Open Access
4 March 2024 Development of a near-infrared wide-field integral field unit by ultra-precision diamond cutting
Author Affiliations +
Abstract

Integral field spectroscopy (IFS) is an observational method for obtaining spatially resolved spectra over a specific field of view (FoV) in a single exposure. In recent years, near-infrared IFS has gained importance in observing objects with strong dust attenuation or at a high redshift. One limitation of existing near-infrared IFS instruments is their relatively small FoV, less than 100 arcsec2, compared with optical instruments. Therefore, we developed a near-infrared (0.9 to 2.5 μm) image-slicer type integral field unit (IFU) with a larger FoV of 13.5×10.4 arcsec2 by matching a slice width to a typical seeing size of 0.4 arcsec. The IFU has a compact optical design utilizing off-axis ellipsoidal mirrors to reduce aberrations. Complex optical elements were fabricated using an ultra-precision cutting machine to achieve root mean square surface roughness of less than 10 nm and a P-V shape error of less than 300 nm. The ultra-precision machining can also simplify the alignment procedures. The on-sky performance evaluation confirmed that the image quality and the throughput of the IFU were as designed. In conclusion, we successfully developed a compact IFU utilizing an ultra-precision cutting technique, almost fulfilling the requirements.

1.

Introduction

Integral field spectroscopy (IFS) is an observational method for obtaining spatially resolved spectra over a specific field of view (FoV) in a single exposure. It can obtain spectral information of spatially extended objects more efficiently than conventional scanning methods, such as slit scanning, scanning Fabry–Perot interferometry, and imaging Fourier transform spectroscopy. It also allows for obtaining homogeneous data over the FoV because it is not affected by temporal variations in atmospheric and instrument conditions. Therefore, many IFS instruments have been developed for optical and infrared astronomy.110

In particular, near-infrared IFS is important for studying nearby objects with strong dust attenuation or obtaining rest-frame UV and optical spectra of high-redshift objects. However, as the FoVs of existing near-infrared IFS instruments for 8 to 10 m telescopes are smaller than 100  arcsec2,510 various science cases that require a wider FoV, such as the observation of extended objects or wide-field surveys, are restricted. This is because they are optimized for observation with high spatial resolution using adaptive optics.

Therefore, we developed a near-infrared image-slicer type integral field unit (IFU) with a large FoV for an existing near-infrared instrument, simultaneous-color wide-field infrared multi-object spectrograph (SWIMS).1116 SWIMS is one of the first generation instruments for the University of Tokyo Atacama Observatory (TAO) 6.5 m telescope, which is being constructed by the Institute of Astronomy, the University of Tokyo, on the summit of Cerro Chajanantor in the Atacama Desert, Chile.1722 SWIMS is capable of simultaneous two-color imaging with a large FoV of ϕ9.6  arcmin and multi-object slit spectroscopy covering the entire 0.9 to 2.5  μm in a single exposure.

In this paper, we report the development and performance evaluation of the IFU for SWIMS, which is named SWIMS-IFU. A basic concept of the IFU is explained in Sec. 2, and its optical design is presented in Sec. 3. Section 4 summarizes the mirror fabrication and assembly using ultra-precision cutting. After showing the performance evaluation by laboratory tests in Sec. 5 and an engineering observation in Sec. 6, we summarize the overall development in Sec. 7.

2.

Concept

2.1.

Overview of SWIMS

The internal structure of SWIMS is shown in Fig. 1. Imaging with a large FoV and spectroscopy with a wide simultaneous wavelength coverage of 0.9 to 2.5  μm are realized by dividing collimated light into two wavelength ranges of 0.9 to 1.45  μm (blue arm) and 1.45 to 2.5  μm (red arm) by a dichroic mirror and re-imaging them simultaneously on focal plane array detectors of each arm. Switching between imaging and spectroscopy modes is done by changing the filters and grisms in the filter wheels. Two 2048×2048 format HAWAII-2RG (H2RG) array detectors for each focal plane cover a 6.7×3.3  arcsec2 FoV, and the goal is to increase them to four in a 2×2 matrix to cover the whole FoV delivered by the optics. Spectra are dispersed over the two detectors to cover the wide wavelength range.

Fig. 1

Internal structure of SWIMS. A slit mask or the IFU is placed on the telescope focal plane indicated by the arrow.

