3 October 2023 Quantitative hemodynamic imaging: a method to correct the effects of optical properties on laser speckle imaging
Thinh Phan, Christian Crouzet, Gordon T. Kennedy, Anthony J. Durkin, Bernard Choi
Author Affiliations +
Abstract

Significance

Studying cerebral hemodynamics may provide diagnostic information on neurological conditions. Wide-field imaging techniques, such as laser speckle imaging (LSI) and optical intrinsic signal imaging, are commonly used to study cerebral hemodynamics. However, they often do not account appropriately for the optical properties of the brain that can vary among subjects and even during a single measurement. Here, we describe the combination of LSI and spatial-frequency domain imaging (SFDI) into a wide-field quantitative hemodynamic imaging (QHI) system that can correct the effects of optical properties on LSI measurements to achieve a quantitative measurement of cerebral blood flow (CBF).

Aim

We describe the design, fabrication, and testing of QHI.

Approach

The QHI hardware combines LSI and SFDI with spatial and temporal synchronization. We characterized system sensitivity, accuracy, and precision with tissue-mimicking phantoms. With SFDI optical property measurements, we describe a method derived from dynamic light scattering to obtain absolute CBF values from LSI and SFDI measurements. We illustrate the potential benefits of absolute CBF measurements in resting-state and dynamic experiments.

Results

QHI achieved a 50-Hz raw acquisition frame rate with a 10 × 10 mm field of view and flow sensitivity up to ∼4 mm / s. The extracted SFDI optical properties agreed well with a commercial system (R2 ≥ 0.98). The system showed high stability with low coefficients of variations over multiple sessions within the same day (<1 % ) and over multiple days (<4 % ). When optical properties were considered, the in-vivo hypercapnia gas challenge showed a slight difference in CBF (−1.5 % to 0.5% difference). The in-vivo resting-state experiment showed a change in CBF ranking for nine out of 13 animals when the correction method was applied to LSI CBF measurements.

Conclusions

We developed a wide-field QHI system to account for the confounding effects of optical properties on CBF LSI measurements using the information obtained from SFDI.

1.

Introduction

The brain is a nutrient-demanding organ whose hemodynamics are well regulated to ensure proper functional activities. Blood flow in the brain is tightly controlled to maintain homeostasis (i.e., vascular reactivity) and to deliver sufficient glucose, oxygen, and other nutrients to support neuronal activities (i.e., neurovascular coupling). The neurovascular unit plays a crucial role in controlling essential vascular activities, such as vasodilation and constriction to regulate cerebral hemodynamics.1,2 Studying cerebral hemodynamics at resting states or in relationship to perturbations may provide diagnostic or prognostic information on the onset or progression of brain conditions, such as stroke and Alzheimer’s disease (AD).1,3,4

In preclinical models, the change in cortical hemodynamics can be quantified using reflectance techniques, such as laser speckle imaging (LSI) and optical intrinsic signal imaging (OISI).58 LSI measures cerebral blood flow (CBF),9 while OISI measures hemoglobin concentrations and their derivative parameters, such as cerebral blood volume and tissue oxygen saturation (StO2).5,1012 With LSI, local speckle contrast values approximate the speckle decorrelation time, which is assumed to be inversely proportional to the speed of moving scatterers, such as red blood cells.9 With OISI, spectroscopic analysis can be applied to the measured reflectance changes at multiple wavelengths to obtain chromophore concentrations such as deoxyhemoglobin (HbR) and oxyhemoglobin (HbO2).10,11 LSI and OISI have been used in tandem to obtain a more complete view of hemodynamics following stimulation and in disease models.6,13,14

Optical scattering complicates the interpretation of OISI and LSI data. Measured differences in reflectance are affected by the absorption and scattering properties of the interrogated tissue. Scattering lengthens the path length over which the detected reflected light travels in an unknown manner. With planar reflectance measurements alone, it is challenging to separate the effects of scattering from absorption. Hence, OISI measurements oftentimes are normalized to the corresponding initial values, to estimate relative changes in HbR and HbO2 during time-resolved reflectance measurements while assuming scattering remains constant. With LSI, speckle contrast values are affected by changes in optical properties.15 Specifically, an increase in absorption can increase the measured speckle contrast despite a constant flow value. The opposite applies to scattering, where an increase in scattering results in a decrease in speckle contrast.1517 Nonetheless, this dependency of speckle contrast on optical properties is not typically considered in studies that employ LSI.

Spatial frequency domain imaging (SFDI) enables the quantitative measurement of wide-field spatially resolved optical properties.18,19 Using a spatially patterned illumination, in combination with an appropriate model of light propagation in turbid media, SFDI can determine pixel-wise mapping of absorption (μa) and reduced scattering (μs) coefficients. Within the context of rodent neuroimaging, SFDI was previously used to study the difference in brain optical properties of wild-type and AD mouse models.20,21 We also reported on a combination of LSI and SFDI in a fast imaging system to image cortical hemodynamics in a rat model of cardiac arrest.16,2224 However, in these investigations, the LSI and SFDI fields of view (FOV) were mismatched in terms of size, temporal synchronization, and spatial overlap and thus required additional post-acquisition image alignment and analysis. Furthermore, the LSI and SFDI data were used strictly in a complementary manner, and tissue’s absorption and scattering effects on LSI data were not addressed.16

Here, we introduce a quantitative hemodynamic imaging (QHI) system that combines LSI and SFDI into an integrated device, with matching FOV and temporal synchronization. Details on hardware setup and software for acquisition and analysis are provided. We provide system characterization results for sensitivity, accuracy, and precision tests using tissue-mimicking silicone phantoms. Using the direct measurement of optical properties from SFDI, we describe a method derived from the dynamic light scattering literature to obtain an absolute CBF value from LSI and SFDI measurements. Finally, we illustrate the potential benefits of such absolute measurements in resting-state and hypercapnia gas challenge mouse models.

2.

System Instrumentation

2.1.

SFDI Hardware

The hardware design and assembly of the SFDI arm of the system were based on the openSFDI guide25 by Applegate et al.,26 with adjustments to enable continuous acquisition. Since the typical mouse heart rate ranges from 5 to 14 Hz, we designed QHI to achieve raw acquisition frame rate of 50 Hz to capture pulsatile signals using LSI. This frame rate allowed for a slower effective SFDI frame rate of 2.8 Hz for three wavelengths and two spatial frequencies. On the detection arm, a fast CMOS camera (BFS-U3-32S4M-C, FLIR, Oregon, United States) was mounted so that the imaging axis was orthogonal to the sample plane [Fig. 1(a)]. We used a 50 mm fixed focal length lens (67-717, Edmund Optics, New Jersey, United States) at maximum zoom together with a 10 mm spacer (CML10, Thorlabs, New Jersey, United States) to achieve a 10×10  mm FOV (magnification = 0.42) at 1200×1200  pixel resolution. The FOV was chosen to match the dimensions of a mouse cranial window. The working distance was 130  mm. A notch filter at 632.8 nm (86-126, Edmund Optics, New Jersey, United States) was used to minimize possible crosstalk with the LSI wavelength [Fig. 1(a)].

Fig. 1

Overall schematic of the QHI system. (a) Hardware setup of QHI with LSI and SFDI components [AL, aspheric lens; AcL, achromatic lens; CL, condenser lens; D, diffuser; DMD, digital micromirror device; F, appropriate notch (for SFDI) and line filters (for LSI) at LSI wavelength, IL, imaging lens with spacers; L, plano-convex lens; LED, light emitting diode; LPDM, long-pass dichroic mirror; P, linear polarizer (cross-polarized between illumination and detection arms); and SPDM, short-pass dichroic mirror]. (b) Triggering scheme of the system where the computer sends a serial trigger to activate an automatic Arduino triggering scheme for the LED and the DMD. The electrical output of the DMD was then used to trigger the camera. The exposure time was set to 10 ms with a 20 ms period per full cycle, allowing for an effective 50 Hz raw frame rate. The Arduino triggering scheme can be deactivated with another serial trigger from the computer.

