|
1.IntroductionOptical coherence tomography (OCT) is a promising noninvasive imaging modality in vivo for label-free three-dimensional tissue imaging in vivo.1 In recent years, methods for calculating attenuation coefficients from OCT signals to identify various tissue types and obtain their pathological status information have begun to attract the attention of scholars. This idea first came from ultrasonic imaging.2 When the ultrasound or infrared light penetrate the tissue, the intensity decays exponentially, and therefore the attenuation coefficient reflects the acoustic/optical characteristics of different tissues or the same type of tissues under different physiological conditions. According to some basic research, due to the lack of oxygen and energy, cerebral ischemia is accompanied by changes in cellular morphology, which optical attenuation coefficient (OAC) is sensitive to. Consequently, OAC increases during cerebral ischemia.3 Observations of ischemia tissue by transmission electron microscopy suggest that the increase in OAC is due to deformation or enlargement of mitochondria and dendrites expansion.4 Through quantitative measurement of OAC, cerebral tissue damaged by ischemia can be effectively labeled and evaluated. This method provides additional information and increases the potential of OCT in exploring cerebral infarction. The calculation of OAC from OCT signals is usually performed by fitting an exponential curve through OCT depth profiles. Nevertheless, this method requires precise layering, and the OACs of different layers are obtained separately.5,6 Because the curve-fitting method needs to be performed within a certain depth range, the image should only contain one tissue type, or the region of interest (ROI) must be selected before fitting the curve. However, the selection of ROI may have a negative impact on the fitting results, so this method may be very limited in practical applications. Some other improvements have also been proposed, including, for example, the axial point spread function (PSF),7 but none so far has fundamentally dealt with the limitations of the curve-fitting approach. Recently, a depth-resolved estimation (DRE) method has been developed to convert each pixel of an OCT signal to a corresponding OAC value.8 This method estimates OAC with much higher resolution by modeling OCT signals based on single scattering by tissue and provides per-pixel OAC values without prior layering. However, increasing errors are observed near the end of the imaging depth range. In addition, it does not take the OCT noise floor into account, which may affect the accuracy of the OAC estimation. Baran et al.9,10 combine the DRE method with the en face sorted average intensity projection (sAIP) method and calculate the average OAC for a specific en face location. They observe changes in the OAC of cerebral cortex during cerebral ischemia-reperfusion in mice and find that, during occlusion, the average OAC in the ischemic region is greater than the baseline condition. After reperfusion, the average OAC at the middle cerebral artery (MCA) side still keeps increasing significantly despite partial blood flow recovery.9,10 This study preliminarily demonstrates the potential of OCT in determining the area of cerebral infarction. But the algorithm itself has not been improved. In this paper, we propose an optimized depth-resolved estimation (ODRE) method to reduce the errors when estimating OAC. Numerical simulations and phantom experiments are used to verify its validity. We also apply this method to define the degree of cerebral damage. The changes in blood perfusion and the degree of brain injury in mice with focal cerebral ischemia are observed in real time, and the relationship between the two is analyzed. 2.Optical Coherence Tomography System and Signal ProcessingIn this study, a spectral-domain OCT (SD-OCT) system was employed, which is similar to our previous work.11,12 Briefly, a superluminescent diode with a central wavelength of 1310 nm and a bandwidth of 65 nm was used as a light source, providing an axial resolution in the tissue. In the sample arm, an achromatic lens with 30-mm focal length was used to achieve theoretical lateral resolution of . The data were acquired with 125 frames per second, where each B-frame datum was composed of 400 A-lines covering total distance. The slow axis (C-scan), consisting of 2000 B-frames, covers the same distance by repeatedly scanning five times at each position (a total of 400 locations). In SD-OCT, the presence of a fixed-pattern noise results in strong erroneous horizontal lines on the image, which usually comes from the non-flatness of the reference spectrum that is to be interfered with the signal light. It may also be produced