JATIS_10_1_015004_f001.png

For multi-object spectroscopy, a slit mask is placed at a telescope focal plane by a multi-object spectroscopy unit (MOSU).23 The MOSU consists of a carousel that stores up to 22 slit masks under vacuum and cryogenic conditions and a robotic arm that handles the slit masks.

SWIMS will be installed at the Nasmyth focus of the TAO 6.5 m telescope, which has a focal ratio (F-ratio) of 12.2. This is the same as the F-ratio of the Cassegrain focus of the Subaru telescope. SWIMS was carried to the Subaru Telescope to evaluate and optimize its on-sky performance before the completion of the TAO 6.5 m telescope. Engineering observations were conducted between 2018 and 2020. Thereafter, the instrument was operated as a PI-type instrument from 2021 to 2022.

2.2.

Concept of the SWIMS-IFU

The SWIMS-IFU is an image-slicer type IFU. It should be stored and operated in a cryogenic environment to suppress thermal background radiation from the IFU itself. By placing the IFU at the telescope focal plane as slit masks, SWIMS can be switched to an IFS mode. Thus, the IFU is also stored in a dedicated slot of the carousel and installed at the telescope focal plane by the MOSU. At the telescope focal plane, the IFU is held in place by magnets embedded in both the IFU and focal plane structure, which is the same mechanism as slit masks. This basic concept of the IFU does not require any modification of the optics of SWIMS. On the other hand, the size of the IFU is restricted to within 235×170×55  mm3 due to space limitations in the carousel.

3.

Optics of the SWIMS-IFU

The SWIMS-IFU aims to realize near-infrared IFS with an FoV larger than 100  arcsec2 and a wide simultaneous wavelength coverage of 0.9 to 2.5  μm. To realize the large FoV, we set a slice width of the IFU to be 0.4 arcsec, which is the same as a typical seeing size at good observation sites. The number of slices is 26 (CH+13 to CH+1 and CH-1 to CH-13), and the length of each slice is 13.5 arcsec, resulting in the FoV of 13.5×10.4  arcsec2, which is larger than that of existing instruments by more than a factor of two. Specifications of the IFU are summarized in Table 1. Note that these specifications are for the Subaru Telescope. We also should note that only the central 12 channels (from CH-6 to CH+6) are currently covered by two H2RG detectors available in each arm of SWIMS, which corresponds to 13.5×4.8  arcsec2. The FoV currently achieved is almost half of the full specification, but it is comparable to the FoV of 8×8  arcsec2 of VLT/ERIS, which is the largest FoV of any existing near-infrared IFS instruments. The full FoV of 13.5×10.4  arcsec2 can be achieved by installing two additional H2RG detectors on each arm.

Table 1

Specifications of the SWIMS-IFU.

Field of view13.5×10.4  arcsec2
(13.5×4.8  arcsec2)a
Slice width0.4 arcsec
Number of slices26
Slice length13.5 arcsec
Wavelength coverage0.9 to 2.5  μm
Spectral resolution λ/Δλ875 to 1500 (blue arm), 750 to 1250 (red arm)
Input and output F-ratio12.2
Weight900 g

aFoV covered by central 12 channels.

Figure 2 shows an optical layout of the IFU. To realize the compact optical design, we adopt the Advanced Image Slicer configuration,24 in which pupil images are formed on pupil mirrors to minimize the size of the pupil mirrors and to reduce the size of the whole optical system.

Fig. 2

Optical layout of the SWIMS-IFU.

JATIS_10_1_015004_f002.png

A prior optics consisting of PO0 (flat mirror), PO1 (diverging doublet lens), and PO2 (concave spherical mirror) magnifies an object image by a factor of 2.75 to match the slice width to the seeing size and forms it on a slice-mirror array. An exit pupil of the prior optics is located behind the slice-mirror array. The PO1 consists of a biconcave lens made of BaF2 and a meniscus lens made of S-TIH14 (OHARA INC.), both with an effective diameter of 9.4 mm.

The slice-mirror array, consisting of 26 flat slice mirrors, divides the magnified object image into 26 rectangular slices. Each slice mirror has different orientations and reflects incoming light to a corresponding pupil mirror to form a pupil image on it. The 26 slice mirrors are labeled as CH+13 to CH+1 and CH-1 to CH-13 from top to bottom in Fig. 2. The width of the slice mirrors is 0.52 mm, corresponding to 0.4 arcsec, and the length is 18.0 mm, corresponding to 13.5 arcsec.