NPh_10_4_045001_f001.png

The SFDI illumination arm was configured with an illumination axis at 24  deg with respect to the imaging arm [Fig. 1(a)] to minimize distortion artifacts (e.g., keystone artifacts) on projected SFDI patterns within the FOV. All optical components were set up at 190 mm away from the mounting breadboard to provide ample space for animal imaging. The projection system included three LEDs at 660, 780, and 850 nm (M660L4, M780LP1, and M850L3, respectively; Thorlabs, New Jersey, United States) coupled into a digital micro-mirror device (DMD) projector (DLP4500 NIR, Keynote Photonics, Texas, United States) using appropriate dichroic mirrors (FF697-SDi01-25x36, Semrock, New York, United States, and DMSP805R Thorlabs, New Jersey, United States) and condenser lenses (ACL2520U-B, Thorlabs, New Jersey, United States) [Fig. 1(a)]. The three wavelengths were chosen to enhance sensitivity to the chromophores of interest (i.e., HbR and HbO2) based on their absorption spectrum and to approximately match the probing depth of the LSI wavelength (633 nm). To achieve sufficient light intensity for fast acquisition (i.e., 50 Hz), we added a plano-convex lens (LA1708-AB, Thorlabs, New Jersey, United States) in front of the combined light beam before the DMD [Fig. 1(a)]. The plano-convex lens position was carefully adjusted to avoid creating a focused image of the LED on the DMD. A 400 to 1100 nm achromatic lens (AC254-075-AB-ML, Thorlabs, New Jersey, United States) was positioned 100  mm from the DMD so that a focused image of roughly 3× magnification of the DMD’s display was formed on the focal plane. We then made fine adjustments to ensure that the illumination and detection arms were confocal using projection alignment steps listed in the openSFDI guide.27 A barrier was then added to block unwanted diffraction patterns from the DMD from the imaging axis. We controlled the LEDs using dedicated drivers (LEDD1B, Thorlabs, New Jersey, United States) with triggering signals from an Arduino Uno (1050-1024-ND, Digi-Key Electronics, Minnesota, United States). Details regarding the electrical configuration are discussed in the Acquisition Software section below.

Finally, linear polarizers (LPNIRE100-B, Thorlabs, New Jersey, United States) were added to achieve crossed polarization between the illumination and detection arms while maintaining appropriate intensity counts (i.e., 40000 to 45000 counts for 16-bit images) from a static silicone phantom of known optical properties (μa=0.019, 0.019, and 0.019  mm1, μs=1.07, 0.83, and 0.72  mm1 at 660, 780, and 850 nm, respectively).

2.2.

LSI Hardware

For the LSI detection arm, an identical CMOS camera and lens-spacer system to that of the SFDI arm were used to achieve a matching FOV. We combined the two detection arms using a 638 nm long pass dichroic mirror (DMLP638R, Thorlabs, New Jersey, United States) [Fig. 1(a)]. For LSI, we adjusted the imaging lens aperture to a f/# of 4 to achieve a proper speckle-to-camera pixel size ratio (approximately 9  μm to 3.45  μm, respectively).28 A 632.8 nm laser line filter (FL632.8-3, Thorlabs, New Jersey, United States) was used to minimize crosstalk with illumination from the SFDI arm [Fig. 1(a)].

For the LSI illumination arm, a long-coherence 633 nm laser (115-81059-554, Coherent, California, United States) was coupled with an aspheric lens (C330TMD-B, Thorlabs, New Jersey, United States) to quickly expand the beam. A 1500-grit diffuser (DG-10-1500-MD, Thorlabs, New Jersey, United States) was used to homogenize the beam. The laser was then mounted to achieve homogeneous illumination over the camera’s FOV with the beam concentric with the FOV [Fig. 1(a)]. We adjusted the laser’s illumination power using a polarizer (LPNIRE100-B, Thorlabs, New Jersey, United States) to achieve the appropriate intensity count (i.e., 60-70 for 8-bit images) on the same static silicone phantom used for the SFDI setup. Similar to the SFDI arm, cross-polarization with linear polarizers was used to minimize specular reflectance [Fig. 1(a)].

2.3.

Computer Specification and Electrical Wiring

The performance and effective frame rate of the system also depend on the specifications of the acquisition computer. Here, we used a laptop with Intel® Core™ i7-9750H CPU (12 CPUs) processor, 16 GB RAM, and a 1TB built-in M.2. solid state drive (SSD). The LED drivers were connected to the Arduino triggering digital pins using BNC cables with the ground leads connected to the Arduino GND pin. On the DMD, the “TRIG IN” connector J7 was connected to the Arduino DMD trigger pin and the “TRIG OUT” connector J11 to the cameras’ trigger-in pin through GPIO cables. All ground outputs from the DMD and the cameras were connected to the Arduino GND pin. Micro-B-to-A-male USB-3.0 cables were used to connect the cameras and the computer. One B-to-A male USB-2.0 cable was used to connect the Arduino to the computer.

3.

Acquisition Software

We developed a master acquisition program in LabVIEW (LabVIEW 2020, NI, Texas, United States) to control the Arduino and camera acquisition. For camera acquisition, we used the NI Vision Acquisition Software suite to set the camera’s acquisition mode to external triggering, gain to zero, and exposure time to 10 ms. Gamma correction was disabled, and trigger delay time was set to 11  μs (i.e., smallest value). The image format was set to uncompressed TIFF at 1200×1200  pixel resolution with a bit depth of either 8-bit (LSI) or 16-bit (SFDI), aligned to the center of the sensor. We used the LabVIEW master program to write a serial trigger for the Arduino to activate an autonomous triggering scheme [Fig. 1(b)]. Here, the autonomous triggering scheme stops after the Arduino receives a deactivation serial trigger from the computer.

The autonomous Arduino triggering scheme includes cycling through wavelengths first and then spatial patterns to mitigate temperature-dependent drifting of LED intensity. The code controlled the timing of the DMD and the LED. Here, we ran the DMD in pattern sequence mode with an external trigger (refer to Keynote Flexlight LC4500 LC4500-RGB-EKT Electronics Kit User’s Guide Sec. 3.3.229). Images of the chosen spatial frequencies (fx=0 and 0.3  mm1) at three phases (0 deg, 120 deg, and 240 deg) were loaded onto the DMD’s internal memory. We note that the DMD’s maximum speed was limited by the loading time of the images. Here, only two 24-bit images could be loaded and displayed at a full frame rate.

Next, an input signal from the Arduino switched the LEDs for every frame and advanced the pattern once every three frames via TTL triggering. The exposure and period time of the DMD were set to match the camera’s exposure time of 10 ms. The trigger output signal from the DMD was delivered 11  μs before the pattern display to account for the camera’s 11  μs minimum trigger-to-exposure delay time. The delay time between two consecutive triggering cycles was set to 20 ms, which effectively resulted in a raw 50 Hz raw acquisition rate for LSI and SFDI [Fig. 1(b)]. For our current computer specifications, this frame rate was appropriate to minimize frame-dropping issues. This led to effective 2.8 and 50 Hz frame rates for SFDI and LSI, respectively.

The images were saved with time stamps and an order index of 0 to 17 (a total of 18 frames needed for SFDI at 3 wavelengths, 2 spatial frequencies, and 3 phases). The indices were later used to check and account for any dropped frames during analysis.

4.

Processing Software

Due to fast acquisition at full uncompressed 1200×1200  pixel resolution at 8- and 16-bit depth (i.e., 216 megabytes per second) on a single SSD hard drive, we expected incidents of dropped frames. We noted that dropped frames happened when there were background tasks such as file transferring and Windows updates. Thus, we used the indices of each successfully saved image to check for a complete set of 18 consecutive frames in LSI and SFDI data (i.e., consecutive indices from 0 to 17) before performing further processing steps listed below. If a dropped image was detected in either LSI or SFDI, the entire set of 18 frames for both modalities was disregarded in further processing steps.

4.1.

SFDI Processing Pipeline

We used MATLAB (R2019b, MathWorks, Massachusetts, United States) to process SFDI data following a previously published protocol.19 The first step of the processing scheme was pixel-wise spatial frequency demodulation using the three phases collected to obtain the AC signal of the reflectance

Eq. (1)

MAC(fx)=23[(I1I2)2+(I2I3)2+(I3I1)2]12,
where I1, I2, and I3 are reflectance images at a spatial frequency fx at 0 deg, 120 deg, and 240 deg phases, respectively. A Gaussian filter (15×15  pixel window size) was applied to the demodulated AC signal images to remove noise. The AC images (MAC) were then calibrated against the calibration phantom MAC,phantom and a diffuse reflectance value of the phantom (Rd,phantom) obtained from a white Monte Carlo (WMC)30 simulation to obtain a diffuse reflectance (Rd) term

Eq. (2)

Rd(fx)=MAC(fx)MAC,phantom(fx)Rd,phantom(fx).

