14 February 2020 Selective retina therapy monitoring by speckle variance optical coherence tomography for dosimetry control
Author Affiliations +
Abstract

Significance: Selective retina therapy (SRT) selectively targets the retinal pigment epithelium (RPE) and reduces negative side effects by avoiding thermal damages of the adjacent photoreceptors, the neural retina, and the choroid. However, the selection of proper laser energy for the SRT is challenging because of ophthalmoscopically invisible lesions in the RPE and different melanin concentrations among patients or even regions within an eye.

Aim: We propose and demonstrate SRT monitoring based on speckle variance optical coherence tomography (svOCT) for dosimetry control.

Approach: M-scans, time-resolved sequence of A-scans, of ex vivo bovine retina irradiated by 1.7-μs duration laser pulses were obtained by a swept-source OCT. SvOCT images were calculated as interframe intensity variance of the sequence. Spatial and temporal temperature distributions in the retina were numerically calculated in a 2-D retinal model using COMSOL Multiphysics. Microscopic images of treated spots were obtained before and after removing the upper neural retinal layer to assess the damage in both RPE and neural layers.

Results: SvOCT images show abrupt speckle variance changes when the retina is irradiated by laser pulses. The svOCT intensities averaged in RPE and photoreceptor layers along the axial direction show sharp peaks corresponding to each laser pulse, and the peak values were proportional to the laser pulse energy. The calculated temperatures in the neural retina layer and RPE were linearly fitted to the svOCT peak values, and the temperature of each lesion was estimated based on the fitting. The estimated temperatures matched well with previously reported results.

Conclusion: We found a reliable correlation between the svOCT peak values and the degree of retinal lesion formation, which can be used for selecting proper laser energy during SRT.

1.

Introduction

Selective retina therapy (SRT) is an effective laser treatment method for various retinal diseases associated with degradation of the retinal pigment epithelium (RPE), such as diabetic macular edema, central serous chorioretinopathy, and age-related macular degeneration.16 The RPE, which contains a high concentration of melanosomes, absorbs 50% to 60% of incident green light. However, in order to selectively target the RPE layer, the laser pulse duration needs to be shorter than a thermal relaxation time of the RPE (10  μs).7 The SRT reduces negative side effects and facilitates healing of the induced retinal lesions by avoiding thermal damages of the adjacent photoreceptors, the neural retina, and the choroid. However, the selection of proper laser energy—which is crucial for successful SRT without excessive burning and collateral damage—is challenging because lesions in the RPE are ophthalmoscopically invisible. In addition, different melanin concentrations among patients, or regions even within an eye,8 make it impossible to set a static threshold value of pulse energy of a therapeutic irradiation window.

Fundus fluorescence angiography (FFA) is an accurate method to detect the lesions, but it requires the use of fluorescent dye injection9 and a long delay between treatment and detection. For real-time noninvasive SRT monitoring, several approaches have been proposed. These include the detection of microbubble formation and collapse, which induce mechanical disruption and damage to RPE cells.10 This approach measures the acoustic transient11 or light reflection changes.12 Although these methods have already been used in several clinical studies,11,13,14 they do not provide visual feedback during the treatment. Optical coherence tomography (OCT), which can provide depth-resolved imaging, was also applied for the SRT monitoring.1519 The treatments are considered successful when OCT signal variations, i.e., intensity decrease, are detected, and the results show good agreement with the evaluation of lesions by FFA.

Speckle variance OCT (svOCT) quantifies the speckle pattern variation caused by moving particles or structural changes in biological tissues. It calculates the interframe intensity variance of a sequence of structural OCT images. The svOCT has been extensively developed in recent years for OCT angiography, which is used to visualize retinal microvasculatures;20 it has also been shown to be able to monitor protein denaturation and coagulation.21 Thus, it is expected that svOCT could be an effective way to detect speckle variation changes induced by morphological and structural changes of retinal tissue during the thermal-induced microbubble formation and collapse by laser irradiation. In this paper, we studied and demonstrated SRT monitoring based on the svOCT. A swept-source OCT imaging system integrated with a microsecond pulsed laser system was used for ex vivo bovine retina study. SvOCT images corresponding to various laser pulse energies and the various number of frames were obtained. The svOCT values of RPE and photoreceptor layers were averaged along the axial direction, and peak values of the svOCT at each pulse laser irradiation were analyzed. The microscopic images of the treated spots were taken before and after removing the upper neural retinal layers to assess the degree of retina and RPE damage. Spatial and temporal thermal effects in retina induced by pulse laser irradiation were simulated and correlated to the peak values of svOCT.