by optical interferences that occur in the middle of the optical pathway of the reference light. It can be removed by generating a reference spectrum through recording the reference arm spectrum or averaging many spectra of an actual measurement.13 However, recording a reference spectrum is not always convenient and effective,14 so we used the latter. Each interferometric spectrum is subtracted by the mean spectrum to remove the static component. This method is popular for SD-OCT, due to its simple practical implementation and inherent immunity to reference changes. OCT signal intensity is affected by the focus location. Therefore, the confocal properties of the OCT system must be taken into account when extracting OAC quantitatively. Recently, Faber et al. derived the general expression of the confocal axial PSF,15,16 which is described as where is the OAC and is the signal depth. The function is the axial PSF, is the position of the confocal gate, and is the “apparent” Rayleigh length used to characterize the axial PSF: where is the minimum beam radius, is the center wavelength of the light source, is the refractive index (we used ), and is used to distinguish specular reflection () from diffuse reflection (). In this paper, . The bias could be removed by dividing it by axial PSF .17 To visualize the volumetric microvasculature down to the capillary level, an OCT angiography technique was utilized.3.Optical Attenuation Coefficient Estimation3.1.Depth-Resolved EstimationVermeer et al.8 introduce the model as shown in Eq. (4), where is the OCT signal of a pixel and is the OAC, both at depth , and is the pixel size (usually related to the axial resolution of the OCT system): By using this equation, the mapping relationship between the OCT signal intensity and the OAC is established. For the DRE method, the number of pixels available for calculation is limited due to the limited detection depth of the OCT system. Therefore, the OAC should be approximated as where is the number of the pixels within a limited depth. The one problem with DRE is that the approximate processing of OAC is impractical. In Vermeer et al.’s model, the boundary condition is , which means that the detection depth is infinite. However, in the actual environment, it is apparent that infinite depth is impossible. In this case, the DRE model uses a finite depth instead of , assuming that the light decays completely in a limited range of penetration. For small depths, it means that is small enough, can be considered satisfactory. As increases, the difference between and decreases, resulting in an increase in the difference between and , which is regarded to be an error. When , the error peaks.Another problem is that the DRE method has not taken the OCT noise floor into account. The noise floor includes electrical noise of the photodetector, shot noise, and relative intensity noise from the reference arm light.18 If the OCT signal is fully attenuated when it reaches the depth , then all of the depths from to may be noise. The DRE method adds a large amount of noise to the denominator of Eq. (5), and the resulting OAC may be smaller than it should be. In this paper, we propose the ODRE method to solve these problems. 3.2.Optimized Depth-Resolved EstimationThe DRE method is suitable for the boundary condition , so Eq. (4) can be used for subsequent calculations and the OAC for the last point should be expressed as where the equation can be rewritten asWith this equation, we can supplement the data after the last point , because can be split into and . Equation (4) can be rewritten as where in Eq. (8) is unknown. To determine it, a piece of data was intercepted from the ranging depth of the system and then fitted with an exponential curve (as shown in Fig. 3). The fitting model is . The fitted result is the average attenuation coefficient of this piece of data and can be considered as the best approximation of . The following is a detailed description of each step of the algorithm:
4.Results4.1.Numerical SimulationsReliable testing and simulation play an important role in the innovation of optical imaging, measurement techniques, and data analysis algorithms. To evaluate the effectiveness of the ODRE method experimentally, a series of numerical simulations were performed with and without noise. We first simulated a beam of light propagating through a homogeneous medium without noise and using three sets of OACs (0.5, 0.7, 1). The observed depth was 3 mm, and the pixel size Δ was 0.01 mm. As shown in Fig. 4(a), for ease of calculation, we assumed that the initial light intensity is 1. The OACs of the simulated data were estimated by the method of DRE and ODRE, respectively. For ODRE method, the number of pixels used for exponential