The next step of processing included an inverse lookup table (LUT) scheme to match the measured pairs of Rd to a pair of μa and μs values. Here, the LUT for Rd(fx) at fx=0 and 0.3  mm1 was generated using the same WMC code mentioned above with μa range of 0.0001 to 0.2  mm1 and μs range of 0.001 to 3  mm1 for 512 linearly spaced values. We employed a rapid LUT method using an interpolation scheme for scattered data from a built-in MATLAB function (scatteredInterpolant) to generate an inverse LUT for a pair of μa and μs at each pixel.

After the extraction of μa and μs, we calculated chromophore concentrations using μa at all three wavelengths. Here, we calculated an inverse matrix of extinction coefficients at each wavelength for HbO2 and HbR, which was then multiplied with μa at each pixel. In Matlab, this method is analogous to least square fitting when the matrices have mismatched dimensions. Finally, we used HbR and HbO2 concentration to calculate physiologically relevant parameters, such as HbT and StO2 using their definitions

Eq. (3)

HbT=HbO2+HbR,

Eq. (4)

StO2=HbO2HbT×100.

4.2.

LSI Processing Pipeline

We employed a typical spatial speckle contrast algorithm with a 5x5 sliding window for analysis.9 We used MATLAB to perform the computations using the following equation:

Eq. (5)

K=σ(I)I,
were σ(I) and I are the standard deviation and mean values of the intensity, respectively, within the sliding window. Then, the center pixel of the window was assigned the contrast value K. We further converted speckle contrast K to speckle flow index (SFI) using the following simplified speckle imaging equation:31,32

Eq. (6)

SFI=12TK2,
where T is the exposure time of the camera. Our previous publications have shown that SFI closely approximates 1τc, where τc is the decorrelation time of the electric field autocorrelation function.31,32 In most laser speckle neuroimaging literature, 1τc is considered an approximation of CBF.5,6,9

For each LSI measurement, β is a normalization factor that takes into account the mismatch between detector and speckle sizes and polarization. Typically, β can be estimated using speckle contrast K (where τcT) measured from a static phantom as βKphantom2.9 We then use β to correct for SFI

Eq. (7)

SFIβcorrected=β·SFI.

4.3.

Obtaining Absolute Flow Values by Correcting for Optical Properties

In this section, we describe in detail a protocol used to obtain an absolute measurement of flow by correcting speckle contrast K with optical properties. We first started with an equation relating K to electric field autocorrelation G1 as described by Mazhar et al.15

Eq. (8)

K2=2βT0TG12(τ)(1τT)dτG12(τ=0).

The associated solution for G1 in the spatial frequency domain is

Eq. (9)

G1(τ)=3P0Aμsμtr(μeffμtr+1)(μeffμtr+3A).

In Eq. (9), μtr=μa+μs is the transport coefficient, and P0 is the illumination power. Here, A was defined as

Eq. (10)

A=1Reff2(1+Reff);where  Reff0.0636n+0.668+0.710n1.440n2.

In the spatial frequency domain, μeff is defined as

Eq. (11)

μeff=  μeff2+(2πfx)2;where  μeff=3(μa+13μs(2πnλ)2(6Dbτ))μtr.

In this equation, fx is the spatial frequency of the illumination pattern; n is the refractive index; λ is the LSI wavelength; and Db is the “effective” Brownian diffusion coefficient describing the motion of scatterers.33 This equation is often used in dynamic light scattering theory to describe the mean-squared displacement of moving particles. Furthermore, the Brownian model yielded a better fit for the observed correlation decay curves in various tissue types.34 In this work, we assumed that LSI illumination is planar; thus, fx is assumed to be 0.

Our goal was to retrieve Db using μa, μs, and speckle contrast K. The overall schematic of deriving Db from measured parameters is described in Fig. 2. First, we used concentrations of HbO2 and HbR determined with SFDI to calculate μa at 633 nm. Then, to determine μs at 633 nm, we fit an exponential function in the form of the Mie-scattering exponential decay function35

Eq. (12)

μs(λ)=A(λ850)blog(μs(λ))=log(A)+b·log(λ850).

Fig. 2

Flow chart of the method of obtaining absolute flow value using the Brownian diffusion coefficient Db from measured K, μa, and μs. Here, a β factor was approximated using a static silicone phantom.

NPh_10_4_045001_f002.png

To perform the fit more efficiently, we rewrote the decay function, as shown in Eq. (12). We then solved for log(A) and b using MATLAB “\” operator using the μs measured at 660, 780, and 850 nm. From log(A) and b, we obtained a value of μs at 633 nm with a forward solution. We used a β factor to create a LUT of K2 values as a function of Db, μa, and μs by numerically integrating Eq. (8). The values of μa and μsn in the LUT were chosen to cover the range that included the values at 633 nm (i.e., μa from 0.0001 to 0.3  mm1 and μs from 0.001 to 3  mm1 with 128 equally spaced values), and the range of Db was set to 0 to 206  mm2/s with 256 equally spaced values. Using the MATLAB function scatteredInterpolant, we then obtained the value of Db from the set of measured K, μa, and μs at 633 nm.

5.

Characterization

For each of the experiments listed below, the system was turned on and warmed up for at least 10 min before imaging. This allowed for the LEDs in the SFDI system to reach a steady-state temperature, which was expected to minimize longitudinal drift in the illumination.

5.1.

Flow Sensitivity Using LSI with a Flow Phantom Setup

We used a flow phantom setup to characterize the flow sensitivity of the LSI arm of QHI as previously described.36,37 The static part of the phantom was made using silicone with no added absorbers and TiO2 as scatterers. The optical properties of the static portion were measured using SFDI, with μa=0.0049, 0.0059, and 0.0062  mm1 and μs=0.87, 0.67, and 0.60  mm1 at 660, 780, and 850 nm, respectively. A 25  μL glass capillary pipet tube (Kimble 71900, inner radius 330  μm) was embedded in the silicone phantom so that the tube was flush with the surface of the phantom [Fig. 3(a)]. A 1% intralipid solution was prepared and infused into the capillary tube using a precision pump (70-4504, Harvard Apparatus, Massachusetts, United States) at incremental speeds of 0, 0.5, 1, 2, 3, 4, and 5  mm/s for 60 s at each speed. The imaging session lasted 7 min in total. We repeated the experiment for 5 trials, with 15 min of wait time between each trial. This wait time is implemented to ensure that the flowing fluid is restored to a baseline state.

Fig. 3

Flow phantom experiment to characterize the sensitivity of the LSI arm. (a) A schematic of the silicone flow phantom with an embedded glass capillary pipet tube where a solution of 1% intralipid was flowed at different speeds using a syringe pump. A representative flow map of the captured FOV was also shown with the chosen ROI (black rectangles) for static and flow regions. (b) Time-course data of the experiment where the flow speed was increased at each 1-min interval (top table). The time course of SFIβ-corrected within the flow region showed an increasing pattern with increasing speed between 0 and 4  mm/s. A plateau pattern was observed at higher speeds (>4  mm/s). For the static region, SFIβ-corrected showed no change during the experiment.

NPh_10_4_045001_f003.png

Two regions of interest (ROIs) were chosen over the dynamic and static region of the phantoms and median values of SFIβ-corrected were calculated at each imaging time point and filtered using a 50-frame (1  s) moving average filter. The resulting time course data sets are shown below in Fig. 3(b). Here, the dynamic ROI showed the expected increase in SFIβ-corrected values with increasing speed, with the sensitivity to flow reaching a plateau starting at 4  mm/s, similar to what has been reported previously with T=10  ms.31 The static ROI showed no change in SFI despite the increasing speeds.

5.2.