2.

Experimental Method

An in-house built swept-source OCT imaging system was integrated with a frequency-doubled Nd:YLF laser-based SRT system (Lutronic, Goyang, Korea). The schematic of the system is shown in Fig. 1. The OCT system used a commercial swept-source engine (Axsun Technologies Inc., Billerica, Massachusetts) operating at 100-kHz sweep rate. The center wavelength and sweeping bandwidth of the system were 1060 and 100 nm, respectively. The OCT laser was combined with the pulse laser using a dichroic mirror; galvano mirrors were used to direct the OCT laser to a treated spot on the retina. The wavelength of the pulse laser was 527 nm, and the pulse laser operated at 100-Hz repetition rate and 1.7-μs duration. The pulse laser energy was adjusted from 22 to 190  μJ using neutral-density filters. In this study, fresh ex vivo bovine eyes were acquired from a local butcher and immersed in a cooled saline solution. Bovine eyes have tapetum fibrosum, which has the retinal epithelial layer completely unpigmented, over the central and mid-region of the retina.22 Because the RPE layer exists in the periphery of the retina, it was difficult to focus the beam on the RPE layer using the crystalline lens of an eye itself. Therefore, the bovine eye’s cornea and lens were removed, and the beam was focused on a tilted eye using an objective lens. A total of 39 treated spots were tested on two eyes. M-mode OCT images of the bovine retina were acquired during the laser pulse irradiation.

Fig. 1

Schematic of a swept-source OCT system integrated into a pulse laser system. BD, balanced detector; OL, objective lens.

JBO_25_2_026001_f001.png

SvOCT images were calculated by

SVij=1Nk=(i1)*N+1i*N[Ik(j)1Nl=(i1)*N+1i*NIl(j)]2,
where i, j, and k are the indices of the frame of the svOCT images, axial position, and the frame of M-scans, respectively, and N is the number of frames used for variance calculation. The photoreceptor and RPE layers, which are highly scattering and absorptive, were set as a region of interest (ROI), and the svOCT values in the ROI were averaged along the axial direction. Microscopic images of the treated spots were obtained using a CCD camera (DCC1645C, Thorlabs, Newton, New Jersey) and a 10× magnification zoom lens.

Spatial and temporal temperature distributions in the retina were numerically calculated by COMSOL Multiphysics software. Spatial distribution was calculated in 2-D, and the geometry of the bovine retina model used for the simulation is shown in Fig. 2(a). The RPE was modeled as a 7-μm layer containing melanosomes that were assumed as spheres of radius 0.3  μm by a discrete absorber model.23 Melanosomes were diagonally distributed, and the distance between adjacent melanosomes was set to 1.2  μm. The model applied a heat equation as shown below:

ρCpTt=·(kT)+Q,
where T is the temperature as a function of time (t) and spatial coordinate x and y, ρ is the density, k is the thermal conductivity, and Cp is the heat capacity of the material. The Q refers to the heat source from laser irradiation. The coefficient values used are shown in Table 1, which include the thickness, absorption coefficient, and thermal physical constant values25 of each retinal layer. The absorption coefficient of retinal melanosomes at 532-nm wavelength was estimated from 2370 to 13000  cm1,10,26 and 6500  cm1, which makes the absorption in RPE around 50% of the total, was used for our simulation. The duration and frequency of the pulse laser were set to 2  μs, 5  μs or 10  μs and 100 Hz, respectively; the peak temperature was calculated over 1.5 ms. Gaussian laser-beam profile with a diameter of 150  μm was used for the simulation.

Fig. 2

(a) The geometry of the bovine retina model. The retina was assumed to consist of two layers, neural retina and RPE, and has immediate contact with a choroid. The RPE was modeled as a 7-μm layer containing melanosomes that were assumed as diagonally distributed spheres of radius 0.3  μm. (b) Temperature–time dependence in the neural retina and at the melanosome surface in RPE when laser pulse of energy 50  μJ irradiated. Spatial distribution of temperature around RPE, when the temperature of the neural retina reached a maximum after (c) 20-μJ, (d) 50-μJ and (e) 100-μJ pulse irradiations.