fitting is 100 (1 mm). The results are shown in Figs. 4(b) and 4(c). Generally, the OACs at different depths in a homogeneous medium should be the same, but there is increasing error in the results of DRE and its maximum is . It should be noted that the estimation error of the DRE method exists at any depth and not just at the end of the data. On the other hand, OACs obtained by the ODRE method are satisfactory and consistent with the preset OACs (0.5, 0.7, and 1). As mentioned above, DRE method can only be used if almost all light is attenuated within the recorded imaging depth range. In practice, the OCT data does not necessarily satisfy such a depth range. As a result, an unacceptable error is generated, when the DRE method is used. Here, we selected three depth ranges of 0.5, 1, and 1.5 mm, respectively. The shallower the depth, the more light is left [as shown in Fig. 5(a)]. The OACs were estimated within these limited depth ranges by DRE and ODRE. The results are shown in Figs. 5(b) and 5(c), respectively. As shown in Fig. 5, a major error can occur if the optical signal attenuation is not completed within the limited depth range. The smaller the depth range, the sooner the error occurs. Instead, the ODRE method can be used to estimate the OACs in any depth range with negligible error. OCT images are often affected by speckle noise, which is a multiplicative noise of the coherent imaging system.20,21 To accurately reflect the performance of the proposed method in OAC estimation in a noisy environment, three sets of numerical signals were set up, and each set was actually a combination of multiplicative noises and the corresponding analog datasets used in Fig. 4. The increasing process of the noise follows the development: Here, is the analog signal at depth , with its counterpart used in Fig. 4. is the signal with multiplicative noise added in and is a uniformly distributed random noise with a specific amplitude. The amplitude distribution of speckles in an OCT signal can be expressed as the ratio of the standard deviation to the average value of the signal.22 According to Pircher’s theory, the contrast of the speckle pattern is 0.52.23 Speckle noise can be offset by multiple averages. In this paper, we repeat the acquisition five times at each position, and the speckle noise is reduced to of the original. Thus, the actual contrast of the speckle pattern is 0.23. So, we next simulated a signal with a multiplicative noise of 23% amplitude, as shown in Fig. 6.Figure 6(a) shows the original simulated data, and Figs. 6(b) and 6(c) show the estimated value of the OACs obtained by the DRE and the ODRE, respectively. According to Fig. 6, this series of results are similar to that of Fig. 4. Again, the results of ODRE are still consistent with the preset OACs, but the difference is that the former fluctuates by noise. These convincing simulation results show that the proposed method can accurately estimate the OACs in a random depth regardless of the noise. 4.2.Phantom Experimental ResultsFour liquid optical phantoms comprising different Intralipid concentrations were prepared. The volume-to-volume (v/v) ratios used were , 10%, 15%, and 20% (diluted in distilled water). Three methods (exponential fitting, DRE, and ODRE) were used to estimate the OAC. For the DRE method, the entire imaging depth was used for OAC estimation. For the ODRE method, only 0.5-mm depth range was used. The number of pixels used for exponential fitting is 60 (0.15 mm). Figure 7 shows the experimental results of 20% phantom. Figure 7(a) shows the OCT intensity image, Fig. 7(b) shows the corresponding OAC image obtained by DRE using Eq. (4), and Fig. 7(c) shows the corresponding OAC image obtained by ODRE using Eq. (8). Since the phantoms are homogeneous, the exponential fitting method can be applied to the entire depth range, and the results are credible and can be used as a standard for evaluating the other methods.8 The blue solid line in Fig. 7(d) is shown as the optical signal data for the leftmost column (A-scan) of Fig. 7(a). The exponential fitting method is used to process the data, and the result is shown in the red solid line in Fig. 7(d). The result () is . Figure 7(e) is the results of ODRE and DRE. It can be seen that the OAC curve within the depth range obtained by the ODRE method is even, and the average OAC value (red dotted line) is consistent with the exponential fitting result. The OAC value obtained by the DRE method is underestimated due to the accumulation of excessive noise during the estimation process. The OAC of each phantom obtained from the exponential fitting, DRE, and ODRE are listed in Table 1. Table 1OAC estimated by the exponential fitting, DRE, and ODRE methods from the uniform phantoms.