Optical Property Accuracy Assessment of SFDI

We fabricated four static silicone phantoms and compared optical property measurements between the SFDI arm of QHI with a commercial SFDI system. The phantoms were made using a base of silicone, TiO2 powder as scatterers, and India ink as absorbers. One phantom had a specific set of μa and μs, and the other three phantoms were fabricated to have the following optical property values relative to the first phantom: (1) same μa, 2×μs; (2) same μs, 2×μa; and (3) 2×μa and 2×μs. Here, the values of μa and μs measured using QHI were compared with measurements from a commercial SFDI system (ReflectRS, Modulim, California, United States). A common circular ROI with a diameter of 3.3 mm was selected and median values of the optical properties were extracted. We performed a linear regression analysis using the median μa and μs values from the four phantoms measured using QHI and the ReflectRS for 660 and 850 nm. This wavelength choice resulted from the fact that the ReflectRS does not have a 780 nm LED. Here, we observed a high linear correlation between QHI and the ReflectRS (R2>0.98) for both μa and μs (Fig. 4). Using a Bland–Altman plot, we observed a mean difference of 0.0059 and 0.0316  mm1 for μa and μs, respectively (Fig. 4). Collectively, these data suggest that QHI slightly underestimated μa and slightly overestimated μs; however, this discrepancy was linear for various combinations of μa and μs (as shown through high R2). One reason for this discrepancy could be the mismatched FOV of the two systems, as QHI had a much smaller total FOV (i.e., 1×1  cm versus 20×15  cm, respectively, 300× smaller) due to its higher magnification, with an increased sensitivity to any heterogeneities within the phantom (e.g., locally dense regions of TiO2), which could skew the median value.

Fig. 4

Correlation (top) and Bland-Altman plots (bottom) between QHI and a commercial ReflectRS system from Modulim of four static silicone phantoms. Here, we compared the median values obtained from the chosen ROI. Only 660 nm and 850 nm wavelengths were compared, as 780 nm is not available with the ReflectRS. For the correlation plot, R2 values were also listed for the linear fit.

NPh_10_4_045001_f004.png

5.3.

Intra-Day Inter-Session Precision Measurements of Static Silicone Phantom

To investigate the stability of the system during a session and among multiple sessions within one day of imaging, we imaged phantom 1 in six separate trials. Here, each trial lasted 5 min, and the time between the start of each trial was 30 min. The total length of the experiment was 3 h. A circular diameter with a radius of 200 pixels (1.6  mm) was chosen at the center of the FOV and median values of μa, μs, and K were recorded at each respective wavelength at each time point. A whisker bar graph of all median values for the five minutes of imaging for each measured optical parameter at each respective wavelength is shown in Fig. 5. We also calculated the coefficient of variation across all trials using the mean value of each trial’s time course. For SFDI measurements of μa and μs of all wavelengths, the standard deviations for each of the trials were <0.001 while the coefficient of variation across all trials was <0.5%. For LSI measurements of K, we observed standard deviations ranging from 0.0014 to 0.0013 for each of the trials with a coefficient of variation across all trials of 0.16%. These values were listed in Table S1 in the Supplementary Material and indicated good intraday stability of the system.

Fig. 5

Measurements of QHI precision were performed for six trials (5 min each) within the same day on the same static silicone phantom (i.e., phantom 1). The mean and standard deviation are plotted here. The recorded values for μa, μs, and K and the calculated coefficient of variation across all trials were shown in Table S1 in the Supplementary Material.

NPh_10_4_045001_f005.png

5.4.

Longitudinal Precision Measurement of Static Silicone Phantom

Using the four static silicone phantoms listed above, we conducted longitudinal precision measurements over 5 days. A single full set of LSI and SFDI frames (a total of 18 frames) was taken and analyzed. A circular ROI with a radius of 200 pixels (1.6  mm) was chosen at the center of the FOV and median values of μa, μs, and K were determined at each respective wavelength. These median values across the 5 days were then used to calculate a coefficient of variation for each measured parameter. Here, we noted that all coefficients of variations were <4% (Fig. 6, Tables S2–S4 in the Supplementary Material), indicating good precision of the system across multiple days. Here, K will theoretically be 1 for a completely static phantom. However, the measured K from the four phantoms ranged between 0.5 and 0.7 due to potential environmental, experimental, or instrumental sources. For phantom 2, K increased with increasing μa values compared to those of phantom 1 due to the decreasing probability of detection of a photon hitting a dynamic scattering event (i.e., a dynamic photon). On the contrary, phantom 3 showed a decrease in contrast K with increasing μs due to the increasing probability of detection of a dynamic photon. Finally, phantom 4 showed similar K values to those of phantom 1 when both increasing μa and μs. The measured K values differed across the four phantoms due to μa and μs changes as expected from Eqs. (8)–(11) (Fig. 6, red dotted line for K graphs).

Fig. 6

Measurements of QHI precision over five days. The median values within the selected ROI for each day are plotted and further used in the analysis. The recorded values for μa, μs, and K and the calculated coefficient of variation across all five days are shown in Tables S2–S4 in the Supplementary Material. K is higher for the phantom with higher μa (phantom 1 versus phantom 2) and lower for the phantom with higher μs (phantom 1 versus phantom 3). With higher μa and μs (phantom 4), K is similar to that of phantom 1. This agrees with the findings of Mazhar et al.15

NPh_10_4_045001_f006.png

6.

In-vivo Neuroimaging Demonstration

We then assessed the in vivo efficacy of the system in two experiments: (1) measuring intra-animal changes in hemodynamics during hypercapnia and (2) inter-animal comparison of resting-state CBF measured using either SFI or Db. For both experiments, the correction of CBF using the derived optical properties showed impactful changes.

6.1.

Intra-Animal CBF Measurement During Hypercapnia Gas Challenge

To assess the impact of the processing methods on dynamic measurements, we subjected a female mouse (C57BL/6, 5 months old) to a cranial window preparation for hypercapnia. Anesthesia was initiated at 4% isoflurane using an induction chamber and the animal was maintained at 2% for the surgical period at 1  L/min flow rate. For the gas challenge, we used 1.5% isoflurane to maintain anesthesia and the following experimental protocol: 5 min of baseline with balance room air, 4 min of 5% CO2 and balance room air, followed by 6 min of the initial isoflurane-room air condition. We extracted the time course of SFDI and LSI data using a semicircular ROI on the left hemisphere (radius of 175 pixels, 1.4  mm). The ROI was chosen with its position along the midpoint between bregma and lambda and an offset of 100 pixels (0.8  mm) from the midline. The representative CBF maps of Db and SFI with the chosen ROI can be seen in Fig. 7(a). We then normalized Db and SFI measurements to time t=0 and obtained a percentage ΔCBF. Due to normalization, β correction is not required for SFI measurements. The time courses were then filtered using a moving average of 200 data points, approximating a 4-s period. In this experiment, as expected we observed a hyperemic response during the hypercapnic period [Fig. 7(b)]. However, there was a slight difference between the filtered SFI and Db time courses [Fig. 7(c)]. This difference ranged from 1.5 to 0.5% during the experimental period, corresponding to the range of changes of 0.005  mm1 (7%) in μa and 0.025  mm1 (2%) in μs [Figs. 7(c) and 7(d)]. This decrease in μa and μs resulted in a minimal change in CBF dynamics between Db and SFI measurements. In fact, Eq. (9) predicts that K2 remains constant with a defined Db if the changes of μa and μs happen with the same multiplicative factor. For example, if μa and μs both increase by 50%, K2 stays the same. Intuitively, the loss of correlation due to scattering cancels the increase of correlation from the loss of the traveling photon due to absorption. However, in the case where only one of the two optical properties remains constant while the other is changing, K2 will change accordingly. A simulation result of this phenomenon was shown in Fig. S1 in the Supplementary Material.

Fig. 7

CBF dynamics slightly differentiate between Db and SFI measurements for a mouse undergoing a hypercapnia gas challenge. (a) Representative images of Db and SFI during hypercapnia with the chosen semicircular ROI (black) on the left hemisphere. Scale bars are 1 mm. b) CBF measurements normalized to t=0 where both SFI and Db analysis was performed. Median data are chosen from the left hemisphere to extract the time course of SFI and Db values. The data is further filtered using a moving filter of 200 frames (4  s). (c) A difference of 1.5 to 0.5% between the normalized Db and SFI curves is observed throughout the gas challenge experiment. (d) Optical properties derived at 633 nm showing a slight decrease in both μa and μs across the time course.

