9 March 2012 Spatio-temporal Monte Carlo modeling of a-Se detectors for breast imaging: energy-weighted Swank noise and detective quantum efficiency
Author Affiliations +
Abstract
We study the effect of energy weighting in Swank noise and Detective Quantum Efficiency (DQE) at zero spatial frequency with a detailed Monte Carlo (MC) transport code that includes the three-dimensional spatial and temporal transport of electron-hole pairs in semiconductor x-ray detectors. The transport model takes into account recombination and trapping of carriers including effects of Coulomb forces and external applied electric field. We report pulse-height spectra (PHS) for mono-energetic x rays from 6 to 28 keV photon energy with 0.5 keV step size, and for clinical mammography spectra. A first-approximation electronic noise model is included in the simulations. The Swank calculations take into account the entire PHS distribution while the DQE(0) is calculated from the simulated Swank factor, and quantum efficiency values from the PENELOPE database of attenuation coefficients. The simulated DQE(0) is based on the entire clinical x-ray spectrum and takes into account the energy distribution following Tapiovaara and Wagner's (Phys. Med. Biol. 30, 1985) description for the weighting of carrier transport processes. Swank and DQE simulations for semiconductor detectors can provide insight into the fundamental limitations and possible optimization of breast imaging systems.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yuan Fang, Andreu Badal, Karim S. Karim, and Aldo Badano "Spatio-temporal Monte Carlo modeling of a-Se detectors for breast imaging: energy-weighted Swank noise and detective quantum efficiency", Proc. SPIE 8313, Medical Imaging 2012: Physics of Medical Imaging, 83135S (9 March 2012); https://doi.org/10.1117/12.912012
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Monte Carlo methods

Quantum efficiency

Sensors

X-rays

X-ray detectors

Breast imaging

Image sensors

RELATED CONTENT

DQE simulation of a-Se x-ray detectors using ARTEMIS
Proceedings of SPIE (March 22 2016)
Dose-efficient system for digital mammography
Proceedings of SPIE (April 25 2000)
Performance of the XMM EPIC MOS CCD detectors
Proceedings of SPIE (November 10 1998)
Gamma-ray detectors for breast imaging
Proceedings of SPIE (July 07 1997)

Back to Top