9 February 2024 Spiral laser scanning photoacoustic microscopy for functional brain imaging in rats
Author Affiliations +
Abstract

Significance

There are many neuroscience questions that can be answered by a high-resolution functional brain imaging system. Such a system would require the capability to visualize vasculature and measure neural activity by imaging the entire brain continually and in rapid succession in order to capture hemodynamic changes. Utilizing optical excitation and acoustic detection, photoacoustic technology enables label-free quantification of changes in endogenous chromophores, such as oxyhemoglobin, deoxyhemoglobin, and total hemoglobin.

Aim

Our aim was to develop a sufficiently high-resolution, fast frame-rate, and wide field-of-view (FOV) photoacoustic microscopy (PAM) system for the purpose of imaging vasculature and hemodynamics in a rat brain.

Approach

Although the most PA microscopy systems use raster scanning (or less commonly Lissajous scanning), we have developed a simple-to-implement laser scanning optical resolution PAM system with spiral scanning (which we have named “spiral laser scanning photoacoustic microscopy” or sLS-PAM) to acquire an 18 mm diameter image at fast frame rate (more than 1 fps). Such a system is designed to permit continuous rat brain imaging without the introduction of photobleaching artifacts.

Conclusion

We demonstrated the functional imaging capability of the sLS-PAM system by imaging cerebral hemodynamics in response to whisker and electrical stimulation and used it for vascular imaging of a modeled brain injury. We believe that we have demonstrated the development of a simple-to-implement PAM system, which could become an affordable functional neuroimaging tool for researchers.

1.

Introduction

Photoacoustic imaging (PAI) is an emerging hybrid imaging modality, combining the advantages of optical contrast with acoustic penetration.13 PAI has been used extensively in preclinical,410 and more recently, clinical studies.1121 In PAI, upon nanosecond laser irradiation of tissue, chromophores, such as oxy-hemoglobin (HbO2) and deoxy-hemoglobin (Hb), absorb energy and generate photoacoustic waves through the thermoelastic effect.1,2225 The generated waves are detected by ultrasound transducers and processed by an image reconstruction algorithm. PAI has previously been used to provide high-resolution structural and functional images of brain vasculature in small and large animal models.2628 Many neuroscience questions can be answered by a high-resolution functional brain imaging system.22,29,30 For functional imaging, PAI relies on the same principle as that of functional magnetic resonance imaging, where neural activity is indirectly captured through imaging of cerebral hemodynamic fluctuations (i.e., changes in oxygen demand). Many applications of PAI for functional brain imaging have been reported. For example, Wang et al.4 imaged cerebral hemodynamic changes in the rat brain in response to whisker stimulation, hyperoxia and hypoxia; Nasiriavanaki et al.22,31 developed a photoacoustic computed tomography (PACT) system to monitor the cortical resting state functional connectivity in the mouse brain;1 Kang et al. imaged neuronal activity during seizures in the mouse brain;32 Liao et al.33 imaged hemodynamic changes in the rat brain in response to forepaw electrical stimulation; and Janggun et al.34 developed a functional PAI system to monitor regional activation in the rat brain induced by cocaine.

There are two major implementations of PAI: PACT and photoacoustic microscopy (PAM), the latter of which is further divided into two categories, acoustic resolution-PAM and optical resolution PAM (OR-PAM), based on how the focus is achieved.1,35 Although PACT is used for imaging hemodynamic parameters in larger vessels in deeper regions,4,22,3639 PAM is used for imaging fine vessels, i.e., capillaries in shallow regions.4045 Analysis of hemodynamic changes in capillaries, as compared to larger vessels in the brain, provides a more detailed understanding of brain functionality.

Early versions of OR-PAM used 2D galvo scanners to provide an imaging area of 6  mm diameter in about 2 min.46 By incorporating faster scanning hardware, such as microelectromechanical system mirrors, second generation OR-PAM systems became faster. For example, Yao et al.47 developed an OR-PAM system able to image an area of 2.5×4  mm2 in 37 s; Lan et al.48 developed a hexagon-based OR-PAM system that imaged an area of 12×15  mm2 in 12 s; Li et al.49 developed an OR-PAM system using a combination of mechanical and optical scanning able to image an area of 10×8  mm2 within 180 s; Lee et al.50 introduced a fully waterproof 2D galvanometer scanner able to image 8×13  mm2 in 100 s; Zhong et al.51 utilized a cylindrically focused transducer imaging 76  μm×4.5  mm slices simultaneously to capture an area of 4.5×1  mm2 in 1 s. Chen et al.52 developed polygonal scanning method to image an area of 12×12  mm2 in 5 s.

In the implementation of OR-PAM, sometimes an unfocused ultrasound transducer is kept stationary, at an inclined angle, and a laser light is raster scanned point-by-point by means of a 2D galvanometer within the field-of-view (FOV) of the transducer. While not a new type of OR-PAM, this method has been referred to as laser scanning PAM (LS-PAM) in the literature.10,52 LS-PAM has lower signal-to-noise ratio (SNR) compared to other configurations of OR-PAM, mainly due to three reasons: (i) noncoaxial opto-acoustic configuration, (ii) unfocused transducer for detection, and (iii) transducer placement further away from the imaging target.53

The three main scanning configurations that have been explored for optical imaging are raster scanning, Lissajous scanning, and spiral scanning.54 In raster scanning, the imaging beam or stage moves in a back-and-forth pattern across the sample in a grid-like fashion. Although raster scanning allows for precise control over the scanning path and image acquisition, it is slower compared to other scanning techniques and may cause motion artifacts because of the need for stopping and starting at the end of each line. This stop-start movement (fly-back) can result in misalignment between consecutive scan lines and can lead to image distortion or artifacts. Some implementations of raster scanning, such as bidirectional triangular and sinusoidal, have improved on the speed of image acquisition.55,56 However, triangular scans still suffers from interruption times (phase lag) and sinusoidal scans display deformations of the image at the margins because of nonlinearities.55 A thorough comparison of raster scanning functions (triangular, sinusoidal, and sawtooth) is explored in a series of papers.5456 Lissajous scanning involves using orthogonal oscillations to scan the imaging beam across the sample in a two-dimensional pattern using sinusoidal oscillations in both the x and y axes to create a complex scanning trajectory resembling Lissajous figures. It is faster than raster scanning, as it avoids the need for repetitive movements, and avoids motion artifacts, but requires additional complexity in the scanning mechanism and control system. Its scanning pattern requires visiting the same regions on multiple passes, which can increase photobleaching, and requires additional postprocessing because of the nonlinearity of pixel spacing.57 Spiral scanning involves a continuous spiral motion of the imaging beam or stage from the center towards the outer edges of the sample.58 The entire FOV is covered in a single continuous motion without the need for repetitive movements. It enables faster image acquisition compared to raster scanning, as it eliminates back-and-forth movements, minimizes exposure time at each point, and reduces motion artifacts. Spiral scanning also provides a more even distribution of sampling points across the entire scanned area and the density of points remains relatively consistent throughout the spiral trajectory. More details on different implementations of spiral scanning including mathematical details and potential areas for distortion are discussed in Refs. 58 and 59. We previously studied brain activities by imaging the main functional cortical regions using ring array PACT;22,60 one of the motivations for the current study was to enable imaging of cortical subregions within the main functional regions. This has been accomplished by developing a robust, sufficiently large FOV, fast system with reasonable resolution.

2.

Material and Methods

Although the most PA microscopy systems use raster scanning (or less commonly Lissajous scanning), here we develop a simple-to-implement laser scanning OR-PAM system with spiral scanning (sLS-PAM) to acquire a large image at a fast frame rate. The design requirements include ability to continuously image the same region on the brain without the introduction of photobleaching artifacts and a high sensitivity acoustic detection unit.

2.1.