A pupil-mirror array forms pseudo-slit images, which are de-magnified and sliced object images, on a slit-mirror array. The magnification of the pupil mirrors is the inverse of that of the prior optics, so the overall magnification of the IFU remains unity. The pupil-mirror array consists of 12 concave spherical mirrors for central channels (CH-6 to CH+6) and 14 concave off-axis ellipsoidal mirrors for outer channels (CH-13 to CH-7 and CH+7 to CH+13). The off-axis ellipsoidal mirrors reduce aberrations caused by large reflection angles for the outer channels.

The slit-mirror array reflects light into the SWIMS optics, and pupil images are formed at a cold stop of SWIMS. The slit-mirror array consists of concave spherical mirrors. The curvature and angle of each slit mirror are adjusted to form pupil images at the correct position of the cold stop.

Figure 3 shows root mean square (RMS) diameters of ray-traced spots formed by the prior optics at the slice-mirror array. All of the spot sizes are significantly smaller than the slice width of 0.52 mm. RMS diameters of ray-traced spots formed by the IFU at the telescope focal plane are shown in Fig. 4. Improvement by the off-axis ellipsoidal pupil mirrors in the outer channels is clearly seen. The largest RMS spot diameter is 125  μm at the edge of CH+6, which corresponds to 0.26 arcsec at the telescope focal plane.

Fig. 3

RMS spot diameters at the slice-mirror array for the center and both edges of each channel.

JATIS_10_1_015004_f003.png

Fig. 4

RMS spot diameters at the telescope focal plane at the center and both edges of each pseudo-slit. Dashed vertical lines indicate the boundaries between channels with a spherical pupil mirror and with an off-axis ellipsoidal pupil mirror. Outer channels adopt the off-axis ellipsoidal pupil mirrors.

JATIS_10_1_015004_f004.png

4.

Manufacturing

4.1.

Mirror Fabrication by Ultra-Precision Cutting

Figure 5 shows a mechanical structure of the SWIMS-IFU. It consists of five optical elements (PO0+slice-mirror array, PO1, PO2, pupil-mirror array, and slit-mirror array) and a base plate. All optical surfaces except for the PO1 and interface surfaces including those of the base plate are fabricated with ultra-precision cutting. Our goal is to achieve sufficient surface quality of the mirrors by only cutting and to guarantee the relative positions and orientations between mirrors within a single optical element as well as between the elements, which will greatly save time and effort for alignment procedures.

Fig. 5

Mechanical structure of the SWIMS-IFU. Vignetting points of pupil images and the FoV described in Sec. 5.1 are also shown.

JATIS_10_1_015004_f005.png

The ultra-precision cutting of the surfaces was done with ULG-100D(5A) (Shibaura Machine Co., Ltd.), operated by RIKEN, using various machining methods, as summarized in Table 2. All mirrors except for the PO2 are made of a rapidly solidified aluminum alloy RSA606125 to achieve a good surface roughness of less than 10 nm. The completed optical elements are shown in Fig. 6. Unique points in our fabrication are (1) the PO0+slice-mirror array, which has two types of flat mirrors in a single work piece, and (2) the pupil-mirror array with spherical and ellipsoidal mirrors and the slit-mirror array with various spherical mirrors. In the former fabrication, to complete two types of flat mirrors in a single operation, two types of tools were mounted simultaneously on the ultra-precision machine. One of the tools is a custom-made flat diamond tool for the slice mirrors, which has the same width and cross-sectional shape as the slice mirrors.26 The latter fabrications adopted a ball-end mill tool with a small cutting-edge radius to fabricate the spherical and ellipsoidal mirrors in a single operation. The method can process various spherical and ellipsoidal mirrors with a radius of curvature larger than that of the tool.

Table 2

Summary of fabrication by ultra-precision cutting.

PO2PO0+slice-mirror arrayPupil-mirror arraySlit-mirror array
Number of mirrors1PO0: 1 slice mirrors: 262626
Mirror shapeSphericalFlatSpherical and ellipsoidalSpherical
Mirror size (mm2)52×41PO0: 14×16 slice mirrors: 18×0.527.25×77.25×6.8
MaterialA6061RSA6061RSA6061RSA6061
Fabrication methodTurningShaper cuttingMillingMilling
ToolRound-tip tool (R=0.5  mm)PO0: round-tip tool (R=2.0  mm), slice mirrors: custom-made flat tool (w=0.52  mm)Ball end mill (R=1.0  mm)Ball end mill (R=0.5  mm)
Finishing pitch (μm)2.8PO0: 16.7, slice mirrors: single cut1010

Fig. 6

Completed mirrors for the SWIMS-IFU: (a) PO2, (b) PO0+slice-mirror array, (c) pupil-mirror array, and (d) slit-mirror array.

