24 April 2017 Effects of increased compression with an ultrasound transducer on the conspicuity of breast lesions in a phantom
Author Affiliations +
Abstract
Ultrasound imaging of the breast is highly operator dependent. The amount of pressure applied with the transducer has a direct impact on the lesion visibility in breast ultrasound. The conspicuity index is a quantitative measure of lesion visibility, taking into account more parameters than standard measures that impact on lesion detection. This study assessed the conspicuity of lesions within a breast phantom using increased transducer compression in breast ultrasound.

Methods

A phantom was constructed of gelatine to represent adipose tissue, steel wool for glandular/blood vessels and silicone spheres to represent lesions, this meant that the lesions were also compressible, but less than the surrounding tissue. The phantom was imaged under increasing transducer compression. The conspicuity index was measured using the Conspicuity Index Software. The distance between the transducer surface and lesion surface was measured as an indication of increased compression.

Results

When moderate compression (17mm) was applied, the conspicuity index increased resulting in better visualisation of the silicone lesions. However, with increased compression the conspicuity index decreased.

New work to be presented

The conspicuity index has never been demonstrated in ultrasound imaging before. This is preliminary phantom work to demonstrate the impact of increased transducer compression on quantitative lesion visibility assessment.

Conclusion

The compression applied should be considered for optimum visualisation, as excessive pressure decreases conspicuity. However, further work needs to be conducted in order to consider other factors, such as density of the breast and lesion location, for a better understanding of the effect of compression on the visualisation of the lesion. A human study is planned.
Conference Presentation
© (2017) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Katy Szczepura, Tahreem Faqir, and David Manning "Effects of increased compression with an ultrasound transducer on the conspicuity of breast lesions in a phantom", Proc. SPIE 10136, Medical Imaging 2017: Image Perception, Observer Performance, and Technology Assessment, 101360T (24 April 2017); https://doi.org/10.1117/12.2254263