Principle of Spiral Scanning

Spiral scanning is a natural solution to mitigate the scanning inefficiencies of raster scanning methods, due to its ability to form a continuous scan pattern. Spiral scanning can be implemented with different spiral scan functions,58 including constant linear velocity Vs. To construct a system with constant linear velocity, a spiral scanning pattern was constructed by driving x and y mirrors of the galvanometer with x(t)=r(t)cos(ν(t)) and y(t)=r(t)sin(ν(t)) as shown in red and black in Fig. 1(a)(i), respectively. Here ν(t) is the frequency at time t, as shown in Eq. (1); r(t) is amplitude at time t, shown in Eq. (2), where Δr is the spiral (sampling) pitch normal to the radial dimension:

Eq. (1)

ν(t)=Vs×4πtΔr,

Eq. (2)

r(t)=VsΔrtπ,

Eq. (3)

Vs=dν(t)dtr(t)=ω(t)r(t).

Fig. 1

Description of spiral scanning implementation: (a) sinusoidal voltage waveforms for driving x (black) and y (red) mirrors of the galvanometer; (b) spiral scan pattern generated by sinusoidal voltage waveforms; (c) frequency of sinusoidal voltage as a function of time; and (d) velocity along the spiral as a function of time. Vs, spiral scan velocity; Δθ, spiral pitch in angular dimension θ (along the spiral); Δr, sampling pitch in radial dimension r; and rend, radius of spiral scan circular field of view. Red arrows denote start of the scan or direction of travel. All scan patterns are undersampled for visual clarity.

NPh_11_1_015007_f001.png

We use the term linear velocity Vs (in units of m/s) to mean the velocity along the path of the scanner and differentiate it from angular velocity ω(t) (in units of rad/s). More details on the use and derivation of Eqs. (1)–(3) can be found in Ref. 59. Equation (3) expresses the relationship between Vs and ω(t) of the spiral pattern. Vs is held constant to maintain homogenous spiral pitch or points separation [Fig. 2(d)]; therefore, the frequency of scanning decreases rapidly as we scan outwards [Fig. 2(c)]. Implementing this decrease in frequency is straightforward using galvanometers, because they can scan very fast, as needed for the beginning of the spiral scan, but can be precisely programmed to slow as the scanning extends outwards. Δθ and Δr are the spiral pitch along the angular and radial axes of the spiral scan, respectively, and are expressed as

Eq. (4)

Δθ=VsfA-scan,

Eq. (5)

Δr=πr2(t)tVs.

Fig. 2

Experimental setup and specifications of wide-field sLS-PAM system: (a) schematic of the system, (b) detection sensitivity map obtained by imaging a piece of black tape, (c) lateral resolution obtained by imaging a sharp blade edge, and (d) two orthogonal intensity profiles across the dotted lines shown in (b). M1–M3: silver mirrors; FL; focusing lens; PC, personal computer; FOV, field of view; FWHM, full-width half-maximum; LSF, line spread function; ESF, edge spread function.

NPh_11_1_015007_f002.png

Equation (4) is analogous to sampling pitch along the fast axis of a raster scan59 where fA-scan is the frequency of generating one A-scan. Equation (5) was derived by solving Eq. (2) for Δr. To achieve a uniform isotropic transverse sampling, Δθ was set equal to Δr. A more thorough explanation of spiral scanning can be found in Refs. 58 and 59.

2.2.

System Development and Characterization

Figure 2(a) shows the schematic of the sLS-PAM system. A pulsed Nd:YAG laser (VPGL-G-20, V-gen, Tel Aviv, Israel) was used as an optical excitation source with a wavelength of 532 nm and a pulse repetition rate of up to 700 kHz. Through a combination of three 90 deg reflective mirrors (MPD019-G01, Thorlabs, Newton, New Jersey, United States), the light from the laser was focused onto a 2D-galvanometer (GVS 202-2D, Thorlabs, Newton, New Jersey, United States). The specifications for the galvanometer are given in Table 1. The XY mirrors in the galvanometer were controlled and operated by a LabVIEW program interfaced with a terminal block (BNC 2110, National Instruments), to generate two orthogonal sinusoidal signals [see Fig. 1(a)] to create a spiral scanning pattern [see Fig. 1(b)].

Table 1

Specifications of galvo scanner.

ParameterValue
Optical scan angle range−25 deg to + 25 deg
Mechanical scan angle range−12.5 deg to +12.5 deg
Scan frequency≥1 kHz for mechanical scan angle < 0.2 deg
≤250 Hz for mechanical scan angle > 0.2 deg
Wavelength400 to 700 nm
Input voltage range± 6.25 V
Input beam diameter1 to 5 mm

An output card was used to generate a frame trigger signal to initiate the start and end of data acquisition. The laser irradiation, spiral scanning, and frame trigger were synchronized by a square shaped signal produced by an external function generator (ATF20B, ATTEN instruments, United States). After the galvanometer, the spirally scanned laser beam was focused onto the imaging target through an achromatic doublet lens (AC127-075-A, Thorlabs, Newton, New Jersey, United States) with a focal length of 7.5 cm following its use in Refs. 6162.63. The AC127-075-A doublet lens was carefully chosen to meet the requirements for continuous imaging (minutes) without damage to the live target or the lens. This lens has a high damage threshold which guarantees its endurance during long (few seconds to few minutes) functional imaging sessions. The relatively low NA lens we have used guards against photobleaching while still providing acceptable resolution (<48  μm). Although imaging with a higher NA lens can provide detail on the finest capillaries, in our experience, upstream capillaries (proximate to veins and arteries) provide more consistent results for measuring hemodynamic change, possibly because hemoglobin volume in the smallest capillaries are not as large as in the upstream capillaries. A water tank was constructed with a Saran wrapped bottom side. The imaging target was pushed up against the Saran wrap to avoid any air gaps. To ensure a perfect coupling, ultrasound gel was used between the Saran wrap and the imaging target.

Upon laser irradiation, the induced PA signals were detected by an ultrasound transducer (Centrascan C306, Olympus NDT; central frequency: 2.25 MHz; 6  dB bandwidth 66%). Centrascan and Videoscan transducers from Olympus are widely used for implementing PAM. The Centrascan transducers, with a piezocomposite element, provide excellent sensitivity with a high SNR. The backing layer in these transducers is composed of a highly attenuative, high-density material that controls the vibration of the transducer by absorbing the energy radiating from the back face of the active element. There is a mismatch in acoustic impedance between the element and the backing, leading to more sound energy reflected forward into the imaging target, which results in greater sensitivity. By comparison, the Videoscan transducer is heavily damped due to good acoustic impedance matching with backing material, which provides a broader bandwidth, but lower sensitivity. The largest available Centrascan transducer is 5 MHz. We ran tests on both the 2.25 and 5 MHz transducers to check differences in resolution and signal strength. The tests demonstrated that the 2.25 MHz model provides 20% higher sensitivity than the 5 MHz, with only 5% loss in resolution. This low-frequency transducer selection is also beneficial for PAM imaging through rodent skull.

The transducer was placed in the water tank, 40 mm away from the imaging target at an angle of 30 deg. Different angles of the transducer give different sensitivity distribution maps and affect the depth shift in the PA signal. An empirical study was performed where using a goniometer, the angle and distance of the transducer was iteratively varied until a maximum area of uniform sensitivity was achieved. Maximum area of uniform sensitivity was determined by measuring full-width half-maximum (FWHM) across two perpendicular line profiles of the imaging area following.64 A maximum uniform sensitivity area (9  mm radius) was achieved when the transducer was placed 30 deg from the y axis and 40 mm from the imaging object [see Figs. 2(a) and 2(b)]. Once the transducer position was fixed with these parameters, it was not changed for all the other studies presented in this manuscript.

A variable gain voltage amplifier with a maximum gain of up to 80 dB (Model 351A, Analog Modules Inc., Longwood, Florida, United States) was used to amplify the detected PA signals. The signals were then digitized at a sampling rate of 100  MS/s using a data acquisition system (GaGe CSE1642, Vitrek Corporation, Poway, California, United States). The data were streamed in real time. The angle within which the galvo mirrors moved, the FOV of the transducer, and the focal length of the objective lens determine a 18 mm diameter FOV, inside which the lateral resolution was within 6 dB fall-off. The FOV and sensitivity map were measured by imaging a piece of flat black tape [see Fig. 2(b)]. To characterize the lateral resolution of the system, we imaged a sharp blade (Rexbeti, Auburn, Washington, United States) [see Fig. 2(c)], where the resolution was measured as 48  μm by calculating the FWHM of the Gaussian profile across the green line shown in Fig. 2(c); the coarse lateral resolution is partially due to transducer frequency, which filters out PA signals generated from finer structures. Two orthogonal intensity profiles [dotted lines shown in Fig. 2(b)] are used to characterize the sensitivity map [Fig. 2(d)].