JATIS_10_1_015004_f006.png

4.2.

Mirror Surface Quality

4.2.1.

Surface roughness

Surface roughness is an important indicator for assessing the scattering loss at a mirror surface. The scattering loss Lscatter by the five mirrors of the IFU at a wavelength λ is estimated by the following equation:8

Eq. (1)

Lscatter=1i=15(1(4πσiλ)2),
where σi is the RMS surface roughness of each mirror. A requirement for the total scattering loss is less than 10% at 0.9  μm, which is translated to a mean RMS surface roughness of less than 10 nm per surface when surface roughness tolerances are equally assigned to all mirror surfaces.

The surface roughness of each mirror was measured using a white light interferometer (Zygo NewView 7200). Figure 7 shows examples of the measured surface images. For the PO0 and the PO2, five and six points in each mirror surface were measured, respectively, with a measurement area of 140×105  μm2. For the three mirror arrays, we used a different objective lens with a long working distance not to interfere with their complex structures, which results in lower spatial sampling but a larger measurement area of 702×526  μm2. Three points in each mirror were measured for the slice-mirror array and the slit-mirror array, whereas five points were measured for the pupil-mirror array. Although all mirrors were measured for the pupil-mirror array and the slit-mirror array, only 10 mirrors from CH-1 to CH+9 were measured for the slice-mirror array due to the interference between the work piece and the objective lens. For the pupil mirrors and the slit mirrors, a low-pass filter was applied to remove patterns with a spatial period larger than 20  μm (twice the finishing pitch), which were considered to be shape errors rather than surface roughness.

Fig. 7

Measured surface images of (a) PO0, (b) PO2, (c) slice mirror of CH-1, (d) pupil mirror of CH-1, and (e) slit mirror of CH-1.

JATIS_10_1_015004_f007.png

As shown in Table 3, the mean RMS surface roughness, except for the slice mirrors, is 5 to 8 nm, which meets the requirement of less than 10 nm. The slice mirrors have the larger mean RMS surface roughness of 12.74 nm, which is due to many impurities and unsmooth patterns of the base material of RSA6061. The resulting RMS surface roughness is comparable to previous studies using a fly-cut method.8,27,28 Combining the mean surface roughness of each mirror in Table 3, the total scattering loss of five mirrors is estimated to be 6.4%, which still meets the requirement of a scattering loss of less than 10%.

Table 3

Measured RMS surface roughness of the mirrors.

PO0PO2Slice mirrorsPupil mirrorsSlit mirrors
Mean (nm)8.084.9012.745.507.36
Min (nm)6.624.228.093.834.42
Max (nm)9.435.4918.849.5312.80

4.2.2.

Shape error

The shape error is an important parameter related to image quality. In particular, the shape error of the pupil mirrors largely affects the image quality because pupil images are formed on the pupil mirrors. Therefore, we simulated degradation of the image quality caused by the shape error of the pupil mirrors using Zemax. In this simulation, we required the degradation of the RMS spot radius at the position of the slit mirror to be less than 10%, and we found that the peak-to-valley (P-V) shape error of each mirror should be less than 300 nm.

The shape errors of the fabricated mirrors were measured with a laser interferometer (Zygo Verifire QPZ). Figure 8 shows examples of the measured shape error profiles. The measurement of the slice mirrors was performed only for six channels, CH±1, ±6, and ±13, due to time constraints. Also, only the spherical surfaces of the pupil mirrors were measured because an aspherical surface is difficult to be measured with an ordinary laser interferometer. Note that P-V values of the pupil mirrors were measured within an effective area of 4.5 mm in diameter.

Fig. 8

Measured shape error profile of (a) PO0, (b) PO2, (c) slice mirror of CH-1, (d) pupil mirror of CH-1, and (e) slit mirror of CH-1.

JATIS_10_1_015004_f008.png

All mirrors except the PO2 meet the requirement for the P-V shape error of less than 300 nm, as summarized in Table 4. The large shape error of the PO2 comes from its thickness of 3  mm due to the limited total size of the IFU; we fixed the work piece to a jig during the process, and slight deformation of the shape may have occurred when it was removed from the jig after finishing. We performed ray-tracing analysis to evaluate the degradation of the image quality caused by the shape error of the PO2, in which the measured shape error of the PO2 was simulated as a vertical astigmatism, and we confirmed that only a few % degradation of the RMS spot radius at the final image plane of the IFU was caused, which is acceptable.

