12 March 2021 Estimation of fluorescence probing depth dependence on the distance between source and detector using Monte Carlo modeling
Author Affiliations +
Abstract
Photosensitizer fluorescence emission during photodynamic therapy (PDT) can be used to estimate for in vivo photosensitizer concentration. We built a surface contact probe with 405nm excitation light source to obtain Photofrin fluorescence signal during clinical PDT. The probe was equipped with multiple detector fibers that were located at distances between 0.14 to 0.87 cm laterally from the excitation source fiber. In this study, we investigated the probing depth of fluorescence in biological tissue with different source-detector separation using our contact probe setup. We used Monte Carlo method to simulate the 405nm excitation light and 630nm fluorescence probing depth at various source and detector (SD) separations. The results provided insight to the most probable depth of origin of detected fluorescence at each SD separation and help to understand the in vivo depth distribution of clinically measured Photofrin concentration.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yi Hong Ong and Timothy C. Zhu "Estimation of fluorescence probing depth dependence on the distance between source and detector using Monte Carlo modeling", Proc. SPIE 11628, Mechanisms and Techniques in Photodynamic Therapy and Photobiomodulation, 116280X (12 March 2021); https://doi.org/10.1117/12.2582818
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Luminescence

Sensors

Monte Carlo methods

Tissues

In vivo imaging

Photodynamic therapy

Natural surfaces

Back to Top