For all subsequent experiments, each frame consisted of 160,000 scanning points, scanned with sampling pitch (Δθ) and spiral pitch (Δr) of 40  μm (beam size on the surface of the imaging target was 16.9  μm). The sinusoidal signals applied to the galvanometer scanners had progressive voltage values and frequency [see Fig. 1(a)], which generated a constant velocity (Vs) of 8  m/s [see Fig. 1(d)]. The segment size per point acquired was 500 samples and the total data set acquired per frame was 200 MB [160,000×500×14  bits (2 bytes)]. For all the animal studies, pulse repetition rate of the laser was set to either 200 kHz (0.8 s acquisition/frame) or 400 kHz (0.4  s/frame) to scan one spiral frame with a diameter of 18  mm.

2.3.

Depth Correction Algorithm

One limitation of an sLS-PAM setup is the noncoaxial arrangement of optical and acoustic focus because the transducer is placed at an angle [Fig. 3(a)]. Due to angled transducer placement, the data acquired are proportionally shifted in the axial direction, leading to some degree of depth misalignment. In our setup, the transducer is placed at an angle of 30 deg from the target, as shown in Fig. 3(a). Due to angled transducer placement, the proximal end of the imaging target is closer than the distal end. Therefore, the PA signals generated from the proximal end of the target reach the ultrasound transducer earlier than the PA signals from the distal end. This introduces a gradual time delay, because the points are scanned from proximal to distal end, in turn affecting axial/depth alignment of the images.

Fig. 3

Schematic demonstrating depth misalignment and correction in sLS-PAM. (a) Angled position of the transducer leading to depth misalignment. (b) Virtually shifted angled transducer from noncoaxial to axial position. Dnc, distance from the noncoaxial transducer to the center of the imaging area. Dc, corrected distance from theoretical coaxial position of the transducer to the center of the imaging area.

NPh_11_1_015007_f003.png

Fig. 4

Flowchart for depth correction algorithm. TR, transducer; P, a single point in the spiral scan; Snc and Sc, sample number computed from the center of angled transducer to point P in spiral scan. Tp, total number of points in spiral scan.

NPh_11_1_015007_f004.png

We developed a depth correction algorithm to virtually shift the transducer from the angled (i.e., noncoaxial) position to a coaxial position [Fig. 3(b)]. In this algorithm, the distance (Dnc) from each point (P) of the spiral scan pattern was calculated from the center of the angled transducer. Similarly, the distance (Dc) from the center of the coaxial position of the transducer was computed. The relationship between Dc and Dnc is Dc=Dnccosφ, where φ is the angle between the two positions. Based on Dnc and Dc, the sample numbers in which we expected to see PA signals were calculated, we call them noncoaxial (Snc) and coaxial samples (Sc), respectively. If Snc and Sc were not equal, data were shifted to match these sample numbers. (The order of the samples was shifted from the order in which they actually reached the target to the order in which they would have reached the target had they been coaxial.) Note that our depth correction algorithm only shifts the data without modifying it. The pixel values of the image before and after image correction were essentially the same (because the depth correction algorithm does not change pixel values). The data correction algorithm is summarized in a flowchart in Fig. 4. Subsequently, to evaluate the performance of the algorithm, a parameter called “depth anomaly,” DA was defined. To calculate DA, an axial line profile was acquired from the proximal to the distal end of the image. The slope of the line profile was then used to calculate DA. The steeper the slope is, the higher DA is.

Fig. 5

Animal protocols for whisker stimulation and electrical stimulation. (a) Method for stimulating whiskers. (b) Protocol for whisker stimulation and image acquisition. (c) Placement of electrodes on the second and forth index finger of rat’s left foot for electrical stimulation (left) and protocol for electrical stimulation (right). WS, whisker stimulation and IAQ, image acquisition.

NPh_11_1_015007_f005.png

2.4.

Lucy–Richardson Deconvolution Algorithm

Another limitation of any PAM setup is that due to the imperfection of the optical components, reaching the theoretical spatial resolution is not possible. Such distortion in resolution appears as blurriness in the PA image. One potential solution to improve image quality is to estimate the point spread function (PSF) of the system and deconvolve it from the PA images. However, this method only reduces the static aberration and not the dynamic aberration introduced by the sample.65 In an imaging system, the distorted output image g(x,y) can be modeled as the convolution of an undistorted input image f(x,y) with a transfer function, or PSF, of the imaging system h(x,y):

Eq. (6)

g(x,y)=f(x,y)*h(x,y).

The Lucy–Richardson algorithm works based on calculating the maximum-likelihood solution for recovering an undistorted image that has been blurred by a known PSF [see Eq. (7)]:

Eq. (7)

fm+1(x,y)=fm(x,y)[h(x,y)*[g(x,y)h(x,y)*fm(x,y)]],
where fm(x,y) is the estimate of the undistorted image in the m’th iteration. The optimization process starts with f0(x,y)=g(x,y) and iteratively modifies fm(x,y). Following our previous work,66 we imaged a very small object (nonfunctionalized colloidal polystyrene microspheres [Alpha Nanotech Inc., United States) of diameter 40  μm] in a gelatin phantom to estimate the PSF of the imaging system. We then utilized the Lucy–Richardson deconvolution algorithm to deconvolve the PSF from the PAM image. For testing the performance of the deconvolution algorithm, a fine branched leaf phantom (Etsy, New York, United States) was imaged.

2.5.

Preparation of Animal Model

Six-week-old male Sprague Dawley rats (Charles River Laboratories, Wilmington, Massachusetts, United States) were used for this study. Because of the aberrating effect of skull, scalp was removed, and the skull was thinned surgically as follows. Rats were first anesthetized using 4% isoflurane inhalation followed by maintenance doses of 2% to 3% isoflurane during surgery (Doctor Oxygen Service, Inc., Franklin, Wisconsin, United States). Hair was removed using a trimmer to expose the scalp. Next, the scalp was removed from the entire dorsal area. Periosteum was also removed by carefully scraping the exposed skull bone using a scalpel blade. The skin was trimmed laterally, and the temporal muscles were gently detached from the bone on both sides of the skull. The fronto-parietal bone was thinned down (2.5  cm2 over the somatosensory cortex) at low speed with a very fine tip dental drill bit (Dremel 8260, Dremel, New York, New York, United States). The remaining thickness of the bone was 0.5  mm from the brain surface. To prevent overheating of the skull, saline was applied repeatedly between the drilling sessions until the skull became flexible, and the major vessels became visible through the wet bone. The animal was introduced to the imaging setup and was kept there for 10 min prior to imaging, to assimilate to the conditions. Ultrasound gel was applied on the exposed region of the skull for acoustic coupling. The experiments were accomplished while the animals were under anesthesia, and the animals were euthanized upon completion of the study. All procedures were conducted in accordance with the “Guide for the Care and Use of Laboratory Animals” and approved by the Wayne State University and University of Illinois Institutional Animal Care and Use Committees.

2.6.