It can be seen from Table 1 that the OAC obtained by the DRE method is generally small, and the error is larger with the increase in the concentration. This is because the faster the signal attenuate, the greater is the noise accumulated by the DRE method on the denominator of Eq. (5) and the greater is the error. To better validate the capabilities of ODRE in practical applications, a multilayer phantom was created. The phantom consisted of three thin layers (15%, 5%, and 10%), 0.25 mm per layer. Each layer is separated by a 0.12-mm coverslip. The multilayer phantom has a total depth of 1.2 mm. The number of pixels used for exponential fitting is 60 (0.15 mm). The experimental results are shown in Fig. 7. Figure 8(a) shows the OCT intensity image of the multilayer phantom, A∼E represent different layers. Among them, A, C, and E are 15%, 5%, and 10% of phantom, respectively. Here, B and D are the coverslips. Figure 8(b) shows the corresponding OAC image obtained by ODRE. Figure 8(c) is a typical average A-scan within Fig 8(a), Fig. 8(d) is the OAC corresponding to Fig. 8(c). The OAC results for the three-layer phantom are , , and , respectively. A -test is performed between the OACs obtained by the ODRE from multilayered samples and the OAC obtained by the exponential curve-fitting from homogenous sample. There is no significant difference for layer A (15%), and there is a significant difference for layer C (5%) and E (10%), . This may be due to a lower SNR at a deeper depth. 4.3.In Vivo Mouse Brain ImagingAbout 3-month-old C57/616 mice (20 to 30 g) were used in the study. All procedures were performed in accordance with the Animal Ethics and Administrative Council of Northeastern University. All efforts were made to minimize animal suffering and to reduce the number of animals used. Surgical anesthesia was induced with sodium pentobarbital (3%, 5 mg/100 g, IP). The scalp was cut along the midline of the skull, and the interparietal bone was exposed by pulling the skin to the side of the head. A standard cranial window, covering the distal branches of MCA and the anterior cerebral artery (ACA), was created in the left parietal cortex, 0.5-mm lateral from the sagittal suture and 0.5-mm posterior from the bregma. A circular coverslip was placed on the exposed brain surface and sealed on bone with a dental cement.24 After surgery, the mice were placed into the sample arm of the OCT system and prepared for baseline data acquisition. The model of the focal cerebral ischemia used in our experiments was based on the previously described endothelin-1-induced middle cerebral artery occlusion (MCAO) model.12 The stereotaxic coordinates of the ET-1 injection were anterior, lateral, and 1.0-mm ventral relative to bregma. A dose of ET-1 (dilute to 100 pmol/L in 0.9% normal saline) was injected at . After injection, we waited for 3 min to ensure that the drug was completely absorbed. The needle was slowly removed, and the mouse was immediately placed under the sample arm to begin the temporal scans. During date acquisition, the mouse was under gas anesthesia (isoflurane) instead of IP anesthesia. In mouse brain imaging, all the datasets were taken at the depth of 1 mm to avoid the noise floor. The number of pixels used for exponential fitting is 120 (0.3 mm). In Fig. 9(a), the orange box represents the ROI in our experiment, highlighted with a black arrow. As shown in Fig. 9(b), the baseline en face OCT angiogram, up to the depth of , was obtained from the ROI, and yellow arrows indicate arteries whereas blue indicate veins. Figure 9(c) shows the en face OAC image obtained from the same ROI by the ODRE method. The attenuation coefficient of the ischemic region, but not the healthy region, increases with the development of ischemia so that they can be distinguished during the process. The en face OAC images were reconstructed by averaging OACs at a specific depth range (151 to ) because there is a better contrast between the damaged and the healthy tissues at this range. The OCT angiography provides the detailed information on cerebral blood flow, whereas the OAC mapping provides information on the optical properties of the tissue. These two techniques can be combined into an indicator to show the status of brain tissue. When the brain tissue is damaged by ischemia, blood perfusion changed first, and OAC changed after a period of time (20 to 40 min). Figure 9(d) shows the changes in cerebral perfusion and OAC of the cerebral cortex in mice with focal