NPh_10_4_045001_f007.png

6.2.

Inter-Animal CBF Measurement at Resting State

We measured the resting-state CBF of a cohort of mice (n=13, B6SJL background, 15 months old). Anesthesia was initiated using 4% isoflurane (balance oxygen) within an enclosed chamber and the animals were maintained at 2% isoflurane (balance room air) at 1  L/min flow rate. We then performed an acute intact-skull cranial window surgical procedure to gain access to the cortical layer of the brain.38 The scalp was shaved, depilated, and sterilized with alcohol and povidone-iodine wipes. We then performed a midline incision, retracted the scalp, and removed the fascial layer to expose the relatively transparent skull. The window was then filled with saline and a coverslip was applied to create a translucent skull for imaging the cortical surface.

Resting-state data were taken for one minute at 2% isoflurane and a common semi-circular ROI (radius of 175 pixels, 1.4  mm) on the left hemisphere was chosen [Fig. 8(a)]. The ROI was chosen with its position along the midpoint between bregma and lambda and an offset of 100 pixels (0.8  mm) from the midline. To calculate SFI and Db, we chose the median value within the ROI to generate time course data for all measured parameters (e.g., K, μa, and μs). We applied the processing methods discussed above to generate optical properties at 633 nm and convert SFI to Db [Fig. 8(a)]. Here, we only applied the processing methods on the 1-min time-course data and not on every pixel within the FOV.

Fig. 8

Baseline CBF measurements from 13 mice (2% isoflurane for 1 min) highlighted the effect of optical properties on the interpretation of CBF measurements using LSI. (a) Median values of data obtained from the left hemisphere were used to generate time course data for the 1-min imaging period. Both the pulsatile CBF waveform and breathing-related motion artifacts could be resolved. On the representative image on the right, the scale bar is 1 mm. (b) Mean values of the SFI time course during each imaging period were then used to rank the animals in ascending order (blue to red circles). Animal IDs (1 to 13) were then assigned based on this ranking. Here, the Db values differed from SFI in ranking for nine out of 13 animals.

NPh_10_4_045001_f008.png

To assess the impact of optical property correction on CBF, we calculated the mean of the time courses of SFI and Db. An initial ranking of resting-state CBF was generated using SFI values [Fig. 8(b), blue to red for lowest to highest SFI measurements]. The animal IDs were also assigned using this ordering (i.e., animal 1 was the one with the lowest SFI). To facilitate a more direct comparison of SFI and Db, we normalized the individual SFI and Db values for all animals to those of animal 1. Due to normalization, β correction is not required for SFI measurements. Using this comparison method, we observed that the ranking of CBF using Db differs from that using SFI. Specifically, 9 out of 13 animals showed a change in the ranking [Fig. 8(b)], thus highlighting the potential effect of optical properties on speckle-based measurements of CBF.

Figure 9(a) showed representative normalized CBF images that were generated using a pixel-wise correction algorithm. All pixels in the Db and SFI images were normalized to the respective values of the center pixel in each image. We also calculated a percentage difference between normalized Db and SFI maps [Fig. 9(a)] and noted higher CBF values in larger vessels (red arrows) in the Db image. Currently, this pixel-by-pixel method of correction is currently not applied to time-course analysis due to the long processing time for images with such high pixel resolution (1.44 MP). We further investigated if these spatial differences would also yield a change in temporal dynamics. On the same representative animal, we used the percentage difference map to define three groups of pixels within the chosen ROIs mentioned in Fig. 8. These three groups correspond to the pixels that showed a decrease [<5%, Fig. 9(b) blue], minimal change [5 to 5%, Fig. 9(b) green], or increase [>5%, Fig. 9(b) red] in the spatial percentage difference map. We then calculated rCBF as relative CBF from t=0 for Db and SFI time courses using the median value of all the pixels within each respective region. As shown in Fig. 9(b) below, the Db and SFI time courses showed similar dynamics at the resting state for all regions. The power spectra, not shown here, did not yield notable differences in the pulsatile signal.

Fig. 9

On one representative animal, Db, when compared to SFI map, showed a spatial difference in CBF measurements but not in temporal dynamics. (a) Representative normalized maps of Db and SFI highlight the spatial difference in estimated CBF. Individual pixels in each image were normalized relative to the respective center pixel (*). A percentage difference map was generated using %Difference=(Db,normSFInorm)/SFInorm×100. Db showed increased blood flow in large vessels (red arrows) compared with SFI. Scale bars are 1 mm. (b) Within the chosen semicircular ROI, temporal dynamics (i.e., rCBF) did not show notable differences between SFI and Db measurements among regions with decreasing (<5%), minimally changing (5 to 5%), and increasing (>5%) spatial differences. The power spectrum of both Db and SFI (not shown here) showed negligible differences in the pulsatile waveforms.

NPh_10_4_045001_f009.png

7.

Discussion

This work outlined the design of a QHI system that combined LSI and SFDI to achieve quantitative CBF measurements independent of optical properties. Db measurements of one mouse under hypercapnia showed a slight difference from SFI one. Additionally, the in vivo resting-state measurements showed a change in the ranking of CBF under anesthesia after correcting LSI measurements using optical properties obtained from SFDI. This discrepancy may impact statistical analyses of proposed hypotheses in experiments that involve comparing the CBF of several groups (e.g., disease-state versus wild-type). The results further illustrate the need to correct for the effects of optical properties on speckle contrast measurements.

In LSI literature, relative CBF (i.e., normalized to baseline or t=0) is the most commonly used quantitation method.5,9,13 Attempts to enable further quantitative measurements have focused on utilizing the dynamics at various exposure times, also known as multi-exposure speckle imaging (MESI), to account for the variability of the optical setup (i.e., β) and the presence of static scatterers.3941 However, the theory of MESI only focused on fitting the shape of the contrast-versus-exposure-time functions and did not account directly for optical properties. Meanwhile, our method utilized the solution of the correlation diffusion equation 42 to relate speckle contrast K to μa, μs, and Db. This method allows for directly estimating Db using the measured LSI and SFDI data. In addition to neuroimaging, QHI can be applied to other applications, such as burn assessment and peripheral blood flow monitoring.17,43,44 In these applications, QHI enables quantitative blood flow measurements independent from potential confounding factors such as pigmentation (i.e., variation in absorption) and collagen denaturation (i.e., changes in scattering) in burn wounds. Despite its advantages, the system is not free from limitations, as discussed below.

7.1.

Current Hardware Limitations

The system has a 50 Hz raw frame rate at 1200×1200  pixel resolution of a co-aligned 10  mm×10  mm FOV. This frame rate enabled an effective 50 Hz and 2.8 Hz frame rate for LSI and SFDI arms, respectively. Using in vitro phantoms, for sensitivity and accuracy measurements, we showed that QHI can detect differences in LSI flow speed from 0 to 4  mm/s while also achieving a strong linear correlation of optical properties (i.e., μa and μs) measurements with a commercial SFDI system (R2>0.95). We observed a slight underestimation of μa and a slight overestimation of μs between QHI and the commercial system.

The openSFDI guide suggested that the exposure time of the camera should be multiples of the DMD’s exposure time to avoid demodulation artifacts.26 However, despite matching our camera’s exposure time to the DMD’s exposure time (i.e., 10 ms), we did not notice significant demodulation artifacts after performing phantom calibration. We achieved this by ensuring that the trigger signal from the DMD to the camera was sent 11  μs before the pattern display to match the minimum 11  μs trigger delay time of the camera used in the current build. And as noted above, the fast frame rate at high pixel resolution may result in dropped frames during acquisition. This could be due to the bottleneck of the writing speeds of the internal SSDs. A potential solution would be saving LSI and SFDI data into separate SSDs to maximize the writing speed.

The flow phantom validation experiment showed that the system is sensitive up to 4  mm/s of directed flow speed for 1% intralipid solution. In live mouse brains, researchers have used fluorescence microscopy to measure absolute flow speed on a vascular level, where the speed of moving red blood cells ranged from 0 to 1  mm/s in capillaries, 1 to 10  mm/s in venules, and 5 to 20  mm/s in arterioles.45 The dynamic range of our system presents a limitation in quantitative measurements of venules and arterioles. Thus, we recommend choosing the median value of a large ROI during ROI selection and processing, such as the chosen semicircles as shown above, as the quantitative factor. This median value typically represents the parenchymal regions of the brain, where capillaries dominate.