Animal Protocols for Whisker Stimulation, Electrical Stimulation, and Hemorrhage Induction

We tested our system using whisker and electrical stimulations, some of the most used techniques for studying cerebral hemodynamic activities in animal models.5,33,67,68 The whisker stimulation protocol, shown in Fig. 5(a), was as follows: the whiskers from one side of the nose were deflected from the whisker pads by means of a tweezer for 20 s, followed by image acquisition, taking 0.8 s. To obtain the strongest functional signals correlated with whisker stimulation, all whiskers on one side of the nose were taped together and deflected simultaneously from the whisker pad. The rat was then rested for 5 min before repeating the next whisker stimulation, to allow hemodynamic changes to subside and return to baseline levels, thereby not interfering with the subsequent image acquisition. The experiment was repeated for four trials. In the electrical stimulation protocol, a pair of metallic needle electrodes (JM needles, Japan) were inserted to the second and fourth finger of the animal’s left hind paw [see Fig. 5(c)]. A 3 mA pulse with 0.5 mA pulse width and 5 Hz frequency was generated from a function generator (ATF20B, ATTEN Instruments, Columbus, Ohio, United States) to provide electrical stimulation. The stimulation protocol was 5 s on followed by an extended resting period [see Fig. 5(b)]. The image acquisition protocol included continuous imaging from 10 s before stimulation until signal returned to baseline (75  s). We also mimicked a brain injury model in rats. For this experiment, the entire skull was removed. Subsequently, we inflicted a small incision in one of the brain vessels using a 1 mm pointed needle (91-01-01, Master Appliance Corp., Racine, Wisconsin, United States).

3.

Results

We compared the performance of our spiral scanning scheme with unidirectional and bidirectional scanning methods. For the spiral pattern to exhibit the same parameters as the unidirectional and bidirectional scans, the velocity of spiral scanning (Vs) was kept equal to Vtri=Vsaw. This consequently achieved Δθ=Δx (separation between scanning points) and, Δr=Δy (line separation or spiral pitch). The length of the square imaging area for the triangular and sawtooth scanning patterns L was adjusted to cover the same total area as the spiral pattern with radius r (eg L2=πrend2), where rend is the radius of the circular imaging area produced by the spiral scan shown in Fig. 1(b), and L is the raster scan length.

We imaged a cross, composed of two pieces of black tape, at different frame rates (length: 16 mm). For all three scanning mechanisms, we kept the overall number of scanning points equal (i.e., 160,000 points) to maintain the spiral pitch of 40  μm. With unidirectional scanning, we imaged a frame in 3.2, 1.6, and 0.8 s, as shown in Figs. 6(a)(i)6(a)(iii). Due to the unidirectional scanning mechanism, a reverse directional image artifact (due to flyback time delay) was introduced, which became more prominent as we scanned at a faster frame rate [see Fig. 6(a)(ii) and 6(a)(iii)]. In contrast, with bidirectional scanning, the problem of reverse directional image artifact was solved [see Fig. 6(b)(i) and 6(b)(ii)]; however, with faster frame rates, the problem of scan misalignment (due to inertial time delays or phase lag delays) became evident, as shown in Fig. 6(b)(iii). With spiral scanning, we were able to scan as fast as 0.4 s [see Fig. 6(c)(iv)] but some distortions including a small blank area were evident.

Fig. 6

Comparison between the performance of (a) unidirectional, (b) bidirectional, and (c) spiral scans (FOV was square (16  mm×16  mm) for unidirectional and bidirectional scans and circular (diameter 18 mm) for spiral scans in LS-PAM when a “X” shaped phantom made of black tape, was imaged at scanning rates of (i) 3.2  s/frame, (ii) 1.6  s/frame, (iii) 0.8  s/frame, and (iv) 0.4  s/frame.

NPh_11_1_015007_f006.png

The distortions that we see at the center of the image are a drawback of the constant velocity spiral scanning method used in this study. Such distortions are hardware-dependent [concerning, for example, the specifications of the galvo mirrors and the data acquisition system (DAQ)]. We refer readers to Ref. 58, which gives mathematical and visual details of distortions associated with spiral scanning. The severity of these distortions is proportional to frame rate of image generation, as seen in Fig. 6(c)(i)6(iv). For our implementation of constant velocity spiral scanning, we have observed that in the center, the rate of trigger is too fast for the DAQ system, and thus data acquisition does not take place. This malfunction generates zero intensity data, which appears as blank regions in the center of the image [Fig. 6(c)(iii) and 6(c)(iv)]. In addition, at these high frequencies, the galvo mirrors lock themselves and the light beam is not directed toward the target (within the FOV). For instance, if the frame rate was 0.8  s/frame, the initial frequency would have been 20 kHz, and the galvo would be locked until the 189th point’s frequency (that is 1  kHz), creating a blank circle of diameter 70  μm. If the frame rate was 0.4  s/frame, the initial frequency would have been 40 kHz, and the galvo would be locked until the 388th point’s frequency (that is 1  kHz), creating a blank circle of diameter 200  μm. Once the DAQ is triggered, if the frequency is still too high for the galvo, the galvo cannot follow accurately, and a distortion similar to the one shown in Fig. 5 of Ref. 58 will also appear near the center of the image. The diameter of the blank region artifact and the nearby distortion decreases with decreasing frame rate. This artifact in many applications, if it is small, can be ignored.

To evaluate the depth correction algorithm, we prepared a test phantom consisting of three copper wires (diameter 0.5 mm) at different depths as shown in Fig. 7(a). Next, Fig. 7(b) shows the photoacoustic image of this phantom before [Fig. 7(b)(i)] and after [Fig. 7(b)(ii)] depth correction. The averaged depth anomaly, the slope of line profile in the axial direction, was improved from 73% to 9%. The depth anomaly was not improved to zero due to manual error in transducer angle measurements. This algorithm was incorporated into all future results.

Fig. 7

Depth correction algorithm results. (a) Experimental setup to evaluate the depth correction algorithm by imaging a three-wire phantom in water with an ultrasound transducer at an inclined angle of 30 deg. The setup includes: (i) water tank, (ii) ultrasound transducer, (iii) objective lens, (iv) xy galvo scanner, (v) two-stage low noise amplifier. (b) Photoacoustic images: (i) PAM image of the phantom without depth correction and (ii) PAM image of the phantom with depth correction. DA, depth anomaly.

NPh_11_1_015007_f007.png

To obtain the system PSF, we imaged nonfunctionalized colloidal polystyrene microspheres (Alpha Nanotech Inc., Vancouver, Washington, United States), of diameter 40  μm, embedded in 8% gelatin in a water phantom as shown in Fig. 8(a). A brightfield image of two microspheres is shown in Fig. 8(b) (we used a 40X-2500X Kohler compound microscope) and the PAM image of microspheres in the gelatin phantom is shown in Fig. 8(c). From the PAM image, we selected a window of 20×20  pixels, enclosing the image of a single microsphere [shown in red box in Fig. 8(c)], representing the PSF of the system. We imaged a leaf phantom as shown in Fig. 8(d) and applied the Lucy–Richardson deconvolution algorithm to correct for image aberrations. The images before and after deconvolution are given in Figs. 8(e) and 8(f), respectively. Line profiles across the line indicated in Figs. 8(e) (in green) and 8(f) (in blue) were compared. The results showed a considerable reduction in the image blurriness after applying the deconvolution algorithm. Therefore, this deconvolution algorithm was incorporated into all future results.

Fig. 8

Deblurring algorithm results. (a) A photograph of polystyrene microspheres embedded in gelatin phantom, (b) brightfield photograph of polystyrene microspheres, (c) PAM image of microspheres in gelatin phantom, (d) photograph of a leaf phantom, (e) PAM image before deconvolution, (f) PAM image after deconvolution, and (g) averaged intensity line profiles obtained from the images in (d) (green) and (e) (blue). LR, Lucy–Richardson.

NPh_11_1_015007_f008.png

To evaluate the functional brain imaging capability of our sLS-PAM, we imaged a rat brain in vivo during intermittent rest and stimulation periods. All of the following experiments were performed by spiral scans with an imaging time of 0.8 s per frame for an 18 mm diameter area. Whiskers and electrical stimulation were performed. The cycle of stimulation and rest periods was repeated 4 times (4 trials). Using the PAM images of the rat’s brain vasculature at rest [See Fig. 9(a)] and whisker stimulation, stimulation maps were generated by subtracting each trial’s PA image after stimulation from each trial’s PA image at rest. The four stimulation maps are shown in Figs. 9(b)9(e). By stimulating the rat’s whiskers, we anticipated a functional change in the somatosensory cortex69 (because of the light illumination used in PAI, the visual cortex may have been stimulated as well). Because left nose whiskers were stimulated and nerves cross upon entering the brain (decussation), we expected to see functional changes in the right side of the rat’s brain. Although there are other areas showing increase (or decrease) in activity, such changes might be due to functional connectivity between different subsystems. In summary, the changes in the averaged PA signal in the activated brain regions are consistent with the stimulation activity described in the protocol. For quantitative purposes, we selected 10 regions of interest containing vasculature and performed pixel intensity and vessel diameter analyses before and after whisker stimulation. These results are shown in Figs. 9(f) and 9(g), respectively.

