2 May 1997 Effects of reduced film granularity on mammographic image quality
Phillip C. Bunch
Author Affiliations +
Abstract
The basic imaging characteristics of a new mammographic screen-film combination have been experimentally determined. These characteristics include absolutely calibrated sensitometry, modulation transfer functions, noise power spectra, noise equivalent quanta, and detective quantum efficiency. Specific image quality advantages of this new system are discussed, including a substantial reduction in film noise, and comparisons are made with an established mammography system.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Phillip C. Bunch "Effects of reduced film granularity on mammographic image quality", Proc. SPIE 3032, Medical Imaging 1997: Physics of Medical Imaging, (2 May 1997); https://doi.org/10.1117/12.273997