1.IntroductionMany key areas of research and development in the biosciences and biomedical technology areas are progressing rapidly. One of the more exciting areas is biomedical imaging. Unfortunately, the light from high-resolving microscopes tends to damage cells. A high resolving power noninvasive microscope, capable of imaging without damaging cells or tissues, would enable physicians and scientists to better see and understand tissue and organ function. The resolving power of a microscope can be improved in several ways. One example was the development of confocal microscopes, which provided better resolving power as well as higher contrast with respect to standard wide-field microscopy. Two other approaches have been proposed and analyzed by Simon and Sergienko1,2 to improve the resolution of standard confocal microscopy. In one of these approaches, Simon and Sergienko used photon correlation, which improved resolution up to 50%.1 In the other approach, they used a degenerate entangled photon pair generated from spontaneous parametric down-conversion (SPDC) to improve the resolution up to 77%.2 Simon and Sergienko suggested the use of a nonlinear crystal after the object to illuminate the specimen with a high pump power to generate sufficient signal and idler photons. To obtain visible-light images with UV-level resolution, they also suggested the use of a UV pump to generate signal and idler photons in the visible range. However, because the specimen was illuminated by high-power UV light, there was a high probability of damaging the specimen, i.e., the cells and tissues studied in biomedical and biosciences. In 1995 at UMBC, while investigating Einstein-Podolsky-Rosen phenomena, Pittman et al.3 discovered ghost imaging using entangled photons to transmissively image a stencil mask internal to their experimental setup. Meyers et al.,4,5 in 2007 and 2008, generated the first ghost image of a remote object by imaging a small toy soldier placed remotely from their setup at the Army Research Laboratory. This demonstration was the first practical application of ghost imaging, and it showed that ghost imaging could be applied to remote sensing from astrophysical to microscopic scales with the potential of increased resolution, increased contrast, and mitigation of adverse effects from distorting media. In 2010, Karmakar and Shih6,7 reported on an experimental demonstration of two-color ghost imaging with enhanced resolving power. The resolution of two-color ghost imaging was also studied theoretically by Rubin and Shih8 and Chan et al.9 Certainly, disturbances around the object can degrade the image contrast of a microscope. However, in 2011, Meyers et al.10,11 showed that disturbances, such as turbulence and distortions, do not affect multiphoton interference ghost imaging. While these studies used the quantum properties of thermal light, the Meyers et al. turbulence-free ghost imaging design also applies to entangled photons. In addition, absorption of photons by an object, though this leads to image information in conventional imaging, can potentially degrade the correlations in certain correlation-based imaging. An advantage of entangled photon imaging is that it eliminates most noise photons that can degrade correlation techniques and relatively few entangled photons are needed to accurately determine features.3 However, accidental coincidences will form a noise background. With good detection equipment, accidental coincidences will be minimized. Furthermore, positive-negative ghost imaging techniques have been developed and experimentally verified, which achieve near 100% contrast by eliminating the effect of the noise background,11,12 which improves images even using thermal light two-photon interference ghost imaging. In this report, we propose a quantum microscope configuration capable of achieving high resolving power without any destruction of cells or tissues by using two-color ghost imaging technology. Such a quantum microscope configuration could provide improved image contrast and would not be affected by disturbances around the object. 2.Proposed ConfigurationIn the proposed experimental configuration, two-color ghost imaging technology is applied to improve the resolving power of a microscope system. Figure 1 shows a schematic of the experimental setup. By recording the joint-detection counts between the bucket detector and the CCD as a function of the transverse plane coordinates of the CCD in the idler beam, the image of the object aperture is observed with a magnification factor that is the same as the degenerate one-color case3 as well as the classical case. In this experimental configuration, entangled photon pairs with generated from SPDC can reproduce a ghost image with enhanced resolving power by a factor of by means of higher imaging amplification, where is the wavelength of the radiation that illuminates the object and is the wavelength of the idler beam. In this case, image amplification is enhanced by a factor of more than that of the one-color ghost imaging or that of a classical imaging setup. Since the lowest limit of transparency of the nonlinear crystals for SPDC is , we propose the use of a 197-nm excimer laser as a pump. A type I BBO crystal together with this pump can be used as an entangled two-photon source to generate a signal beam of 695 nm in the visible range and an idler beam of 275 nm in the UV range. This configuration would improve the magnification power of a quantum microscope by a factor of 2.5. If one could produce a two-photon source capable of generating in the visible region and in the x-ray region, then the magnification power could be improved by a factor of 500. 