JBO_25_2_026001_f002.png

Table 1

Thickness, absorption coefficient, and thermal properties of each retinal layer.

Neural retinaRPE layer outside of melanosomeMelanosomeChoroidSclera
Thickness (μm)27070.3 (radius)200700
Absorption coefficient (cm1)10.42406500245244.9
Heat capacity (J/kg·K)36803680368036804178
Density (kg/m3)10001000100010001000
Thermal conductivity (W/m·K)0.5650.5650.5650.5300.58

The temperature of the neural retina was calculated by averaging the temperature in a rectangular region (30  μm×20  μm) 2.5  μm away from RPE, and the temperature of the melanosome surface in RPE was calculated by averaging the temperature on seven melanosome surfaces at the first two melanosome layers located in the center of the Gaussian beam. Figure 2(b) shows the time-dependent temperature variation in the neural retina (in orange) and at the melanosome surface in RPE (in blue). Spatial distributions of temperature around RPE, when the temperature of neural retina reached a maximum after a 20-, 50-, and 100-μJ pulse irradiation, are shown in Figs. 2(c)2(e). The calculated peak temperatures in each region were correlated to the peak values of svOCT and tissue damage.

3.

Results

Figures 3(a) and 3(b) show an M-scan OCT image and the corresponding svOCT image of the bovine retina when 108-μJ energy per pulse irradiated. The M-scan OCT image shows the visible temporal signal variations induced by the laser pulse at the moment marked with white triangles. The signal variation increases the svOCT value. Figure 3(c) shows svOCT values averaged in ROI, and it shows a distinctive peak for each pulse irradiation.

Fig. 3

(a) M-mode OCT image of the bovine retina and corresponding (b) svOCT image. Photoreceptor and RPE layers, which are highly scattering and absorptive, were set as an ROI. (c) Axially averaged svOCT values in the ROI during pulse laser irradiation. White triangles mark the moment when each laser pulse (108  μJ) irradiated.

JBO_25_2_026001_f003.png

To find appropriate N for calculating the svOCT image, we first tested different values of N and compared the average peak values of the svOCT spikes, the standard deviation of those peaks, and background noise level. Since the speckle variation induced by each laser pulse lasts for only around 50  μs, only five frames of the speckle variances show high signal while the rest show very low signal. Intuitively, if we choose a larger window size N, the lower portion of sv signal in the window will be significant, and it will decrease the overall value of the sv signal inside the window, i.e., the peak value of svOCT spikes. Furthermore, the variance of those peak values will also decrease. Figures 4(a)4(d) show how the average peak values (shapes) and standard deviation (error bar) change depending on laser energy level when N is 2, 5, 10, and 20, respectively. We can see that, as expected, both the average peak value and the standard deviation decrease when N increases. Relative standard deviation, defined as the ratio of the standard deviation to the mean, was 0.53, 0.46, 0.43, and 0.46 for N of 2, 5, 10, and 20, respectively. Therefore, a mid-range of N values between 5 and 10 are suitable for calculating svOCT in terms of the precision and repeatability. In addition, note that the background noise level increases with increasing N. For N of 2, 5, 10, and 20, the average upper bound levels of background noise were 0.54±0.24, 0.74±0.31, 1.03±0.41, and 1.64±0.66, respectively. The background noise levels were bounded by μ+3σ upper limit, where μ was svOCT values averaged in ROI before or after laser irradiation and σ was the standard deviation. The increase in background noise was caused by the bulk motion of the sample and other environmental changes. Shorter integration time will decrease these effects and so will the smaller value of N. Considering both the effects of relative standard deviation and background noise levels, we choose N as 5.

Fig. 4

Mean (shapes) and standard deviation (error bar) of peak values of svOCT values averaged in ROI depending on laser pulse energy when window size N is (a) 2, (b) 5, (c) 10, and (d) 20.

JBO_25_2_026001_f004.png

Figure 5 shows svOCT values averaged in ROI depending on pulse laser energy when N is 5. Peak values increased with increasing pulse laser energy, and the distinctive peaks were observed when the laser pulse energy was as low as 54  μJ. Figures 6(a) and 6(b) show microscopic images of the treated spots before and after peeling upper neural retinal layers off, respectively. The energy level of each spot is shown in Fig. 6(c). The denaturation of neural retina can be observed as whitish spots (pointed by white triangles) in Fig. 6(a), but the lesions confined only to the RPE layer are invisible. The leftmost and rightmost columns are high-energy lesions marking pattern of the spots. In Fig. 6(b), lesions in the RPE layer (pointed by arrows) can be detected by peeling off the upper neural retinal layers.