cerebral ischemia over time. As shown in Fig. 9(d), when ET-1 was injected, due to a high concentration of ET-1, the arteries and most capillaries near the injection point (a red dot) were blocked rapidly and the blood supply reached a minimum level within 10 min. However, the arteries far from the injection site were not obstructed, which may be due to some compensatory mechanisms, such as anastomosis of MCA and ACA.24 The veins also remained unchanged. Part of the reason might be that their vascular walls are relatively thin and the walls are almost unaffected by ET-1. During the period from 20 to 120 min, arteries recovered gradually, as indicated by the yellow arrows in Fig. 9(d), with part of the capillaries also returning to normal. As shown in Fig. 9(d), the white area gradually appeared in the OAC images over time, indicating that the backscattering of light increased and reflected that the optical properties of the tissue have changed. The white region was located near the blocked artery, suggesting that tissue in this area has been infarcted due to insufficient supply of blood and oxygen. In the ischemic region, lesions did not occur immediately at the onset of occlusion, and the lesion appeared and the size of the lesion area increased over time. In the later stage, after the effect induced by ET-1 disappeared, blood vessels began to recover gradually. However, we can see that the injured brain cells have not recovered, which indicates that the damage to the brain cells is irreversible. This result is consistent with the conclusion of the reference.9 To investigate the effects of cerebral ischemia on OAC in more detail, we studied the OAC of brain tissue in damaged and undamaged areas. The results are shown in Fig. 10. Figure 10(a) shows a 40-min OCT angiogram, and Fig. 10(b) shows a corresponding time OAC image. The white line in Fig. 10(a) and the yellow line in Fig. 10(b) represent the same cross-sectional location that passes through the damaged cerebral region. The OCT structural images [Figs. 10(c)–10(e)], OCT angiograms [Figs. 10(f)–10(h)], and OAC images [Figs. 10(i)–10(k)] all are at this cross-sectional location. From the OCT structural images, we can see that the penetration depth of the OCT light (green box) at baseline is significantly greater than that at 40 and 120 min. Penetration depth refers to the depth at which the signal intensity is reduced to 5% of its initial value. The penetration depth is at baseline and at 40 and 120 min. This indicates that the attenuation of light in the damaged cerebral tissue is enhanced. In the OCT angiograms, the blood vessels are abundant at baseline, and they are minimal at 40 min and recover at 120 min. It is not easy to identify changes in OAC in the damaged region from the cross-sectional images, so we chose a specific depth range of 151 to (yellow box) and plotted the average OAC curve in that range. Figures 10(l)–10(n) are the average OAC curves for the baseline, 40 min, and 120 min, respectively. From these curves, it is noticeable that the average OACs across the 151- to depth of the cross-section are approximately the same at the baseline. In contrast, at 40 and 120 min after ischemia, the OACs in the middle part (injured region) are significantly higher than those on both sides of the image. The red and blue curves in Fig. 10(o), respectively, correspond to the damaged region (red region 1) and the nondamaged region (blue region 2) shown in Fig. 10(b). The OAC curve for region 1 significantly increases after ischemia, peaks at 40 min, and then fluctuates slightly. The curve for region 2 does not change noticeably before 50 min, after which the OAC gradually increases and the gap between it and the red curve narrows. This indicates that the brain tissue in region 2 was also affected by ischemia, although it was located relatively far from the ET-1 injection site. Figures 10(p) and 10(q) are a series of OAC curves for a typical A-scan of region 1 and region 2, respectively, all of which vary with depth. Different colors and shapes are intended to represent different times. The above results show that OAC can be used as an indicator for brain injury. 