3.Discussion and AnalysisIn the two-color case, the signal and the idler photon pair from the SPDC source are emitted at different angles and opposite directions. Furthermore, if one photon is emitted at a certain angle, the emission angle of its conjugate is determined by the certain angle of that photon with unit probability. This permits us to explain the experiment in terms of geometrical optics by unfolding the schematic in Fig. 1 into that shown in Fig. 2 by considering the SPDC source as a hinge. The geometrical rays in Fig. 2 represent the two-photon amplitudes, therefore, the image is formed in terms of coincidence counts when the object aperture, imaging lens, and CCD are located according to the modified Gaussian thin lens equation. The relationship among the object aperture’s optical distance from the lens, , the image plane’s optical distance from the lens, (which is actually the distance from the lens back to SPDC source and then straight forward to the image plane), and the focal length of the imaging lens, , satisfies a modified Gaussian thin-lens equation for two colors:6,8 where and and are the angular frequencies of the signal and idler beam, respectively, and the Gaussian thin-lens equation in classical as well as in the degenerate case is where and are the object distance and image distance of the standard degenerate ghost imaging and classical imaging experimental setup.By using the Glauber theory of photodetection, the joint detection probability between two detectors and CCD located at and , respectively, can be expressed as follows:14,15 where , are the negative and positive frequency fields at space-time points and , respectively, and is the two-photon amplitude, also called the two-photon effective wavefunction. Since imaging takes place on the transverse plane, we can calculate the transverse part of the two-photon effective wavefunction at the point (): where and are the Green’s functions evaluated by propagating the signal from the output plane of the two-photon source to the object plane and the idler from the output plane of the two-photon source to a CCD, where defines the transverse coordinates of the object plane.16,17 Consider that an arbitrary phase of is introduced on the transverse coordinates of the object plane by disturbances around the object. Then the two-photon wavefunction becomes:Since , the disturbances have no effect on the formed ghost image in terms of coincidences. After simplification and using the modified Gaussian thin-lens equation shown in Eq. (1) in Eq. (5), we obtain the generalized two-photon effective wavefunction for this case as: where the point-spread function is a result of the finite size of the lens (radius ) and defines the transverse coordinates of the image plane.A collection lens along with detector can be treated as a bucket detector, which is shown in Figs. 1 and 2. The single photon detectors used in quantum imaging are usually very fast, have low dead time, high time resolution, and support high count rates (typically 32 ns dead time, 300 ps timing resolution, and also support up to ). These features of single-photon detector allow fast accumulation of necessary counts. These features of single-photon detectors allow fast accumulation of necessary counts. Fast single photon detectors allow an integration time of approximately a few minutes for the reported quantum microscopy. Generally, this integration time is less than the life time of the specimen. Faster detection methods currently in development, such as those described by Pittman,18 will further reduce the integration times needed for quantum microscopy. The bucket detector integrates all reflected and scattered from the object aperture as a joint-detection event. This process can be written as in the following convolution: where the CCD is scanned in the image plane . The image is observed in terms of coincidence counts () as a function of the transverse plane coordinates of the CCD ().The finite size of the image spot defined by a point-spread function in Eq. (7) determines the angular resolution of the imaging. According to Rayleigh’s criterion, the images of two nearby point objects are said to be unresolvable when the center of one point-spread function falls on the first minimum of the point-spread function of the other. To clarify this situation, consider a point located at on the object plane. The corresponding unique spot in the image plane is determined by the point-spread function defined in Eq. (7). Rayleigh’s criterion defines the size of the point-spread function by considering its first minimum, i.e., Now, we consider another nearby point on the object plane at . In order to separate the image of from that of , the value of cannot be smaller than , and hence where is the magnification factor of the image. Thus, we have the angular limit of spatial resolution for two color ghost imaging asLast, under the approximation , the two-photon effective wavefunction becomes with the modified image forming equation as where .Compared to the degenerate ghost imaging system17 and classical imaging system, we found three interesting features for this quantum microscope system.