Fig. 9

Evaluation of the functional brain imaging capability of sLS-PAM with whisker stimulation. (a) Image of rat’s brain before whisker stimulation. (b)–(e) Activation maps of four separate trials showing an increase in pixel intensity mainly on the right side of the rat brain (left whiskers were stimulated). Maps were developed by subtracting each trial’s PA image after stimulation from each trial’s PA image at rest, graphs showing (f) change in pixel intensity and (g) vessel diameter before and after whisker stimulation for ten separate regions of interest identified on the images in (a).

NPh_11_1_015007_f009.png

Electrical stimulation was also performed, and images were acquired continuously (see Fig. 5). A PAM image of the whole rat brain is shown in Fig. 10(a). Figures 10(b)10(g) shows an enlarged version of the inset box in Fig. 10(a), at a range of times. We note increased pixel intensity (over baseline) at the onset of electrical stimulation at t=0 Fig. 10(d). At t=16, we observed some vessels in the enlarged regions become activated, denoted by a further increase in pixel intensities, and the effect of stimulation became more pronounced at t=26  s. The effect of electrical stimulation persisted for about 26 s before starting to diminish. Finally, at 75 s, we see the same intensity as prior to stimulation [Fig. 10(g)]. Figure 10(h) shows the change in the hemodynamics in terms of PA signal intensity from 10 s before to 75 s after stimulation, and Fig. 10(i) shows change in vessel diameter for a single vessel [shown in Fig. 10(b)]. The experiment was repeated 10 times and the rat was rested for 5 min between experiments to allow hemodynamic changes to subside and return to baseline levels. The extent of electrical stimulation was fairly strong (500  μs pulses of amplitude 3 mA delivered in 5 s) compared with a recently reported rodent brain PA imaging system70 (250  μs of amplitude 2 mA delivered over 20 s). The rationale for using the more intense stimulation for this study was to demonstrate the capability of the system to capture a prolonged animal response with high time resolution, which we were not able to accomplish with a simpler whisker stimulation protocol, and gentler electrical stimulation.

Fig. 10

sLS PAM images of rat brain with electrical stimulation: (a) sLS-PAM image at t=0 (beginning of electrical stimulation); (b)–(g) enlarged area of sLS-PAM image denoted by yellow dotted inset, at (b) t=0  s (resting state), (c) t=8  s, (d) t=16  s, (e) t=26  s, (f) t=52  s, and (f) t=75  s; (h) percent change in average pixel intensity over area shown in (b)–(g), over time; and (i) percent change in vascular diameter over time [vessel identified by dashed circle in (b)].

NPh_11_1_015007_f010.png

We also mimicked a brain injury model in a rat by vascular incision. For this experiment only, the rat skull was entirely removed. The rat brain before and after vascular incision is shown in Figs. 11(a)(i) and 11(a)(ii), respectively. The sLS-PAM images before [Fig. 11(b)(i)] and after vascular injury [Fig. 11(b)(ii)] were acquired and the injured area was clearly distinguishable, as marked by a white pointer. Furthermore, blood leakage around the injured area was observed, specified by a purple pointer.

Fig. 11

Brain injury model. (a) (i) Rat brain prior to vascular incision and (ii) rat brain after vascular incision. (b) (i) sLS-PAM image before vascular incision and (ii) sLS-PAM image after vascular incision. White pointer detonates injured area and purple pointer denotes area with blood leakage.

NPh_11_1_015007_f011.png

4.

Discussion

Neuroimaging techniques play a crucial role in exploring functional brain activity by measuring cerebral hemodynamic parameters and enhancing our fundamental understanding of neuroscience. Rats have become extensively utilized in neuroscience research due to the availability of numerous transgenic rat models tailored for various studies. In our previous research, we investigated brain activities by employing ring array PACT to image the primary functional cortical regions. As a result, we were inspired to conduct the current study with the goal of extending this imaging capability to visualize finer regions, i.e., the cortical subregions within the main functional regions. We developed a system which we call sLS-PAM, with a sufficiently high resolution, fast frame rate (>1  Hz), and wide FOV (18  mm) for visualizing the vasculature and hemodynamics in an entire rat brain.71 Our system can work continuously we have tested is use imaging phantoms for up to 5 min and the optics were intact and showed no damage. Moreover, it appears reasonably safe for animals, as we saw no sign of damage to the rat’s brain, even after 75 s of continuous imaging (and 10 consecutive imaging session each lasting over 75 s). Although we used a 2.25 MHz Centrascan transducer, a higher frequency transducer with the same or higher sensitivity would improve the lateral resolution of the current system, and we are working with a company to develop a customized transducer for this purpose. Although we do not claim our system is the fastest reported in the literature, it is fast enough for rat brain functional imaging applications. To demonstrate its place among other designs, we list in Table 2 some existing OR-PAM systems with their imaging area, temporal resolution, and temporal resolution per imaging area, compared to our system (written in bold).

Table 2

Comparison between OR-PAM systems. Characteristics of our system are written in bold.

Ref.MethodImaging area (mm2)Temporal resolution (s)Temporal resolution per imaging area (s mm−2)
47Water immersible MEMS scanning PAM2 × 51.250.13
49Fully motorized scanning PAM8 × 101802.25
50Waterproof 2D galvanometer scanning PAM8 ×131000.96
51Cylindrical focused wide field PAM4.5 × 114.5
52Wide field polygon scanning PAM12 × 1250.03
72Semi-water immersible galvanometer scanning PAM12.9 × 8 mm150.15
73LS-PAM with ZnO detector2 × 21.70.43
74Fiber optics-based PAM8 × 13320.31
75LS-PAM with transparent detector20 × 201800.45
61Water immersible 12 facet polygon scanning PAM11 × 7.50.5 (unprocessed)0.006
This reportsLS-PAMDiam = 18 cm (circular)0.80.003

Some of the limitations of sLS-PAM are as follows. Because of the inclined angle of the transducer, the depth-resolved information needs correction. Another shortcoming is that the angle and distance between the transducer and the imaging target needs optimization for each application to ensure adequate sensitivity and FOV. These concerns may be resolved by motorizing the transducer holder. Finally, at very fast imaging speeds, the current implementation of the spiral scanning system has distortion and a blank area artifact in the center of the image, due to high angular frequency at scan inception. This can be overcome by developing a distortion correction algorithm.

5.

Conclusion

We developed a high-resolution, fast frame rate, and wide FOV OR-PAM spiral scanning system. We do not claim the system has the fastest frame rate, highest resolution, largest FOV, and highest sensitivity altogether, but we demonstrated it is capable of imaging a circular region with 18 mm diameter in 0.4 s with negligible artifact (i.e., a small blank circle in the center of the image) and some distortion, or in 0.8 s virtually artifact free with negligible distortion, which is ample for the purpose of imaging vasculature and hemodynamics in a rat brain. We demonstrated the functional imaging capability of the sLS-PAM system by imaging cerebral hemodynamics in response to whisker and electrical stimulation and used it for vascular imaging of a modeled brain injury. We believe that there are many neuroscience questions that can be answered by the described system.

Disclosures

Authors declared no conflicts of interest.

Code and Data Availability

Code and Data can be made available upon reasonable request.

Funding

This work was supported by the National Institutes of Health (Grant Nos. R01EB027769-01 and R01EB028661-01).

Acknowledgments

We would like to thank Art Camire for his help regarding the reconfiguration of the V-gen laser.

References

1. 