Fig. 5

SvOCT values averaged in ROI when pulse laser energy is (a) 150, (b) 108, (c) 86, (d) 69, (e) 54, and (f) 30  μJ.

JBO_25_2_026001_f005.png

Fig. 6

Microscopic image of the retina (a) before and (b) after peeling neural retinal layers off. (c) The energy level of treated spots.

JBO_25_2_026001_f006.png

Average peak values of svOCT depend on the laser pulse energy; the result is summarized in Fig. 7(a). The data with blue circles indicate when the laser pulses induce a lesion in the upper neural layers in addition to inducing a lesion in the RPE layer [corresponding to the indicated spots in Fig. 6(a)]. The data represented by orange squares represent the cases when the lesions were induced only on the RPE layer, indicated by white triangle arrows in Fig. 6(b). The laser-induced lesion confined only to the RPE layer can be considered as a successful treatment. Fifteen peaks for each spot are averaged, and the standard deviation is shown by the error bar. As expected, it was difficult to define the threshold energy level that induces lesion only in the RPE layer. If the null hypothesis is defined as a successful treatment and the treatment is decided to be successful when the energy level is in the range from 41.9 to 92.1  μJ, which is determined by logistic regression, the type I error and the type II error were 26.7% and 25%, respectively. Compared with the energy level, the average peak values of svOCT showed a better correlation with the lesion creation. Average sv peak values ranged from 12.4 to 38.7 when the induced lesions were observed in both neural and RPE layers; sv values ranged from 0.3 to 18.3 when the lesion was confined only to the RPE layer. In the range of sv peak values from 0.7 to 1.9, no induced lesion was observed at all. When the treatment is decided to be successful with the average svOCT peak value between 1.88 and 15.3 based on logistic regression, the type I error and the type II error were 20% and 0.083%, which was better than the case when the threshold was set by the energy level. For dosimetry control, the method is designed to be used with a power ramped pulsed mode, in which laser energy increases linearly from pulse to pulse within one pulse train and automatically stops the next pulse irradiation when the svOCT peak values reached the predetermined threshold value.

Fig. 7

(a) Averaged peak values depending on pulse laser energy and damage range. (b) Simulated (lines) and estimated temperature from the svOCT intensity (shapes) at neural retina and RPE. (c) Simulated temperature at the neural retina and the RPE as a function of laser energy level for three pulse durations—2, 5, and 10  μs.

JBO_25_2_026001_f007.png

In addition, the average peak value of svOCT intensities was correlated to the simulated temperature of the neural retina and melanosome surface in RPE. The linear regression of average peak values of svOCT on laser energy was calculated for photocoagulated lesions and selectively damaged lesions. Then, each of them was correlated to the simulated temperature of neural retina and RPE because the tissue damage process was different from each other. Since the simulated temperature of melanosome surface in the RPE was also linear to pulse laser energy, it was correlated to the average peak values of selectively damaged lesions as

TM=124.5+12.4P,
where TM is the temperature of melanosome surface in RPE and P is the average peak values of svOCT. Figure 7(b) shows the simulated temperature (solid line) and estimated temperature (square shapes) from the average peak values of svOCT. Most of the estimated temperatures at RPE were higher than 150°C; the lowest temperature was 127.7°C. This is reasonable since the microvaporization is known to occur at around 150°C. Similarly, the average peak values of the photocoagulated lesions were correlated to the temperature of the neural retina and empirically fitted to a line as
TN=25.5+1.5P.

Figure 7(b) shows that the temperature of the neural retina was mostly estimated to increase higher than 50°C when the neural retina was photocoagulated.

We also simulated temperatures of the neural retina and the melanosome surface in RPE as a function of laser energy level for three pulse durations—2, 5, and 10  μs, as shown in Fig. 7(c). The temperature of the neural retina does not change significantly, but the temperature of the melanosome surface in RPE decreases as pulse duration increases. The decrease in temperature can be explained by less heat confinement with longer pulse duration due to heat diffusion during irradiation.

4.