Also, the described method of deriving μa and μs at the LSI wavelength from values obtained using SFDI can be prone to errors. This would then affect the deduced Db values. One solution to this problem is to perform LSI and SFDI sequentially, where LSI frames are interlaced with SFDI frames. However, this method would significantly increase the time required for the acquisition or reduce the exposure time to accommodate for a faster acquisition speed. Decreasing exposure time can increase the detectable noise levels and also affect the spatial speckle contrast measurements when the exposure time starts to approximate the correlation time (i.e., the flow would appear more static at a faster exposure time). Additionally, the sequential triggering of SFDI frames limits the SFDI frame rate to one for every 18 LSI frames. This leads to a temporal mismatch in the correction method, where 18 LSI frames, which can span the pulsatile signal, have the same optical properties. This limitation poses a potential explanation for the lack of temporal dissimilarity between Db and SFI, as shown in Fig. 9(b) above. Furthermore, the mismatched wavelengths raised the question of whether the two modalities probe similar tissue volumes. Here, we calculated the estimated effective depth penetration of the SFDI system using the derived optical properties at 633 nm for the 13 mice measured at resting state. The depth penetration ranges from 0.5 to 2  mm depending on the spatial frequencies (Table S5 in the Supplementary Material). It is acknowledged that the penetration depth for SFDI falls within this specific range. This range overlaps with the typical penetration depth of an LSI measurement.46 However, the exact probing volumes remain dissimilar. To address these concerns, there have recently been reports of utilizing speckle patterns as a function of spatial frequency to extract optical properties.4749 However, to the best of our knowledge, a method to decouple the effect of blood flow from optical properties on speckle contrast measurements has not been reported.

7.2.

Limitations of the Homogeneous, Semi-Infinite Tissue Assumption

A limitation of the current processing method is the employment of a semi-infinite homogeneous model for SFDI data. Furthermore, the solution from Mazhar et al. was derived for the same tissue geometry.15 Thus, the current correction scheme does not include the effects of volumetric heterogeneities such as the skull on the measurements. To improve the accuracy of QHI measurements of CBF, the skull should be modeled as a scattering layer with minimal absorption and a fixed Db value that resembles that of a static phantom.

Furthermore, it should be noted that the equations used in this publication were derived from dynamic light scattering theory with an assumption that the photons encounter multiple scattering events before being detected. This is typically true for the parenchymal region where small capillaries act as the main source of dynamics, resulting in a more diffusive behavior typically observed from multiple and unordered scattered photons.50 However, it was also shown that larger vessels follow a more single and ordered scattered regime. Finally, the authors also noted a mixed dynamic (i.e., multiply and ordered scattering, and single and unordered scattering) in the immediate regions surrounding these vessels. Thus, our assumption of multiple and unorder scattering potentially results in the mischaracterization of the true dynamics of the larger vessels and their surrounding regions. Nevertheless, this assumption should be appropriate for the parenchymal region, where capillaries dominate.

7.3.

Understanding the Diffusion Coefficient and its Relevance in CBF Measurements

The diffusion coefficient Db is used to describe the mean squared displacement of the moving particles using the Brownian diffusion model. In diffuse correlation spectroscopy, a fiber-based analog of LSI, Db has been used to quantify CBF with assumed optical properties.5153 Researchers have found that this Brownian model better fits the electric field autocorrelation curves compared to a random flow model when measuring the movement of red blood cells inside tissues. A potential explanation is that, in tissue, red blood cells do not simply move from one position to another but also experience shear stresses, causing them to “roll and tumble” in the vasculature.34 This is particularly true for the parenchymal region of the brain, where the imaging system is mainly probing small capillaries. From a macroscopic point of view, this highly randomized motion exhibits behaviors similar to Brownian motion. However, the typical values of Db for blood flow are magnitudes higher than that of pure thermally induced Brownian motion. As for large vessels, as discussed above, a more rigorous model should be utilized to correctly account for the more single-directional motion of the red blood cells.

The unit of Db is area over time (e.g., mm2/s). This unit differs from the clinically used units of CBF such as perfusion rate per tissue volume per time (e.g., mL/g/min). However, these clinical measurements are often obtained using a volumetric scanner, such as MRI or PET. This is different from an en-face widefield optical imaging measurement, such as LSI. While LSI can probe CBF at deeper brain regions, this information is usually integrated into the superficial cortical measurements. Hence, each LSI pixel represents mainly the motion detected at the surface and partial subsurface contributions. Both contribute to the blurriness of the speckle pattern at the analyzed pixel. Thus, the unit of Db (i.e., area per time) is appropriate to measure such dynamics. In fact, it has been shown that LSI measurements of directed motions do not measure volumetric flow directly but a combination of the channel size and the speed of the particles (i.e., mostly single-directional area flux).54 However, the work in that study specifically investigated the flow dynamics of large vessels, ranging from 20 to 100  μm for in-vivo and 65 to 175  μm for in-vitro demonstrations and is not directly applicable to measurements of the parenchymal regions as performed here. Future work is warranted to fully understand which model works best for each selected vascular region of the cortex.

Finally, we are interested in leveraging Db as a non-unitless measurement of CBF in calculating the cerebral metabolic rate of oxygen (CMRO2). In studies combining LSI with spectroscopic imaging, relative measurements of rCMRO2 can be estimated with rCBF and hemoglobin concentrations using certain assumptions.5,13,27 A new model utilizing Db as a quantitative CBF surrogate will allow for a direct calculation of CMRO2 instead of a relative value. However, this new method requires changes to the typical CMRO2 equation (i.e., Fick’s law) due to the mismatched dimensions and will be the subject of future work.

8.

Conclusion

In this work, we reiterate that the interpretation of LSI measurements can be confounded by changes in optical properties. To enable measurement of absorption, scattering, and speckle contrast, we developed and described in detail our QHI system for wide-field, small-animal neuroimaging to achieve temporal and spatial synchronization for LSI and SFDI. We demonstrated that QHI has sufficient sensitivity and precision to enable reproducible and accurate longitudinal measurements. Our in-vivo data suggest that CBF is most accurately described with the combination of LSI and SFDI, which has implications for comparing CBF measurements among subjects and within a given subject over time. Future work will involve the use of QHI in longitudinal studies of CBF associated with neurodegenerative disease progression.

Disclosures

Any opinions, findings, and conclusions or recommendations expressed in this material are those of the authors and do not necessarily reflect the views of the United States Air Force, the National Institutes of Health (NIH), the BrightFocus Foundation, the National Science Foundation, the National Research Foundation of Korea, or the Arnold and Mabel Beckman Foundation. Dr. Durkin is a co-founder of Modulim Inc., Irvine, California, United States, which developed the Reflect RS. However, Dr. Durkin does not participate in the operation or management of Modulim. He is in compliance with the conflict of interest policy of the University of California, Irvine, and that of the NIH (policies revisited annually). The other authors have no financial interests or commercial associations that might pose or create a conflict of interest with the information presented in this article.

Code, Data, and Materials Availability

All code, data, and materials used in this study are available by reasonable request to the corresponding author.

Acknowledgments

We thank Dr. Matthew Applegate for valuable conversations on the openSFDI platform. We would like to thank the Arnold and Mabel Beckman Foundation, the National Institutes of Health [Grant Nos. R21AG066000 and 5TL1TR001415 (CC)], the National Science Foundation Graduate Research Fellowship Program [Grant No. DGE-1839285 (TP)], the donors of the Alzheimer’s Disease Research (Grant No. A2022005F), a program of the BrightFocus Foundation, and the National Research Foundation of Korea (Grant No. NRF-2018K1A4A3A02060572) for support of this research. In addition, this material is based, in part, upon technology development supported by the U.S. Air Force Office of Scientific Research (Grant No. FA9550-20-1-0052).

References

1. 

K. Kisler et al., “Cerebral blood flow regulation and neurovascular dysfunction in Alzheimer disease,” Nat. Rev. Neurosci., 18 419 –434 https://doi.org/10.1038/nrn.2017.48 NRNAAN 1471-003X (2017). Google Scholar

2. 

