a postdoctoral fellow at ARL (W911NF-11-2-0074).

References

1. 

D. S. SimonA. V. Sergienko, “The correlation confocal microscope,” Opt. Express, 18 9765 (2010). http://dx.doi.org/10.1364/OE.18.009765 OPEXFF 1094-4087 Google Scholar

2. 

D. S. SimonA. V. Sergienko, “Twin-photon confocal microscopy,” Opt. Express, 18 22147 (2010). http://dx.doi.org/10.1364/OE.18.022147 OPEXFF 1094-4087 Google Scholar

3. 

T. B. Pittmanet al., “Optical imaging by means of two-photon quantum entanglement,” Phys. Rev. A, 52 R3429 (1995). http://dx.doi.org/10.1103/PhysRevA.52.R3429 PLRAAN 1050-2947 Google Scholar

4. 

R. E. MeyersK. S. DeaconY. Shih, “A new two-photon ghost imaging experiment with distortion study,” J. Mod. Opt., 54 (16), 2381 –2392 (2007). http://dx.doi.org/10.1080/09500340701400117 JMOPEW 0950-0340 Google Scholar

5. 

R. E. MeyersK. S. DeaconY. Shih, “Ghost-imaging experiment by measuring reflected photons,” Phys. Rev. A, 77 041801 (2008). http://dx.doi.org/10.1103/PhysRevA.77.041801 PLRAAN 1050-2947 Google Scholar

6. 

S. KarmakarY. Shih, “Two-color ghost imaging with enhanced angular resolving power,” Phys. Rev. A, 81 033845 (2010). http://dx.doi.org/10.1103/PhysRevA.81.033845 PLRAAN 1050-2947 Google Scholar

7. 

S. KarmakarY. Shih, “Observation of two-color ghost imaging,” Proc. SPIE, 7702 770204 (2010). http://dx.doi.org/10.1117/12.851188 PSISDG 0277-786X Google Scholar

8. 

M. H. RubinY. Shih, “Resolution of ghost imaging for nondegenerate spontaneous parametric down-conversion,” Phys. Rev. A, 78 033836 (2008). http://dx.doi.org/10.1103/PhysRevA.78.033836 PLRAAN 1050-2947 Google Scholar

9. 

K. W. C. ChanM. N. O’SullivanR. W. Boyd, “Two-color ghost imaging,” Phys. Rev. A, 79 033808 (2009). http://dx.doi.org/10.1103/PhysRevA.79.033808 PLRAAN 1050-2947 Google Scholar

10. 

R. E. MeyersK. S. DeaconY. H. Shih, “Turbulence-free ghost imaging,” Appl. Phys. Lett., 98 111115 (2011). http://dx.doi.org/10.1063/1.3567931 APPLAB 0003-6951 Google Scholar

11. 

R. E. MeyersK. S. DeaconY. H. Shih, “Positive-negative turbulence-free ghost imaging,” Appl. Phys. Lett., 100 131114 (2012). http://dx.doi.org/10.1063/1.3698158 APPLAB 0003-6951 Google Scholar

12. 

H. ChenT. PengY. Shih, “100% correlation of chaotic thermal light,” Phys. Rev. A, 88 023808 (2013). http://dx.doi.org/10.1103/PhysRevA.88.023808 PLRAAN 1050-2947 Google Scholar

13. 

D. N. Klyshko, “Combine EPR and two-slit experiments: interference of advanced waves,” Phys. Lett. A, 132 299 (1988). http://dx.doi.org/10.1016/0375-9601(88)90856-0 PYLAAG 0375-9601 Google Scholar

14. 

R. J. Glauber, “The quantum theory of optical coherence,” Phys. Rev., 130 2529 (1963). http://dx.doi.org/10.1103/PhysRev.130.2529 PHRVAO 0031-899X Google Scholar

15. 

R. J. Glauber, “Coherent and incoherent states of the radiation field,” Phys. Rev., 131 2766 (1963). http://dx.doi.org/10.1103/PhysRev.131.2766 PHRVAO 0031-899X Google Scholar

16. 

Y. H. Shih, “Entangled photons,” IEEE J. Sel. Topics Quantum Electron., 9 1455 (2003). http://dx.doi.org/10.1109/JSTQE.2003.820927 IJSQEN 1077-260X Google Scholar

17. 

Y. Shih, An Introduction to Quantum Optics, Taylor & Francis, Boca Raton, Florida (2011). Google Scholar

18. 

T. Pittman, “It’s a good time for time-bin qubits,” Physics, 6 110 (2013). http://dx.doi.org/10.1103/Physics.6.110 PHYSGM 1943-2879 Google Scholar

Biography

Sanjit Karmakar is a postdoctoral fellow at the U.S. Army Research Laboratory, Adephi, Maryland, USA, and the University of Maryland, Baltimore County, Maryland, USA. He received his PhD in applied physics from the University of Maryland, Baltimore County, USA, in 2012. His current research interests include quantum networks, quantum imaging, and applications of quantum optics in biomedical sciences. He is a member of SPIE.

Ronald E. Meyers is a physicist at the U.S. Army Research Laboratory, Adephi, Maryland, USA. He is a principle investigator in physics research. He has more than 100 scientific publications and more than 20 patents in science and technology addressing significant problems of national interest. His current research interests are focused on fundamental quantum physics, quantum imaging, quantum networking, quantum technology, and quantum information. He is a conference chair of SPIE.

Yanhua Shih is a professor of physics at the University of Maryland, Baltimore County, USA. He received his PhD in physics from the University of Maryland, College Park, USA, in 1987. His recent research interests include fundamental problems of quantum theory, experimental general relativity, quantum information processing, quantum optics, nonlinear optics, and laser physics. He is a conference chair of SPIE.

© 2015 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2015/$25.00 © 2015 SPIE
Sanjit Karmakar, Ronald E. Meyers, and Yanhua Shih "Noninvasive high resolving power entangled photon quantum microscope," Journal of Biomedical Optics 20(1), 016008 (8 January 2015). https://doi.org/10.1117/1.JBO.20.1.016008
Published: 8 January 2015
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Microscopes

Spectral resolution

Confocal microscopy

Photonic microstructures

Tissues

Charge-coupled devices

Imaging systems

Back to Top