5.DiscussionThe method of calculating OAC from OCT signals can be used for in vivo characterization of various tissue types with pathological conditions, such as atherosclerosis, burn scar, glaucoma,25 and ischemic brain.9,10 Vermeer et al. have confirmed that the DRE method is capable of estimating the OAC in uniform and layered, heterogeneous phantoms. However, the DRE method produces two types of errors. The first type of error increases with depth, mainly because the DRE model uses a finite depth instead of infinity. This error has no significant effect on the shallow layer when the optical signal is completely attenuated and the SNR is high enough. Noise floor exists in the OCT signal, but the DRE method uses the full-depth OCT signal as an input to estimate the OAC. As a result, a large amount of noise is added to the denominator of Eq. (5), which results in a second error: the shallow OAC is underestimated. Therefore, for the DRE method, the result of the shallow layer is smaller than it should be, and the result of the deep layer is larger than the value it should have. Cutting signals with a low SNR will shorten the estimated depth of OAC and will increase the error of the first type (see Fig. 5). The smaller the depth range, the sooner the error occurs. The existence of these two errors greatly affects the accuracy of the DRE method and limits its application. The ODRE method has solved these two errors at the same time. First, we improved the estimation model by compensating the summation of to infinity by fitting the final end of the OCT signal, which solves the first error. In addition, it is important to note that the ODRE method does not require complete attenuation of the light, so OAC can be estimated in any depth range. Therefore, we can eliminate signals with low SNR and solve the second type of error problem. These advantages make ODRE an attractive alternative for OAC calculation and may be suitable for a wider range of applications. To verify the performance of ODRE, numerical simulations and phantom experiments were performed, both of which confirmed that ODRE is capable of estimating the OAC from the OCT signals with high accuracy. In the multilayer phantom experiment, we found that for the same concentration of the phantom, the results of the multilayer phantom are less than that of the uniform phantom. It may be that the SNR is lower at a deeper depth. SD-OCT is characterized by a rapid fall-off in sensitivity along depth due to insufficient spectral resolution. Moreover, the aliasing artifact may appear at the maximum imaging depth. It is because the sampling frequency of the signal is close to or even lower than the Nyquist sampling rate at the maximum imaging depth. Increasing the upsampling ratio would decrease the amount of aliasing.26 The coherent light attenuation transmitted through turbid media is attenuated due to scattering and absorption. Most OCT systems work in near-infrared regions; they use the optical window from 800 to 1400 nm in tissue. Hence, the largest part of the OAC might be caused by scattering.27,28 In addition, because the OCT system can only receive backscattered light, the absorption coefficient cannot be detected. As a result, the DRE method assumes that the attenuation is caused by scattering, that is, the backscattered light is a fixed fraction of the attenuated light. However, blood has a strong infrared light absorption characteristic. Ignoring the absorption coefficient will result in the resulting OAC being underestimated. There are two forms of cell death in the ischemic area: apoptosis and necrosis. Researchers have proposed some convincing conclusions on the reasons for the changes in optical properties. Plesnila et al.29 claimed that a mitochondrial protein, apoptosis-inducing factor (AIF), plays a role in programmed cell death after focal cerebral ischemia. They found that AIF is localized only in the cytoplasm of almost all striatum and cortical cells 45 min after MCAO. About 1 h after MCAO, individual cortical cells showed significant AIF staining. Van der Meer et al.16 used OCT to image porcine carotid arteries in vitro and observed an increase in backscattered light in the medial layer after balloon dilation. They hypothesized that the changes in OCT signal could be attributed to this characteristic form of cell death. They then examined the optical properties of granular human fibroblasts in which necrosis or apoptosis was induced. The OAC in necrotic cells decreased from to , whereas in apoptotic cells, an increase to was observed.30 This study indicates that, in our experiment, the white region with increased OAC, as shown in Fig. 9(d), may be the result of apoptosis. This is our preliminary conclusion and we need to conduct follow-up confirmation studies. From the experiments, we observed