B. V. Zlokovic, “Neurovascular pathways to neurodegeneration in Alzheimer’s disease and other disorders.,” Nat. Rev. Neurosci., 12 723 –738 https://doi.org/10.1038/nrn3114 NRNAAN 1471-003X (2011). Google Scholar

3. 

M. D. Sweeney et al., “The role of brain vasculature in neurodegenerative disorders,” Nat. Neurosci., 21 1318 –1331 https://doi.org/10.1038/s41593-018-0234-x NANEFN 1097-6256 (2018). Google Scholar

4. 

H. Girouard and C. Iadecola, “Neurovascular coupling in the normal brain and in hypertension, stroke, and Alzheimer disease,” J. Appl. Physiol., 100 328 –335 https://doi.org/10.1152/japplphysiol.00966.2005 (2006). Google Scholar

5. 

A. K. Dunn et al., “Spatial extent of oxygen metabolism and hemodynamic changes during functional activation of the rat somatosensory cortex,” Neuroimage, 27 279 –290 https://doi.org/10.1016/j.neuroimage.2005.04.024 NEIMEF 1053-8119 (2005). Google Scholar

6. 

A. K. Dunn et al., “Simultaneous imaging of total cerebral hemoglobin concentration, oxygenation, and blood flow during functional activation,” Opt. Lett., 28 28 https://doi.org/10.1364/OL.28.000028 OPLEDP 0146-9592 (2003). Google Scholar

7. 

A. Cho et al., “Laser speckle contrast imaging for measuring cerebral blood flow changes caused by electrical sensory stimulation,” J. Opt. Soc. Korea, 20 88 –93 https://doi.org/10.3807/JOSK.2016.20.1.088 1226-4776 (2016). Google Scholar

8. 

C. Crouzet et al., “Intrinsic, widefield optical imaging of hemodynamics in rodent models of Alzheimer’s disease and neurological injury,” Neurophotonics, 10 020601 https://doi.org/10.1117/1.NPh.10.2.020601 (2023). Google Scholar

9. 

D. A. Boas and A. K. Dunn, “Laser speckle contrast imaging in biomedical optics,” J. Biomed. Opt., 15 011109 https://doi.org/10.1117/1.3285504 JBOPFO 1083-3668 (2010). Google Scholar

10. 

D. Malonek and A. Grinvald, “Interactions between electrical activity and cortical microcirculation revealed by imaging spectroscopy: implications for functional brain mapping,” Science, 272 551 –554 https://doi.org/10.1126/science.272.5261.551 SCIEAS 0036-8075 (1996). Google Scholar

11. 

D. Malonek et al., “Vascular imprints of neuronal activity: Relationships between the dynamics of cortical blood flow, oxygenation, and volume changes following sensory stimulation,” Proc. Natl. Acad. Sci. U. S. A., 94 14826 –14831 https://doi.org/10.1073/pnas.94.26.14826 (1997). Google Scholar

12. 

J. Mayhew et al., “Increased oxygen consumption following activation of brain: Theoretical footnotes using spectroscopic data from barrel cortex,” Neuroimage, 13 975 –987 https://doi.org/10.1006/nimg.2001.0807 NEIMEF 1053-8119 (2001). Google Scholar

13. 

P. B. Jones et al., “Simultaneous multispectral reflectance imaging and laser speckle flowmetry of cerebral blood flow and oxygen metabolism in focal cerebral ischemia,” J. Biomed. Opt., 13 044007 https://doi.org/10.1117/1.2950312 JBOPFO 1083-3668 (2008). Google Scholar

14. 

H. Lu et al., “Single-trial estimation of the cerebral metabolic rate of oxygen with imaging photoplethysmography and laser speckle contrast imaging,” Opt. Lett., 40 1193 https://doi.org/10.1364/OL.40.001193 OPLEDP 0146-9592 (2015). Google Scholar

15. 

A. Mazhar et al., “Laser speckle imaging in the spatial frequency domain.,” Biomed. Opt. Express, 2 1553 –1563 https://doi.org/10.1364/BOE.2.001553 BOEICL 2156-7085 (2011). Google Scholar

16. 

R. H. Wilson et al., “High-speed quantitative optical imaging of absolute metabolism in the rat cortex,” Neurophotonics, 8 025001 https://doi.org/10.1117/1.NPh.8.2.025001 (2021). Google Scholar

17. 

B. S. Lertsakdadet et al., “Assessing multimodal optical imaging of perfusion in burn wounds,” Burns, 48 (4), 799 –807 https://doi.org/10.1016/j.burns.2021.08.026 BURND8 0305-4179 (2021). Google Scholar

18. 

D. J. Cuccia et al., “Modulated imaging: quantitative analysis and tomography of turbid media in the spatial-frequency domain,” Opt. Lett., 30 1354 https://doi.org/10.1364/OL.30.001354 OPLEDP 0146-9592 (2005). Google Scholar

19. 

D. J. Cuccia et al., “Quantitation and mapping of tissue optical properties using modulated imaging,” J. Biomed. Opt., 14 024012 https://doi.org/10.1117/1.3088140 JBOPFO 1083-3668 (2009). Google Scholar

20. 

A. J. Lin et al., “Spatial frequency domain imaging of intrinsic optical property contrast in a mouse model of alzheimer’s disease,” Ann. Biomed. Eng., 39 1349 –1357 https://doi.org/10.1007/s10439-011-0269-6 ABMECF 0090-6964 (2011). Google Scholar

21. 

A. J. Lin et al., “Optical imaging in an Alzheimer’s mouse model reveals amyloid-β-dependent vascular impairment,” Neurophotonics, 1 011005 https://doi.org/10.1117/1.NPh.1.1.011005 (2014). Google Scholar

22. 

R. H. Wilson et al., “High-speed spatial frequency domain imaging of rat cortex detects dynamic optical and physiological properties following cardiac arrest and resuscitation,” Neurophotonics, 4 045008 https://doi.org/10.1117/1.NPh.4.4.045008 (2017). Google Scholar

23. 

R. H. Wilson et al., “Spreading depolarization and repolarization during cardiac arrest as an ultra-early marker of neurological recovery in a preclinical model,” (2019). Google Scholar

24. 

R. H. Wilson et al., “Cerebral perfusion and metabolism coupling during a critical time window provides rapid assessment of cardiac arrest severity and prognosis in a preclinical model,” (2019). Google Scholar

25. 

M. B. Applegate et al., “OPEN SFDI, Step-by-step instructions for building an SFDI system,” www.opensfdi.org (September 2023). Google Scholar

26. 

M. B. Applegate et al., “OpenSFDI: an open-source guide for constructing a spatial frequency domain imaging system,” J. Biomed. Opt., 25 016002 https://doi.org/10.1117/1.JBO.25.1.016002 JBOPFO 1083-3668 (2020). Google Scholar

27. 

J. Mayhew et al., “Spectroscopic analysis of neural activity in brain: Increased oxygen consumption following activation of barrel cortex,” Neuroimage, 12 664 –675 https://doi.org/10.1006/nimg.2000.0656 NEIMEF 1053-8119 (2000). Google Scholar

28. 

S. J. Kirkpatrick, D. D. Duncan and E. M. Wells-Gray, “Detrimental effects of speckle-pixel size matching in laser speckle contrast imaging,” Opt. Lett., 33 2886 https://doi.org/10.1364/OL.33.002886 OPLEDP 0146-9592 (2008). Google Scholar

29. 

K. Elliott et al., Keynote Flexlight LC4500: LC4500-RGB-EKT Electronics Kit User ’s Guide, 1 –48 Keynote Photonics( (2016). Google Scholar

30. 

M. Martinelli et al., “Analysis of single Monte Carlo methods for prediction of reflectance from turbid media,” Opt. Express, 19 19627 https://doi.org/10.1364/OE.19.019627 OPEXFF 1094-4087 (2011). Google Scholar

31. 

B. Choi et al., “Linear response range characterization and in vivo application of laser speckle imaging of blood flow dynamics,” J. Biomed. Opt., 11 041129 https://doi.org/10.1117/1.2341196 JBOPFO 1083-3668 (2006). Google Scholar

32. 

J. C. Ramirez-San-Juan et al., “Impact of velocity distribution assumption on simplified laser speckle imaging equation,” Opt. Express, 16 3197 https://doi.org/10.1364/OE.16.003197 OPEXFF 1094-4087 (2008). Google Scholar

33. 