Conclusion

It was shown that the SRT could be successfully monitored by the svOCT imaging system when integrated with the SRT system. We tested our system performance using ex vivo bovine eyes; the svOCT showed distinctive signal variation corresponding to each laser pulse irradiation. The signal variations were proportional to pulse energy levels, and it had a reliable correlation with the creation of lesion within the retina. The temperature at the neural retina and RPE was estimated by svOCT peak values using temperature simulation results, which was consistent with the observed lesion creation. However, we could have missed some minor tissue damages when assessing the photocoagulation and RPE cell damage from the microscopic images. Therefore, more studies that incorporate further analysis supported by histology or fluorescence microscopy would be needed to obtain a more accurate correlation between the svOCT signal and retinal damage range. In addition, we plan to perform in vivo studies using a live animal model to fully validate the utility of this method as an automatic dosimetry control in clinical SRT systems.

Disclosures

The work at Johns Hopkins University was partly supported by Lutronic Corporation; however, the authors have no relevant financial interests in this article and no other potential conflicts of interest to disclose.

Acknowledgments

This work was supported by the World Class 300 Project (No. S2340708) of the Small and Medium Business Administration, Republic of Korea.

References

1. 

J. Roider et al., “Selective retina therapy (SRT) for clinically significant diabetic macular edema,” Graefe’s Arch. Clin. Exp. Ophthalmol., 248 (9), 1263 –1272 (2010). https://doi.org/10.1007/s00417-010-1356-3 Google Scholar

2. 

Y. G. Park et al., “Safety and efficacy of selective retina therapy (SRT) for the treatment of diabetic macular edema in Korean patients,” Graefe’s Arch. Clin. Exp. Ophthalmol., 254 1703 –1713 (2016). https://doi.org/10.1007/s00417-015-3262-1 Google Scholar

3. 

H. Elsner et al., “Selective retina therapy in patients with central serous chorioretinopathy,” Graefe’s Arch. Clin. Exp. Ophthalmol., 244 (12), 1638 –1645 (2006). https://doi.org/10.1007/s00417-006-0368-5 Google Scholar

4. 

C. Klatt et al., “Selective retina therapy for acute central serous chorioretinopathy,” Br. J. Ophthalmol., 95 83 –88 (2011). https://doi.org/10.1136/bjo.2009.178327 BJOPAL 0007-1161 Google Scholar

5. 

C. Framme et al., “Selective retina therapy in acute and chronic-recurrent central serous chorioretinopathy,” Ophthalmologica, 234 177 –188 (2015). https://doi.org/10.1159/000439188 OPHTAD 0030-3755 Google Scholar

6. 

C. Framme et al., “Autofluorescence imaging after selective RPE laser treatment in macular diseases and clinical outcome: a pilot study,” Br. J. Ophthalmol., 86 1099 –1106 (2002). https://doi.org/10.1136/bjo.86.10.1099 BJOPAL 0007-1161 Google Scholar

7. 

R. Brinkmann et al., “Selective retina therapy (SRT): a review on methods, techniques, preclinical and first clinical results,” Bull. Soc. Belge. Ophtalmol., 302 51 –69 (2006). Google Scholar

8. 

J. J. Weiter et al., “Retinal pigment epithelial lipofuscin and melanin and choroidal melanin in human eyes,” Invest. Ophthalmol. Vis. Sci., 27 (2), 145 –152 (1986). IOVSDA 0146-0404 Google Scholar

9. 

A. Chopdar and T. Aung, “Fundus fluorescein angiography,” Multimodal Retinal Imaging, JP Medical Ltd., Victoria (2014). Google Scholar

10. 

R. Brinkmann et al., “Origin of retinal pigment epithelium cell damage by pulsed laser irradiance in the nanosecond to microsecond time regimen,” Lasers Surg. Med., 27 451 –464 (2000). https://doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

11. 

G. Schuele et al., “Optoacoustic real-time dosimetry for selective retina treatment,” J. Biomed. Opt., 10 (6), 064022 (2005). https://doi.org/10.1117/1.2136327 JBOPFO 1083-3668 Google Scholar

12. 

E. Seifert et al., “Automatic irradiation control by an optical feedback technique for selective retina treatment (SRT) in a rabbit model,” Proc. SPIE, 8803 880303 (2013). https://doi.org/10.1117/12.2033560 Google Scholar

13. 

S. Kang et al., “Selective retina therapy in patients with chronic central serous chorioretinopathy: a pilot study,” Medicine (Baltimore), 95 (3), e2524 (2016). https://doi.org/10.1097/MD.0000000000002524 Google Scholar

14. 