Comparing the two-color case with the degenerate case and with the classical imaging system with a longer wavelength , we achieve the same angular resolution limit . When the image plane is moved away from the imaging lens by a factor of in order to form an image while keeping the above conditions steady, the imaging amplification increases by a factor of . In this case, the improved amplification factor is 2.5. As stated above, greater imaging amplification under the enhanced resolving power of an imaging system is a useful and important feature in a microscope. In case of the one-color ghost imaging as well as the classical case, we can resolve with visible light when the object distance () from the imaging lens is 4 cm. The size of human cells and tissues are of the order of . Again, this proposed entangled photon quantum microscope technique would improve the resolving power by a factor of 2.5 over that of a classical wide-field microscope. In this case, the image contrast would be unaffected by disturbances around the object. Furthermore, in this proposed configuration, cells and tissues would be illuminated by visible light instead of UV light, and this would minimize the potential for cell or tissue damage. 4.Comparison of Features of MicroscopyIn this section, we present a table to compare the resolving power of different kinds of microscopes in terms of the resolving power of a standard wide-field microscope (). We also compare the image contrast and specimen safety. The invention of confocal microscopy has provided two important advantages over wide-field microscopes: (1) better resolving power and (2) very good contrast. Later on, the advent of a twin-photon confocal microscope reported improved resolving power significantly beyond the standard confocal microscope. Although twin-photon confocal microscopes can provide very good resolving power, there remains a high probability for cell or tissue damage. In this article, we have reported on a quantum noninvasive, high-resolution microscope that comprises all of the desired features, as shown in Table 1. Table 1Comparison of features of microscopes.
5.ConclusionThe main goal of a modern microscope is to achieve high resolving power, high contrast, and noninvasive imaging. Yet, it has been difficult to design a microscope to achieve all of these features together as shown in the comparison of current microscopes above. Nevertheless, the proposed quantum microscope reported here using two-color ghost imaging technology will be able to achieve all the desired features. Here, cells and tissues will be illuminated by one color in the visible range. The resolving power will be determined by the ratio of the signal and idler photon wavelengths, where, for this case, the ratio is 2.5. As a result, the proposed quantum microscope would be noninvasive with high resolving power. In addition, configurations involving virtual images are conceivable, which may increase magnification. Note that in alternate configurations, the resolving power could be improved further by increasing the wavelength ratio of the signal and idler photons since images are formed in terms of coincidence counts between a bucket detector and the CCD. Significantly, image disturbances, such as from motion of intervening fluid in biological samples or from media turbidity, will have virtually no effect on the contrast of the images in this system. We find that such a quantum microscope would be widely applicable in advancing the progress of biomedical research and medical technologies. AcknowledgmentsThe authors thank K. Deacon, A. Tunick, P. Hemmer, and J. Sprigg for helpful discussions. S. Karmakar is a postdoctoral fellow at ARL (W911NF-11-2-0074). ReferencesD. S. SimonA. V. Sergienko,
“The correlation confocal microscope,”
Opt. Express, 18 9765
(2010). http://dx.doi.org/10.1364/OE.18.009765 OPEXFF 1094-4087 Google Scholar
D. S. SimonA. V. Sergienko,
“Twin-photon confocal microscopy,”
Opt. Express, 18 22147
(2010). http://dx.doi.org/10.1364/OE.18.022147 OPEXFF 1094-4087 Google Scholar
T. B. Pittmanet al.,
“Optical imaging by means of two-photon quantum entanglement,”