that OAC mapping can distinguish the boundaries between infarcts and salvageable tissues at different stages of stroke development. After 2 h of real-time monitoring of mice with focal cerebral ischemia, it can be seen that when the blood flow was at a minimum, apoptosis of cerebral cortical cells did not occur, but it gradually occurred after reperfusion. It is possible that reperfusion aggravates brain damage or that apoptosis occurs only after a period of ischemia. The OAC of nondamaged area also increased over time, which explained why the contrast of the OAC image in Fig. 9(d) (120 min) is lower than before. At the same time, further studies are needed to correlate the local contrast changes with the histological results and explore the relationship between brain damage and OAC. In conclusion, OCT can provide high-resolution microvascular and structural images of the rodent brains in vivo. The OAC mapping estimated from the OCT signals can help in depicting the pathological changes in brain tissue and may serve as an indicator of tissue injury, which helps us to improve our understanding of human cerebral diseases. AcknowledgmentsThis work was supported in part by the National Natural Science Foundation of China (61771119 and 61601106), the Hebei Provincial Natural Science Foundation of China (H2018501087 and H2018501067), and Fundamental Research Funds for the Central Universities (N172304034 and N172304029). ReferencesP. H. Tomlins and R. K. Wang,
“Theory, developments and applications of optical coherence tomography,”
J. Phys. D, 38
(15), 2519
–2535
(2005). https://doi.org/10.1088/0022-3727/38/15/002 JPAPBE 0022-3727 Google Scholar
F. L. Lizzi et al.,
“Relationship of ultrasonic spectral parameters to features of tissue microstructure,”
IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 34
(3), 319
–329
(1987). https://doi.org/10.1109/T-UFFC.1987.26950 ITUCER 0885-3010 Google Scholar
M. Judith and B. Irving,
“Elastic-scattering spectroscopy and diffuse reflectance,”
Biomedical Photonics Handbook, CRC Press, Boca Raton
(2003). Google Scholar
S. Kawauchi et al.,
“Light scattering change precedes loss of cerebral adenosine triphosphate in a rat global ischemic brain model,”
Neurosci. Lett., 459
(3), 152
–156
(2009). https://doi.org/10.1016/j.neulet.2009.05.014 NELED5 0304-3940 Google Scholar
D. J. Faber, M. C. Aalders and T. G. van Leeuwen,
“Curve fitting for quantitative measurement of attenuation coefficients from OCT images,”
Proc. SPIE, 5690 325
–334
(2005). https://doi.org/10.1117/12.592371 Google Scholar
P. P. Dan et al.,
“Signal attenuation and box-counting fractal analysis of optical coherence tomography images of arterial tissue,”
Biomed. Opt. Express, 1
(1), 268
–277
(2010). https://doi.org/10.1364/BOE.1.000268 BOEICL 2156-7085 Google Scholar
M. Almasian et al.,
“Validation of quantitative attenuation and backscattering coefficient measurements by optical coherence tomography in the concentration-dependent and multiple scattering regime,”
J. Biomed. Opt., 20
(12), 121314
(2015). https://doi.org/10.1117/1.JBO.20.12.121314 JBOPFO 1083-3668 Google Scholar
K. A. Vermeer et al.,
“Depth-resolved model-based reconstruction of attenuation coefficients in optical coherence tomography,”
Biomed. Opt. Express, 5
(1), 322
–337
(2014). https://doi.org/10.1364/BOE.5.000322 BOEICL 2156-7085 Google Scholar
U. Baran, Y. Li and R. K. Wang,
“In vivo tissue injury mapping using optical coherence tomography based methods,”
Appl. Opt., 54
(21), 6448
–6453
(2015). https://doi.org/10.1364/AO.54.006448 APOPAI 0003-6935 Google Scholar
U. Baran et al.,
“Automated segmentation and enhancement of optical coherence tomography-acquired images of rodent brain,”
J. Neurosci. Methods, 270 132
–137
(2016). https://doi.org/10.1016/j.jneumeth.2016.06.014 JNMEDT 0165-0270 Google Scholar
Z. Ma et al.,
“In vivo assessment of wall strain in embryonic chick heart by spectral domain optical coherence tomography,”
Appl. Opt., 54
(31), 9253
–9257
(2015). https://doi.org/10.1364/AO.54.009253 APOPAI 0003-6935 Google Scholar
J. Liu et al.,
“Hemodynamic changes in a rat parietal cortex after endothelin-1-induced middle cerebral artery occlusion monitored by optical coherence tomography,”
J. Biomed. Opt., 21
(7), 075014
(2016). https://doi.org/10.1117/1.JBO.21.7.075014 JBOPFO 1083-3668 Google Scholar
Y. Watanabe,
“Real time processing of Fourier domain optical coherence tomography with fixed-pattern noise removal by partial median subtraction using a graphics processing unit,”