M. Zafar et al., “Development of low-cost fast photoacoustic computed tomography: system characterization and phantom study,” Appl. Sci., 9 (3), 374 https://doi.org/10.3390/app9030374 (2019). Google Scholar

2. 

A. Fatima et al., “Review of cost reduction methods in photoacoustic computed tomography,” Photoacoustics, 15 100137 https://doi.org/10.1016/j.pacs.2019.100137 (2019). Google Scholar

3. 

D. Nasri et al., “Photoacoustic imaging for investigating tumor hypoxia: a strategic assessment,” Theranostics, 13 (10), 3346 –3367 https://doi.org/10.7150/thno.84253 (2023). Google Scholar

4. 

X. Wang et al., “Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain,” Nat. Biotechnol., 21 (7), 803 –806 https://doi.org/10.1038/nbt839 NABIF9 1087-0156 (2003). Google Scholar

5. 

S. Hu and L. V. Wang, “Neurovascular photoacoustic tomography,” Front. Neuroenerg., 2 10 https://doi.org/10.3389/fnene.2010.00010 (2010). Google Scholar

6. 

J. G. Laufer et al., “Photoacoustic imaging of vascular networks in transgenic mice,” Proc. SPIE, 7564 75641A https://doi.org/10.1117/12.842204 PSISDG 0277-786X (2010). Google Scholar

7. 

J. Laufer et al., “Three-dimensional noninvasive imaging of the vasculature in the mouse brain using a high resolution photoacoustic scanner,” Appl. Opt., 48 (10), D299 –D306 https://doi.org/10.1364/AO.48.00D299 APOPAI 0003-6935 (2009). Google Scholar

8. 

J. Levi et al., “Molecular photoacoustic imaging of follicular thyroid carcinoma,” Clin. Cancer Res., 19 (6), 1494 –1502 https://doi.org/10.1158/1078-0432.CCR-12-3061 (2013). Google Scholar

9. 

R. Manwar et al., “An application of simulated annealing in compensation of nonlinearity of scanners,” Appl. Sci., 9 (8), 1655 https://doi.org/10.3390/app9081655 (2019). Google Scholar

10. 

M. Zafar, “Instrumentation of fast widefield photoacoustic microscopy system with high fidelity data enhancement and storage,” (2022). Google Scholar

11. 

K. S. Valluru, K. E. Wilson and J. K. Willmann, “Photoacoustic imaging in oncology: translational preclinical and early clinical experience,” Radiology, 280 (2), 332 –349 https://doi.org/10.1148/radiol.16151414 RADLAX 0033-8419 (2016). Google Scholar

12. 

A. Dima and V. Ntziachristos, “Non-invasive carotid imaging using optoacoustic tomography,” Opt. Express, 20 (22), 25044 –25057 https://doi.org/10.1364/OE.20.025044 OPEXFF 1094-4087 (2012). Google Scholar

13. 

E. Fakhrejahani et al., “Clinical report on the first prototype of a photoacoustic tomography system with dual illumination for breast cancer imaging,” PloS One, 10 (10), e0139113 https://doi.org/10.1371/journal.pone.0139113 POLNCL 1932-6203 (2015). Google Scholar

14. 

S. A. Ermilov et al., “Laser optoacoustic imaging system for detection of breast cancer,” J. Biomed. Opt., 14 (2), 024007 https://doi.org/10.1117/1.3086616 JBOPFO 1083-3668 (2009). Google Scholar

15. 

G. C. Langhout et al., “Detection of melanoma metastases in resected human lymph nodes by noninvasive multispectral photoacoustic imaging,” Int. J. Biomed. Imaging, 2014 163652 https://doi.org/10.1155/2014/163652 (2014). Google Scholar

16. 

R. Manwar et al., “Development and characterization of transfontanelle photoacoustic imaging system for detection of intracranial hemorrhages and measurement of brain oxygenation: ex-vivo,” Photoacoustics, 32 100538 https://doi.org/10.1016/j.pacs.2023.100538 (2023). Google Scholar

17. 

R. Manwar et al., “Development of ex-vivo brain hemorrhage phantom for photoacoustic imaging,” J. Biophotonics, 16 (7), e202200313 https://doi.org/10.1002/jbio.202200313 (2023). Google Scholar

18. 

R. Manwar et al., “Transfontanelle photoacoustic imaging for in-vivo cerebral oxygenation measurement,” Sci. Rep., 12 (1), 15394 https://doi.org/10.1038/s41598-022-19350-x SRCEC3 2045-2322 (2022). Google Scholar

19. 

J. Benavides-Lara et al., “Transfontanelle photoacoustic imaging of intraventricular brain hemorrhages in live sheep,” Photoacoustics, 33 100549 https://doi.org/10.1016/j.pacs.2023.100549 (2023). Google Scholar

20. 

R. Manwar et al., “Transfontanelle photoacoustic imaging: ultrasound transducer selection analysis,” Biomed. Opt. Express, 13 (2), 676 –693 https://doi.org/10.1364/BOE.446087 BOEICL 2156-7085 (2022). Google Scholar

21. 

S. Mahmoodkalayeh et al., “Wavelength and pulse energy optimization for detecting hypoxia in photoacoustic imaging of the neonatal brain: a simulation study,” Biomed. Opt. Express, 12 (12), 7458 –7477 https://doi.org/10.1364/BOE.439147 BOEICL 2156-7085 (2021). Google Scholar

22. 

M. Nasiriavanaki et al., “High-resolution photoacoustic tomography of resting-state functional connectivity in the mouse brain,” Proc. Natl. Acad. Sci. U. S. A., 111 (1), 21 –26 https://doi.org/10.1073/pnas.1311868111 (2014). Google Scholar

23. 

Y. Zhou, J. Yao and L.V. Wang, “Tutorial on photoacoustic tomography,” J. Biomed. Opt., 21 (6), 061007 https://doi.org/10.1117/1.JBO.21.6.061007 JBOPFO 1083-3668 (2016). Google Scholar

24. 

M. Zafar et al., “Photoacoustic signal enhancement using a novel adaptive filtering algorithm,” Proc. SPIE, 10878 108785S https://doi.org/10.1117/12.2510557 PSISDG 0277-786X (2019). Google Scholar

25. 

H. Salehi et al., “Design of optimal light delivery system for co-registered transvaginal ultrasound and photoacoustic imaging of ovarian tissue,” Photoacoustics, 3 114 –122 https://doi.org/10.1016/j.pacs.2015.08.003 (2015). Google Scholar

26. 

M. Xu and L. V. Wang, “Photoacoustic imaging in biomedicine,” Rev. Sci. Instrum., 77 (4), 041101 https://doi.org/10.1063/1.2195024 RSINAK 0034-6748 (2006). Google Scholar

27. 

L. V. Wang and L. Gao, “Photoacoustic microscopy and computed tomography: from bench to bedside,” Annu. Rev. Biomed. Eng., 16 155 –185 https://doi.org/10.1146/annurev-bioeng-071813-104553 ARBEF7 1523-9829 (2014). Google Scholar

28. 

L. V. Wang and S. Hu, “Photoacoustic tomography: in vivo imaging from organelles to organs,” Science, 335 (6075), 1458 –1462 https://doi.org/10.1126/science.1216210 SCIEAS 0036-8075 (2012). Google Scholar

29. 

J. I. Matchynski et al., “Photoacoustic imaging of fear conditioning neuronal ensembles using a Fos-LacZ gene reporter system in the rat brain in vivo after intrathecal X-gal injection,” Proc. SPIE, 11629 116292E https://doi.org/10.1117/12.2579526 PSISDG 0277-786X (2021). Google Scholar

30. 

J. Yao et al., “Noninvasive photoacoustic computed tomography of mouse brain metabolism in vivo,” NeuroImage, 64 257 –266 https://doi.org/10.1016/j.neuroimage.2012.08.054 NEIMEF 1053-8119 (2013). Google Scholar

31. 

A. Hariri et al., “Development of low-cost photoacoustic imaging systems using very low-energy pulsed laser diodes,” J. Biomed. Opt., 22 (7), 075001 https://doi.org/10.1117/1.JBO.22.7.075001 JBOPFO 1083-3668 (2017). Google Scholar

32. 

