20 May 2021 Validation of a mammographic image quality modification algorithm using 3D-printed breast phantoms
Author Affiliations +
Abstract

Purpose: To validate a previously proposed algorithm that modifies a mammogram to appear as if it was acquired with different technique factors using realistic phantom-based mammograms.

Approach: Two digital mammography systems (an indirect- and a direct-detector-based system) were used to acquire realistic mammographic images of five 3D-printed breast phantoms with the technique factors selected by the automatic exposure control and at various other conditions (denoted by the original images). Additional images under other simulated conditions were also acquired: higher or lower tube voltages, different anode/filter combinations, or lower tube current–time products (target images). The signal and noise in the original images were modified to simulate the target images (simulated images). The accuracy of the image modification algorithm was validated by comparing the target and simulated images using the local mean, local standard deviation (SD), local variance, and power spectra (PS) of the image signals. The absolute relative percent error between the target and simulated images for each parameter was calculated at each sub-region of interest (local parameters) and frequency (PS), and then averaged.

Results: The local mean signal, local SD, local variance, and PS of the target and simulated images were very similar, with a relative percent error of 5.5%, 3.8%, 7.8%, and 4.4% (indirect system), respectively, and of 3.7%, 3.8%, 7.7%, and 7.5% (direct system), respectively.

Conclusions: The algorithm is appropriate for simulating different technique factors. Therefore, it can be used in various studies, for instance to evaluate the impact of technique factors in cancer detection using clinical images.

© 2021 Society of Photo-Optical Instrumentation Engineers (SPIE) 2329-4302/2021/$28.00 © 2021 SPIE
Joana Boita, Alistair Mackenzie, Ruben E. van Engen, Mireille J. M. Broeders, and Ioannis Sechopoulos "Validation of a mammographic image quality modification algorithm using 3D-printed breast phantoms," Journal of Medical Imaging 8(3), 033502 (20 May 2021). https://doi.org/10.1117/1.JMI.8.3.033502
Received: 28 December 2020; Accepted: 28 April 2021; Published: 20 May 2021
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Breast

Image quality

Detection and tracking algorithms

Mammography

Picosecond phenomena

Polymethylmethacrylate

Computer simulations

Back to Top