T. Durduran and A. G. Yodh, “Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement,” Neuroimage, 85 51 –63 https://doi.org/10.1016/j.neuroimage.2013.06.017 NEIMEF 1053-8119 (2014). Google Scholar

34. 

T. Durduran et al., “Diffuse optics for tissue monitoring and tomography,” Rep. Prog. Phys., 73 076701 https://doi.org/10.1088/0034-4885/73/7/076701 RPPHAG 0034-4885 (2010). Google Scholar

35. 

T. D. O’Sullivan et al., “Diffuse optical imaging using spatially and temporally modulated light,” J. Biomed. Opt., 17 0713111 https://doi.org/10.1117/1.JBO.17.7.071311 JBOPFO 1083-3668 (2012). Google Scholar

36. 

C. Regan et al., “Fiber-based laser speckle imaging for the detection of pulsatile flow,” Lasers Surg. Med., 47 520 –525 https://doi.org/10.1002/lsm.22370 LSMEDI 0196-8092 (2015). Google Scholar

37. 

J. C. Ramirez-San-Juan et al., “Spatial versus temporal laser speckle contrast analyses in the presence of static optical scatterers,” J. Biomed. Opt., 19 106009 https://doi.org/10.1117/1.JBO.19.10.106009 JBOPFO 1083-3668 (2014). Google Scholar

38. 

B. Choi et al., “Cerebral blood flow in chronic kidney disease,” J. Stroke Cerebrovasc. Dis., 30 105702 https://doi.org/10.1016/j.jstrokecerebrovasdis.2021.105702 (2021). Google Scholar

39. 

A. B. Parthasarathy et al., “Robust flow measurement with multi-exposure speckle imaging,” Opt. Express, 16 1975 –1989 https://doi.org/10.1364/OE.16.001975 OPEXFF 1094-4087 (2008). Google Scholar

40. 

L. M. Richards et al., “Intraoperative multi-exposure speckle imaging of cerebral blood flow,” J. Cereb. Blood Flow Metab., 37 3097 –3109 https://doi.org/10.1177/0271678X16686987 (2017). Google Scholar

41. 

S. M. Shams Kazmi, R. K. Wu and A. K. Dunn, “Evaluating multi-exposure speckle imaging estimates of absolute autocorrelation times,” Opt. Lett., 40 3643 –3646 https://doi.org/10.1364/OL.40.003643 OPLEDP 0146-9592 (2015). Google Scholar

42. 

D. A. Boas, Diffuse Photon Probes of Structural and Dynamical Properties of Turbid Media: Theory and Biomedical Application, 260 University of Pennsylvania ProQuest Dissertations Publishing( (1996). Google Scholar

43. 

C. E. Dunn et al., “Comparison of speckleplethysmographic (SPG) and photoplethysmographic (PPG) imaging by Monte Carlo simulations and in vivo measurements,” Biomed. Opt. Express, 9 4306 https://doi.org/10.1364/BOE.9.004306 BOEICL 2156-7085 (2018). Google Scholar

44. 

M. Ghijsen et al., “Wearable speckle plethysmography (SPG) for characterizing microvascular flow and resistance,” Biomed. Opt. Express, 9 3937 https://doi.org/10.1364/BOE.9.003937 BOEICL 2156-7085 (2018). Google Scholar

45. 

X. Han et al., “In vivo two-photon imaging reveals acute cerebral vascular spasm and microthrombosis after mild traumatic brain injury in mice,” Front. Neurosci., 14 210 https://doi.org/10.3389/fnins.2020.00210 1662-453X (2020). Google Scholar

46. 

M. A. Davis, S. M. S. Kazmi and A. K. Dunn, “Imaging depth and multiple scattering in laser speckle contrast imaging,” J. Biomed. Opt., 19 086001 https://doi.org/10.1117/1.JBO.19.8.086001 JBOPFO 1083-3668 (2014). Google Scholar

47. 

P. Jain and S. E. Sarma, “Measuring light transport properties using speckle patterns as structured illumination,” Sci. Rep., 9 11157 https://doi.org/10.1038/s41598-019-47256-8 SRCEC3 2045-2322 (2019). Google Scholar

48. 

M. T. Chen, T. L. Bobrow and N. J. Durr, “Towards SFDI endoscopy with structured illumination from randomized speckle patterns,” in Proc. SPIE, 116310Y (2021). https://doi.org/10.1117/12.2578940 Google Scholar

49. 

M. T. Chen, M. Papadakis and N. J. Durr, “Speckle illumination SFDI for projector-free optical property mapping,” Opt. Lett., 46 673 https://doi.org/10.1364/OL.411187 OPLEDP 0146-9592 (2021). Google Scholar

50. 

D. D. Postnov et al., “Dynamic light scattering imaging,” Sci. Adv., 6 1 –10 https://doi.org/10.1126/sciadv.abc4628 STAMCV 1468-6996 (2020). Google Scholar

51. 

E. M. Buckley et al., “Cerebral hemodynamics in preterm infants during positional intervention measured with diffuse correlation spectroscopy and transcranial Doppler ultrasound,” Opt. Express, 17 12571 https://doi.org/10.1364/OE.17.012571 OPEXFF 1094-4087 (2009). Google Scholar

52. 

H. Zhao, E. Sathialingam and E. M. Buckley, “Accuracy of diffuse correlation spectroscopy measurements of cerebral blood flow when using a three-layer analytical model,” Biomed. Opt. Express, 12 7149 –7161 https://doi.org/10.1364/BOE.438303 BOEICL 2156-7085 (2021). Google Scholar

53. 

R. C. Mesquita et al., “Direct measurement of tissue blood flow and metabolism with diffuse optics,” Philos. Trans. R. Soc. A Math. Phys. Eng. Sci., 369 4390 –4406 https://doi.org/10.1098/rsta.2011.0232 (2011). Google Scholar

54. 

S. M. S. Kazmi et al., “Flux or speed? Examining speckle contrast imaging of vascular flows,” Biomed. Opt. Express, 6 2588 https://doi.org/10.1364/BOE.6.002588 BOEICL 2156-7085 (2015). Google Scholar

Biography

Thinh Phan is a PhD student of biomedical engineering at the University of California, Irvine, USA. He is currently advised by Professor Bernard Choi at the Beckman Laser Institute and Medical Clinic. His research is focused on developing a multimodal widefield optical imaging system to achieve quantitative measurements of cerebral hemodynamics to study rodent models with neurodegenerative diseases and cancer therapies.

Christian Crouzet is a project scientist at the Beckman Laser Institute at the University of California, Irvine, USA. His research is focused on using optical technologies to study rodent models of neurodegeneration and neurological injury across spatial scales.

Gordon T. Kennedy has a BSc degree in applied physics from Strathclyde University and a PhD in laser physics from the University of St. Andrews, Scotland. He was a postdoc at St Andrews University (1993 to 1997) then senior scientist at Diomed (1997 to 2002) before returning to academia as a postdoc at Imperial College, London, from 2002 to 2013. Since 2014, he has been a project scientist at the Beckman Laser Institute where his research interests include optical characterization of biological tissue.

Anthony J. Durkin is the director of the Spatial Frequency Domain Imaging lab at the Beckman Laser Institute, UC Irvine. His research is focused on development and application of optical spectroscopic and quantitative wide-field imaging techniques to characterize in-vivo tissue. He holds a PhD in biomedical engineering from the UT Austin with emphasis on biomedical optics and spectroscopy. He is a senior member of the JBO editorial board and a fellow of SPIE.

Bernard Choi is a professor of biomedical engineering and surgery at the University of California, Irvine, and the acting director of the Beckman Laser Institute and Medical Clinic. His group is focused on the development of technologies and methods to study microvascular structure and function, with a recent neurological focus on the study of age- and disease-related changes to cerebral blood flow and the blood–brain barrier.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Thinh Phan, Christian Crouzet, Gordon T. Kennedy, Anthony J. Durkin, and Bernard Choi "Quantitative hemodynamic imaging: a method to correct the effects of optical properties on laser speckle imaging," Neurophotonics 10(4), 045001 (3 October 2023). https://doi.org/10.1117/1.NPh.10.4.045001
Received: 8 May 2023; Accepted: 12 September 2023; Published: 3 October 2023