Y. J. Kim et al., “Selective retina therapy with real-time feedback-controlled dosimetry for treating acute idiopathic central serous chorioretinopathy in Korean patients,” J. Ophthalmol., 2018 6027871 (2018). https://doi.org/10.1155/2018/6027871 Google Scholar

15. 

P. Steiner et al., “Time-resolved ultra-high resolution optical coherence tomography for real-time monitoring of selective retina therapy,” Invest. Ophthalmol. Vis. Sci., 56 (11), 6654 –6662 (2015). https://doi.org/10.1167/iovs.15-17151 IOVSDA 0146-0404 Google Scholar

16. 

S. Zbinden et al., “Automatic assessment of time-resolved OCT images for selective retina therapy,” Int. J. Comput. Assist. Radiol. Surg., 11 863 –871 (2016). https://doi.org/10.1007/s11548-016-1383-6 Google Scholar

17. 

D. Kaufmann et al., “Dosimetry control and monitoring of selective retina therapy using optical coherence tomography,” Proc. SPIE, 10416 1041604 (2017). https://doi.org/10.1117/12.2283067 PSISDG 0277-786X Google Scholar

18. 

T. Fountoukidou et al., “Motion-invariant SRT treatment detection from direct M-scan OCT imaging,” Int. J. Comput. Assist. Radiol. Surg., 13 683 –691 (2018). https://doi.org/10.1007/s11548-018-1720-z Google Scholar

19. 

D. Kaufmann et al., “Selective retina therapy enhanced with optical coherence tomography for dosimetry control and monitoring: a proof of concept study,” Biomed. Opt. Express, 9 (7), 3320 –3334 (2018). https://doi.org/10.1364/BOE.9.003320 BOEICL 2156-7085 Google Scholar

20. 

M. S. Mahmud et al., “Review of speckle and phase variance optical coherence tomography to visualize microvascular networks,” J. Biomed. Opt., 18 (5), 050901 (2013). https://doi.org/10.1117/1.JBO.18.5.050901 JBOPFO 1083-3668 Google Scholar

21. 

C. Lee et al., “Feasibility study: protein denaturation and coagulation monitoring with speckle variance optical coherence tomography,” J. Biomed. Opt., 21 (12), 125004 (2016). https://doi.org/10.1117/1.JBO.21.12.125004 JBOPFO 1083-3668 Google Scholar

22. 

F. J. Ollivier et al., “Comparative morphology of the tapetum lucidum (among selected species),” Vet. Ophthalmol., 7 11 –22 (2004). https://doi.org/10.1111/vop.2004.7.issue-1 Google Scholar

23. 

J. Roider et al., “Microphotocoagulation: selective effects of repetitive short laser pulses,” Proc. Natl. Acad. Sci. U. S. A., 90 (18), 8643 –8647 (1993). https://doi.org/10.1073/pnas.90.18.8643 Google Scholar

24. 

D. K. Sardar et al., “Optical characterization of bovine retinal tissues,” J. Biomed. Opt., 9 (3), 624 –631 (2004). https://doi.org/10.1117/1.1688813 JBOPFO 1083-3668 Google Scholar

25. 

S. A. Mirnezami et al., “Temperature distribution simulation of the human eye exposed to laser radiation,” J. Lasers Med. Sci., 4 (4), 175 –181 (2013). Google Scholar

26. 

S. L. Jacques, R. D. Glickman and J. A. Schwartz, “Internal absorption coefficient and threshold for pulsed laser disruption of melanosomes isolated from retinal pigment epithelium,” Proc. SPIE, 2681 1 –10 (1996). https://doi.org/10.1117/12.239608 PSISDG 0277-786X Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Soohyun Lee, Shuwen Wei, Shoujing Guo, Jongmin Kim, Bongkyun Kim, Gihoon Kim, and Jin U. Kang "Selective retina therapy monitoring by speckle variance optical coherence tomography for dosimetry control," Journal of Biomedical Optics 25(2), 026001 (14 February 2020). https://doi.org/10.1117/1.JBO.25.2.026001
Received: 1 June 2019; Accepted: 29 January 2020; Published: 14 February 2020
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Retina

Optical coherence tomography

Pulsed laser operation

Laser energy

Speckle

Imaging systems

Laser irradiation

Back to Top