J. Kang et al., “Transcranial recording of electrophysiological neural activity in the rodent brain in vivo using functional photoacoustic imaging of near-infrared voltage-sensitive dye,” Front. Neurosci., 13 579 https://doi.org/10.3389/fnins.2019.00579 1662-453X (2019). Google Scholar

33. 

L. D. Liao et al., “Imaging brain hemodynamic changes during rat forepaw electrical stimulation using functional photoacoustic microscopy,” Neuroimage, 52 (2), 562 –570 https://doi.org/10.1016/j.neuroimage.2010.03.065 NEIMEF 1053-8119 (2010). Google Scholar

34. 

J. Janggun and X. Yang, “Functional photoacoustic imaging to observe regional brain activation induced by cocaine hydrochloride,” J. Biomed. Opt., 16 (9), 090506 https://doi.org/10.1117/1.3626576 JBOPFO 1083-3668 (2011). Google Scholar

35. 

R. Manwar, M. Zafar and Q. Xu, “Signal and image processing in biomedical photoacoustic imaging: a review,” Optics, 2 1 –24 https://doi.org/10.3390/opt2010001 (2021). Google Scholar

36. 

C. Li et al., “Real-time photoacoustic tomography of cortical hemodynamics in small animals,” J. Biomed. Opt., 15 (1), 010509 https://doi.org/10.1117/1.3302807 JBOPFO 1083-3668 (2010). Google Scholar

37. 

L. Li et al., “Label-free photoacoustic tomography of whole mouse brain structures ex vivo,” Neurophotonics, 3 (3), 035001 https://doi.org/10.1117/1.NPh.3.3.035001 (2016). Google Scholar

38. 

P. Omidi et al., “A novel dictionary-based image reconstruction for photoacoustic computed tomography,” Appl. Sci., 8 (9), 1570 https://doi.org/10.3390/app8091570 (2018). Google Scholar

39. 

K. Kratkiewicz et al., “Development of a stationary 3D photoacoustic imaging system using sparse single-element transducers: phantom study,” Appl. Sci., 9 (21), 4505 https://doi.org/10.3390/app9214505 (2019). Google Scholar

40. 

X. Bai et al., “Photoacoustic computed tomography with lens-free focused fiber-laser ultrasound sensor,” Biomed. Opt. Express, 10 (5), 2504 –2512 https://doi.org/10.1364/BOE.10.002504 BOEICL 2156-7085 (2019). Google Scholar

41. 

M. Zafar, R. Manwar and K. Avanaki, “High-fidelity compression for high-throughput photoacoustic microscopy systems,” J. Biophotonics, 15 (5), e202100350 https://doi.org/10.1002/jbio.202100350 (2022). Google Scholar

42. 

M. Zafar et al., “Ultra-widefield and high-speed spiral laser scanning OR-PAM: system development and characterization,” J. Biophotonics, 16 (7), e202200383 https://doi.org/10.1002/jbio.202200383 (2023). Google Scholar

43. 

W. Zhang et al., “Miniaturized photoacoustic probe for in vivo imaging of subcutaneous microvessels within human skin,” Quant. Imaging Med. Surg., 9 (5), 807 –807 https://doi.org/10.21037/qims.2019.05.07 (2019). Google Scholar

44. 

W. Zhang et al., “High-speed dual-view photoacoustic imaging pen,” Opt. Lett., 45 (7), 1599 –1602 https://doi.org/10.1364/OL.388863 OPLEDP 0146-9592 (2020). Google Scholar

45. 

H. Guo et al., “Detachable head-mounted photoacoustic microscope in freely moving mice,” Opt. Lett., 46 (24), 6055 –6058 https://doi.org/10.1364/OL.444226 OPLEDP 0146-9592 (2021). Google Scholar

46. 

Z. Xie et al., “Laser-scanning optical-resolution photoacoustic microscopy,” Opt. Lett., 34 (12), 1771 –1773 https://doi.org/10.1364/OL.34.001771 OPLEDP 0146-9592 (2009). Google Scholar

47. 

J. Yao et al., “Water-immersible MEMS scanning mirror enhanced optical-resolution photoacoustic microscopy,” in Biomed. Opt. and 3-D Imaging, (2012). https://doi.org/10.1364/BIOMED.2012.BSu3A.54 Google Scholar

48. 

B. Lan et al., “High-speed widefield photoacoustic microscopy of small-animal hemodynamics,” Biomed. Opt. Express, 9 (10), 4689 –4701 https://doi.org/10.1364/BOE.9.004689 BOEICL 2156-7085 (2018). Google Scholar

49. 

L. Li et al., “Fully motorized optical-resolution photoacoustic microscopy,” Opt. Lett., 39 (7), 2117 –2120 https://doi.org/10.1364/OL.39.002117 OPLEDP 0146-9592 (2014). Google Scholar

50. 

J. Lee et al., “Fully waterproof two-axis galvanometer scanner for enhanced wide-field optical-resolution photoacoustic microscopy,” Opt. Lett., 45 (4), 865 –868 https://doi.org/10.1364/OL.380032 OPLEDP 0146-9592 (2020). Google Scholar

51. 

F. Zhong et al., “High-speed wide-field multi-parametric photoacoustic microscopy,” Opt. Lett., 45 (10), 2756 –2759 https://doi.org/10.1364/OL.391824 OPLEDP 0146-9592 (2020). Google Scholar

52. 

J. Chen et al., “Wide-field polygon-scanning photoacoustic microscopy of oxygen saturation at 1-MHz A-line rate,” Photoacoustics, 20 100195 https://doi.org/10.1016/j.pacs.2020.100195 (2020). Google Scholar

53. 

T. J. Allen et al., “Large area laser scanning optical resolution photoacoustic microscopy using a fibre optic sensor,” Biomed. Opt. Express, 9 (2), 650 –660 https://doi.org/10.1364/BOE.9.000650 BOEICL 2156-7085 (2018). Google Scholar

54. 

Q. Li et al., “MEMS mirror based omnidirectional scanning for LiDAR optical systems,” Opt. Lasers Eng., 158 107178 https://doi.org/10.1016/j.optlaseng.2022.107178 (2022). Google Scholar

55. 

V.-F. Duma et al., “Experimental investigations of the scanning functions of galvanometer-based scanners with applications in OCT,” Appl. Opt., 50 (29), 5735 –5749 https://doi.org/10.1364/AO.50.005735 APOPAI 0003-6935 (2011). Google Scholar

56. 

V.-F. Duma et al., “Optimization of galvanometer scanning for optical coherence tomography,” Appl. Opt., 54 (17), 5495 –5507 https://doi.org/10.1364/AO.54.005495 APOPAI 0003-6935 (2015). Google Scholar

57. 

Y. R. Teo, Y. Yong and A. J. Fleming, “A comparison of scanning methods and the vertical control implications for scanning probe microscopy,” Asian J. Control, 20 (4), 1352 –1366 https://doi.org/10.1002/asjc.1422 (2018). Google Scholar

58. 

X. Sang et al., “Dynamic scan control in STEM: spiral scans,” Adv. Struct. Chem. Imaging, 2 (1), 6 https://doi.org/10.1186/s40679-016-0020-3 (2016). Google Scholar

59. 

O. M. Carrasco-Zevallos et al., “Constant linear velocity spiral scanning for near video rate 4D OCT ophthalmic and surgical imaging with isotropic transverse sampling,” Biomed. Opt. Express, 9 (10), 5052 –5070 https://doi.org/10.1364/BOE.9.005052 BOEICL 2156-7085 (2018). Google Scholar

60. 

M. Nasiriavanaki, “Resting-state functional connectivity measurement in the mouse brain using a low cost photoacoustic computed tomography,” in Front. Opt., (2016). https://doi.org/10.1364/FIO.2016.JW4A.62 Google Scholar

61. 