J. Biomed. Opt., 17
(5), 050503
(2012). https://doi.org/10.1117/1.JBO.17.5.050503 JBOPFO 1083-3668 Google Scholar
S. Moon, S. W. Lee and Z. Chen,
“Reference spectrum extraction and fixed-pattern noise removal in optical coherence tomography,”
Opt. Express, 18
(24), 24395
–24404
(2010). https://doi.org/10.1364/OE.18.024395 OPEXFF 1094-4087 Google Scholar
T. G. Van Leeuwen, D. J. Faber and M. C. Aalders,
“Measurement of the axial point spread function in scattering media using single-mode fiber-based optical coherence tomography,”
IEEE J. Sel. Topics Quantum Electron., 9
(2), 227
–233
(2003). https://doi.org/10.1109/JSTQE.2003.813299 IJSQEN 1077-260X Google Scholar
F. J. van der Meer et al.,
“Quantitative optical coherence tomography of arterial wall components,”
Laser Med. Sci., 20
(1), 45
–51
(2005). https://doi.org/10.1007/s10103-005-0336-z Google Scholar
G. T. Smith et al.,
“Automated, depth-resolved estimation of the attenuation coefficient from optical coherence tomography data,”
IEEE Trans. Med. Imaging, 34
(12), 2592
–2602
(2015). https://doi.org/10.1109/TMI.2015.2450197 ITMID4 0278-0062 Google Scholar
S. H. Yun et al.,
“High-speed spectral-domain optical coherence tomography at wavelength,”
Opt. Express, 11
(26), 3598
–3604
(2003). https://doi.org/10.1364/OE.11.003598 OPEXFF 1094-4087 Google Scholar
G. Liu et al.,
“A comparison of Doppler optical coherence tomography methods,”
Biomed. Opt. Express, 3
(10), 2669
–2680
(2012). https://doi.org/10.1364/BOE.3.002669 BOEICL 2156-7085 Google Scholar
G. Gong, H. Zhang and M. Yao,
“Speckle noise reduction algorithm with total variation regularization in optical coherence tomography,”
Opt. Express, 23
(19), 24699
–24712
(2015). https://doi.org/10.1364/OE.23.024699 OPEXFF 1094-4087 Google Scholar
A. Pižurica et al.,
“Multiresolution denoising for optical coherence tomography: a review and evaluation,”
Curr. Med. Imaging Rev., 4
(4), 270
–284
(2008). https://doi.org/10.2174/157340508786404044 Google Scholar
B. Karamata et al.,
“Speckle statistics in optical coherence tomography,”
J. Opt. Soc. Am. A, 22
(4), 593
–596
(2005). https://doi.org/10.1364/JOSAA.22.000593 JOAOD6 0740-3232 Google Scholar
M. Pircher et al.,
“Speckle reduction in optical coherence tomography by frequency compounding,”
J. Biomed. Opt., 8
(3), 565
–569
(2003). https://doi.org/10.1117/1.1578087 JBOPFO 1083-3668 Google Scholar
U. Baran, Y. Li and R. K. Wang,
“Vasodynamics of pial and penetrating arterioles in relation to arteriolo–arteriolar anastomosis after focal stroke,”
Neurophotonics, 2
(2), 025006
(2015). https://doi.org/10.1117/1.NPh.2.2.025006 Google Scholar
D. S. J. Van et al.,
“The effect of glaucoma on the optical attenuation coefficient of the retinal nerve fiber layer in spectral domain optical coherence tomography images,”
Invest. Ophthalmol. Visual Sci., 53
(4), 2424
–2430
(2012). https://doi.org/10.1167/iovs.11-8436 IOVSDA 0146-0404 Google Scholar
K. K. H. Chan and S. Tang,
“High-speed spectral domain optical coherence tomography using non-uniform fast Fourier transform,”
Biomed. Opt. Express, 1
(5), 1309
–1319
(2010). https://doi.org/10.1364/BOE.1.001309 BOEICL 2156-7085 Google Scholar
S. J. Matcher, M. Cope and D. T. Delpy,
“In vivo measurements of the wavelength dependence of tissue-scattering coefficients between 760 and 900 nm measured with time-resolved spectroscopy,”
Appl. Opt., 36
(1), 386
–396
(1997). https://doi.org/10.1364/AO.36.000386 APOPAI 0003-6935 Google Scholar
M. Hohmann et al.,
“Extension of depth-resolved reconstruction of attenuation coefficients in optical coherence tomography for slim samples,”
Proc. SPIE, 9792 97920P
(2015). https://doi.org/10.1117/12.2205030 PSISDG 0277-786X Google Scholar
N. Plesnila et al.,
“Nuclear translocation of apoptosis-inducing factor after focal cerebral ischemia,”
J. Cerebr. Blood Flow Metab., 24
(4), 458
–466
(2004). https://doi.org/10.1097/00004647-200404000-00011 Google Scholar
F. J. van der Meer et al.,
“Apoptosis- and necrosis-induced changes in light attenuation measured by optical coherence tomography,”
Laser Med. Sci., 25
(2), 259
–267
(2010). https://doi.org/10.1007/s10103-009-0723-y Google Scholar
BiographyZhenhe Ma is an associate professor and a master student mentor at the School of Control Engineering, Northeastern University, Qinhuangdao, China. He was a postdoctoral fellow at the Oregon Health & Science University (2009-2010). His main research directions are optical coherence tomography, infrared spectroscopy, and Raman spectroscopy. |