Table 4

Measured P-V shape error of mirrors.

PO0PO2Slice mirrorsPupil mirrorsSlit mirrors
Mean (nm)240507138211169
Min (nm)86155117
Max (nm)228297277

4.3.

Assembly

The optical elements of the IFU should be assembled within the tolerances listed in Table 5. They were derived from a tolerance analysis using Zemax, in which the sensitivity of each degree of freedom to throughput, the image quality at the slice-mirror array and at the detector, and the image displacement at the detector were evaluated first. Then, a Monte Carlo simulation with 1000 trials was performed to evaluate the effect of combined errors. Within the tolerances in Table 5, 99.7% of the trials show a throughput degradation of less than 10%; a variations of the RMS spot radius at the slice-mirror array and at the detector of less than 20% and 10%, respectively; and an image misalignment at the detector of a few pixels.

Table 5

Installation tolerances for the optical elements of the IFU. Coordinates are illustrated in Fig. 2, except for the PO1, which has tilted coordinates with the Y axis aligned with the optical axis at the PO1.

Shift X (μm)Shift Y (μm)Shift Z (μm)Tilt X (arcmin)Tilt Y (arcmin)Tilt Z (arcmin)
PO0+slice-mirror array1020100.90.80.7
PO11020202.21.61.6
PO22020100.80.40.4
Pupil-mirror array202050.40.40.4
Slit-mirror array1020101.10.40.4

The ultra-precision cutting was also used in the assembly process to realize simple and precise assembly. First, mounting surfaces of the base plate were machined by ultra-precision cutting to control their height with a few μm accuracy [Fig. 9(a)]. Fixing flanges of the pupil-mirror array and the slit-mirror array were also machined in the same operation as the mirrors [Fig. 9(b)], so relative positions and orientations between the mirrors and the base plate are guaranteed. For unmachined fixing surfaces of the pupil-mirror array and the slit-mirror array and for the optical elements with fixing flanges that cannot be machined, shims with a thickness that was controlled by ultra-precision cutting were used for accurate positioning. Positioning pins were used to fix the optical elements in the plane of the machined surfaces or shims with an accuracy of 10 to 20  μm. A lens holder of the PO1 was fixed to fixing flanges next to the PO0 mirror.

Fig. 9

Ultra-precision cutting for assembly. Solid boxes show machined surfaces. (a) Mounting surfaces of the base plate. Blue boxes show mounting surfaces for the pupil-mirror array and red boxes for the slit-mirror array. Black boxes show machined surfaces for unmachined surfaces of the pupil-mirror array and the slit-mirror array shown in dashed black boxes in panel (b). (b) Fixing flanges of the pupil-mirror array (top) and the slit-mirror array (bottom).

JATIS_10_1_015004_f009.png

As a result, the assembly of the IFU was completed with only one adjustment of the PO1 lens holder. The completed IFU is shown in Fig. 10. The assembly of the SWIMS-IFU using ultra-precision cutting is simple and does not require a precise adjustment. The concept is applicable to not only IFUs but also a variety of optical units with complex designs.

Fig. 10

Completed IFU: (a) without a cover and (b) with a cover.

JATIS_10_1_015004_f010.png

5.

Laboratory Performance Tests

5.1.

Vignetting

There are two vignetting points for the SWIMS-IFU: one is vignetting of the pupil images at the edges of the FoV by an aperture of the PO1 and the other is vignetting of the FoV by a lens holder of the PO1 (refer to Fig. 5).

To measure the vignetting of the pupil images by the aperture of the PO1, we set a light source in the visible wavelength with an input F-ratio of 12.2 at the input side of the IFU to simulate incident light of the Subaru telescope, and we took an illumination pattern at the pupil-mirror array from the output side by a CMOS camera, as shown in Fig. 11(a). Figure 11(b) shows the observed illumination patterns for various channels with different positions in it. The vignetting of the pupil images at the edges of the FoV is larger than at the center, resulting in a lower throughput there. This is caused by an insufficient effective clear aperture diameter of the PO1 lens due to space limitations.