X. Zhu et al., “Real-time whole-brain imaging of hemodynamics and oxygenation at micro-vessel resolution with ultrafast wide-field photoacoustic microscopy,” Light Sci. Appl., 11 (1), 138 https://doi.org/10.1038/s41377-022-00836-2 (2022). Google Scholar

62. 

M. Chen et al., “High-speed functional photoacoustic microscopy using a water-immersible two-axis torsion-bending scanner,” Photoacoustics, 24 100309 https://doi.org/10.1016/j.pacs.2021.100309 (2021). Google Scholar

63. 

M. Donnachie et al., “A photoacoustic microscopy system using MEMS and fibre tip transducers for all-optical control,” Proc. SPIE, 11960 1196018 https://doi.org/10.1117/12.2609765 PSISDG 0277-786X (2022). Google Scholar

64. 

M. L. Rynes et al., “Miniaturized head-mounted microscope for whole-cortex mesoscale imaging in freely behaving mice,” Nat. Methods, 18 (4), 417 –425 https://doi.org/10.1038/s41592-021-01104-8 1548-7091 (2021). Google Scholar

65. 

M. Singh, U. S. Tiwary and K. Young-Hoon, “An adaptively accelerated lucy-richardson method for image deblurring,” EURASIP J. Adv. Signal Process., 2008 1 –10 https://doi.org/10.1155/2008/674038 (2008). Google Scholar

66. 

S. A. Hojjatoleslami, M. R. N. Avanaki and A. G. Podoleanu, “Image quality improvement in optical coherence tomography using Lucy–Richardson deconvolution algorithm,” Appl. Opt., 52 (23), 5663 –5670 https://doi.org/10.1364/AO.52.005663 APOPAI 0003-6935 (2013). Google Scholar

67. 

V. Tsytsarev et al., “Photoacoustic microscopy of microvascular responses to cortical electrical stimulation,” J. Biomed. Opt., 16 (7), 076002 https://doi.org/10.1117/1.3594785 JBOPFO 1083-3668 (2011). Google Scholar

68. 

K. -W Chang, “Photoacoustic imaging of squirrel monkey cortical and subcortical brain regions during peripheral electrical stimulation,” Photoacoustics, 25 100326 https://doi.org/10.1016/j.pacs.2021.100326 (2022). Google Scholar

69. 

J. L. Walker et al., “Precision rodent whisker stimulator with integrated servo-locked control and displacement measurement,” J. Neurosci. Methods, 196 (1), 20 –30 https://doi.org/10.1016/j.jneumeth.2010.12.008 JNMEDT 0165-0270 (2011). Google Scholar

70. 

S. Na, Y. Zhang and L. V. Wang, “Cross-ray ultrasound tomography and photoacoustic tomography of cerebral hemodynamics in rodents,” Adv. Sci., 9 (25), 2201104 https://doi.org/10.1002/advs.202201104 (2022). Google Scholar

71. 

M. N. Kim et al., “Continuous optical monitoring of cerebral hemodynamics during head-of-bed manipulation in brain-injured adults,” Neurocrit. Care, 20 (3), 443 –453 https://doi.org/10.1007/s12028-013-9849-7 (2014). Google Scholar

72. 

J. Kim et al., “Super-resolution localization photoacoustic microscopy using intrinsic red blood cells as contrast absorbers,” Light Sci. Appl., 8 (1), 103 https://doi.org/10.1038/s41377-019-0220-4 (2019). Google Scholar

73. 

K. Ishikawa et al., “Optical resolution photoacoustic microscopy with fast laser scanning and fixed photoacoustic detector*,” in 40th Annu. Int. Conf. of the IEEE Eng. in Med. and Biol. Soc. (EMBC), (2018). https://doi.org/10.1109/EMBC.2018.8513104 Google Scholar

74. 

T. J. Allen et al., “Ultrafast laser-scanning optical resolution photoacoustic microscopy at up to 2 million A-lines per second,” J. Biomed. Opt., 23 (12), 126502 https://doi.org/10.1117/1.JBO.23.12.126502 JBOPFO 1083-3668 (2018). Google Scholar

75. 

T. Liao et al., “Centimeter-scale wide-field-of-view laser-scanning photoacoustic microscopy for subcutaneous microvasculature in vivo,” Biomed. Opt. Express, 12 (5), 2996 –3007 https://doi.org/10.1364/BOE.426366 BOEICL 2156-7085 (2021). Google Scholar

Biography

Mohsin Zafar is a postdoctoral research associate in the Department of Biomedical Engineering at the University of Illinois Chicago. He specializes in the development of advanced photoacoustic microscopy systems with heightened diagnostic capabilities. His research is dedicated to a comprehensive study of cerebral hemodynamics in small animal models, contributing to a deeper understanding of brain functional dynamics.

Laura Stone McGuire is a vascular and endovascular fellowship-trained neurosurgeon currently practicing at the University of Illinois Chicago. Her research interests include the study of cerebral blood flow in a variety of neurosurgical conditions and the development of visualization methods.

Seyed Mohsen Ranjbaran is currently pursuing his PhD in the Biomedical Engineering Department at the University of Illinois Chicago. He holds his master's degree in photonics from the University of Isfahan. His focus is on photoacoustic tomography (PAT), specifically in optimizing and advancing optical components of PAT systems. Additionally, he has played a role in enhancing reconstruction algorithms through the introduction of innovative methods.

James I. Matchynski graduated with high distinction from the University of Michigan Dearborn in 2015. He is now pursuing his dual MD/PhD at Wayne State University School of Medicine in translational neuroscience. His area of research includes using behavioral models of fear-learning to study posttraumatic stress disorder and developing a functional imaging technique that uses photoacoustic and transgenic technology to track changes in Fos neuronal ensembles in vivo.

Rayyan Manwar received his PhD from the University of Windsor, Windsor, Ontario, in 2017. His bachelor’s degree in electrical and electronic engineering from Islamic University of Technology, Bangladesh, in 2011. He is a research specialist at OPIRA lab in the Biomedical Engineering Department at the University of Illinois Chicago. Previously, he was a postdoctoral fellow at Wayne State University, Detroit, Michigan, United States. His research interests include MEMS-based design, fabrication, and characterization, photoacoustic and ultrasound imaging.

Alana C. Conti is a tenured professor of psychiatry and behavioral neurosciences at Wayne State University and a health science specialist at the John D. Dingell VA Medical Center, where she studies experimental trauma and substance abuse, focusing on neuroinflammatory mechanisms underlying pain and substance misuse susceptibility following traumatic brain injury and/or traumatic stress. Her recent studies involve the development and characterization of translationally relevant interventions to mitigate behavioral and pathological outcomes after brain injury.

Shane A. Perrine is a tenured associate professor at Wayne State University School of Medicine, leading the Perrine Lab in preclinical research on addiction, traumatic stress, and traumatic brain injury. His work is utilizing animal models, behavioral testing, molecular neuroscience, and neuroimaging, focuses on understanding brain–behavior relationships. Recent studies involve photoacoustic imaging and transgenic rodent models to explore neuronal ensembles in learned behaviors, such as conditioned fear and drug self-administration

Kamran Avanaki is an associate professor at the University of Illinois in Chicago and previously held the same position at Wayne State University. He earned his PhD from the University of Kent in 2012 with outstanding achievement in medical optical imaging and computing. Specializing in the design and the development of photoacoustic imaging technology and optical coherence tomography, his expertise addresses critical challenges in brain and skin imaging for biomedical applications.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Mohsin Zafar, Laura Stone McGuire, Seyed Mohsen Ranjbaran, James I. Matchynski, Rayyan Manwar, Alana C. C. Conti, Shane A. Perrine, and Kamran Avanaki "Spiral laser scanning photoacoustic microscopy for functional brain imaging in rats," Neurophotonics 11(1), 015007 (9 February 2024). https://doi.org/10.1117/1.NPh.11.1.015007
Received: 14 December 2022; Accepted: 8 January 2024; Published: 9 February 2024
Advertisement
Advertisement
KEYWORDS
Brain

Imaging systems

Neuroimaging

Transducers

Laser scanners

Photoacoustic microscopy

Hemodynamics

Back to Top