Phys. Rev. A, 52 R3429
(1995). http://dx.doi.org/10.1103/PhysRevA.52.R3429 PLRAAN 1050-2947 Google Scholar
R. E. MeyersK. S. DeaconY. Shih,
“A new two-photon ghost imaging experiment with distortion study,”
J. Mod. Opt., 54
(16), 2381
–2392
(2007). http://dx.doi.org/10.1080/09500340701400117 JMOPEW 0950-0340 Google Scholar
R. E. MeyersK. S. DeaconY. Shih,
“Ghost-imaging experiment by measuring reflected photons,”
Phys. Rev. A, 77 041801
(2008). http://dx.doi.org/10.1103/PhysRevA.77.041801 PLRAAN 1050-2947 Google Scholar
S. KarmakarY. Shih,
“Two-color ghost imaging with enhanced angular resolving power,”
Phys. Rev. A, 81 033845
(2010). http://dx.doi.org/10.1103/PhysRevA.81.033845 PLRAAN 1050-2947 Google Scholar
S. KarmakarY. Shih,
“Observation of two-color ghost imaging,”
Proc. SPIE, 7702 770204
(2010). http://dx.doi.org/10.1117/12.851188 PSISDG 0277-786X Google Scholar
M. H. RubinY. Shih,
“Resolution of ghost imaging for nondegenerate spontaneous parametric down-conversion,”
Phys. Rev. A, 78 033836
(2008). http://dx.doi.org/10.1103/PhysRevA.78.033836 PLRAAN 1050-2947 Google Scholar
K. W. C. ChanM. N. O’SullivanR. W. Boyd,
“Two-color ghost imaging,”
Phys. Rev. A, 79 033808
(2009). http://dx.doi.org/10.1103/PhysRevA.79.033808 PLRAAN 1050-2947 Google Scholar
R. E. MeyersK. S. DeaconY. H. Shih,
“Turbulence-free ghost imaging,”
Appl. Phys. Lett., 98 111115
(2011). http://dx.doi.org/10.1063/1.3567931 APPLAB 0003-6951 Google Scholar
R. E. MeyersK. S. DeaconY. H. Shih,
“Positive-negative turbulence-free ghost imaging,”
Appl. Phys. Lett., 100 131114
(2012). http://dx.doi.org/10.1063/1.3698158 APPLAB 0003-6951 Google Scholar
H. ChenT. PengY. Shih,
“100% correlation of chaotic thermal light,”
Phys. Rev. A, 88 023808
(2013). http://dx.doi.org/10.1103/PhysRevA.88.023808 PLRAAN 1050-2947 Google Scholar
D. N. Klyshko,
“Combine EPR and two-slit experiments: interference of advanced waves,”
Phys. Lett. A, 132 299
(1988). http://dx.doi.org/10.1016/0375-9601(88)90856-0 PYLAAG 0375-9601 Google Scholar
R. J. Glauber,
“The quantum theory of optical coherence,”
Phys. Rev., 130 2529
(1963). http://dx.doi.org/10.1103/PhysRev.130.2529 PHRVAO 0031-899X Google Scholar
R. J. Glauber,
“Coherent and incoherent states of the radiation field,”
Phys. Rev., 131 2766
(1963). http://dx.doi.org/10.1103/PhysRev.131.2766 PHRVAO 0031-899X Google Scholar
Y. H. Shih,
“Entangled photons,”
IEEE J. Sel. Topics Quantum Electron., 9 1455
(2003). http://dx.doi.org/10.1109/JSTQE.2003.820927 IJSQEN 1077-260X Google Scholar
Y. Shih, An Introduction to Quantum Optics, Taylor & Francis, Boca Raton, Florida
(2011). Google Scholar
T. Pittman,
“It’s a good time for time-bin qubits,”
Physics, 6 110
(2013). http://dx.doi.org/10.1103/Physics.6.110 PHYSGM 1943-2879 Google Scholar
BiographySanjit Karmakar is a postdoctoral fellow at the U.S. Army Research Laboratory, Adephi, Maryland, USA, and the University of Maryland, Baltimore County, Maryland, USA. He received his PhD in applied physics from the University of Maryland, Baltimore County, USA, in 2012. His current research interests include quantum networks, quantum imaging, and applications of quantum optics in biomedical sciences. He is a member of SPIE. Ronald E. Meyers is a physicist at the U.S. Army Research Laboratory, Adephi, Maryland, USA. He is a principle investigator in physics research. He has more than 100 scientific publications and more than 20 patents in science and technology addressing significant problems of national interest. His current research interests are focused on fundamental quantum physics, quantum imaging, quantum networking, quantum technology, and quantum information. He is a conference chair of SPIE. Yanhua Shih is a professor of physics at the University of Maryland, Baltimore County, USA. He received his PhD in physics from the University of Maryland, College Park, USA, in 1987. His recent research interests include fundamental problems of quantum theory, experimental general relativity, quantum information processing, quantum optics, nonlinear optics, and laser physics. He is a conference chair of SPIE. |