Fig. 11

(a) Configuration to measure vignetting of the pupil images by the aperture of the PO1. (b) Observed pupil images at the center and both edges of CH+13, +6, +1, 1, 6, and 13. The pupil image of the central part of CH+13 is missing due to the vignetting of the FoV.

JATIS_10_1_015004_f011.png

To measure the vignetting of the FoV by the lens holder of the PO1, we set a diffused visible light source at the output side of the IFU, shown in Fig. 12(a), and took an illumination pattern at the slice-mirror array from the input side by a CMOS camera. Figure 12(b) shows the observed illumination pattern at the slice-mirror array. It can be seen that the centers of CH+13 and CH+12 are vignetted, which is caused by obstruction of the light between the slice-mirror array and the PO2 by a lens holder of the PO1 lens.

Fig. 12

(a) Configuration to measure vignetting of the FoV by the lens holder of the PO1. (b) Observed image of the slice-mirror array. There is vignetting at the top of the image, affecting the central part of CH+13 and CH+12.

JATIS_10_1_015004_f012.png

5.2.

Image Quality

To measure the image quality, we again set the light source with F/12.2 and took the pseudo-slit images of all 26 channels by the CMOS camera from the output side of the IFU. The setup for this test was the same as Fig. 11(a), except that the CMOS camera focused on the pseudo-slit image instead of the pupil image.

To evaluate the image quality, cross sections of each pseudo-slit image were fitted with a convolution function of a top-hat function with the designed width and a Gaussian function. The analysis was done for the center and both edges of each pseudo-slit image.

Figure 13 shows double the standard deviations of the Gaussian functions, which we interpret as RMS spot diameters, for all channels, as well as the RMS spot diameters from a ray-tracing simulation at the position of the pseudo-slit image. We should note that these values include degradation caused by various elements. The ray-tracing simulation simulates a spot for a point source at infinity degraded by the telescope and all optical elements of the IFU. On the other hand, the laboratory measurements consider only the degradation by the pupil mirrors. Thus, some measurements at the edge of the outer channels show an RMS spot diameter smaller than the ray-tracing simulation. Still, the overall trend of the measurement is consistent with the ray-tracing simulation, demonstrating the effect of the off-axis ellipsoidal pupil mirrors for the outer channels.

Fig. 13

Measured image quality of the pupil mirrors at the center and both edges of the pseudo-slit image and ray-tracing simulated values in 700 nm. Black circles show the measured value, and white squares show the ray-tracing simulated values. Dashed vertical lines show the boundaries between channels with a spherical pupil mirror and with an off-axis ellipsoidal pupil mirror. Measured values for the center of CH+12 and CH+13 are missing due to vignetting of the FoV, and measured values for edge1 of CH+13 and edge2 of CH-13 are missing due to the configuration of the measurement.

JATIS_10_1_015004_f013.png

6.

Engineering Observation

The engineering observation at the Subaru telescope was carried out on the first half of the night of March 27, 2022, and on the first half of the night of December 2, 2022. Through these engineering observations, we evaluated the on-sky performance of the SWIMS-IFU and demonstrated that the development utilizing ultra-precision cutting can complete the IFU as designed. This allows for improving the performance by revising the IFU optics and mechanics and quickly starting the operation of the SWIMS-IFU at the TAO 6.5 m telescope.

Figure 14 shows observed two-dimensional spectra of a standard star, HIP51684, placed at the center of the FoV with integration times of 20 s for the blue arm and 10 s for the red arm of SWIMS. The spectra of the standard star were obtained at nine positions in the FoV: the center and both edges of CH-1, CH-4, and CH+4. Figures 15(a) and 15(b) are images of a globular cluster M3 reconstructed from the pseudo-slit images of the IFU with an integration time of 3 s for the J- and KS-band filters, respectively. Four pseudo-slit images were obtained in the engineering observation.

Fig. 14

Observed two-dimensional spectra of a standard star, HIP51684, of (a) the blue arm covering from 0.9 to 1.45  μm and (b) the red arm covering from 1.45 to 2.5  μm.

JATIS_10_1_015004_f014.png

Fig. 15

Comparison of images of M3. White arrows indicate stars used in the scale measurement. Reconstructed images of the IFU are binned every 5 pixels in the vertical direction. (a) Reconstructed J-band image of the FoV of the IFU. (b) Reconstructed KS-band image of the FoV of the IFU. (c) Pan-STARRS y-band image.

JATIS_10_1_015004_f015.png

6.1.

Spatial Scales and FoV

The pseudo-slit imaging data of the globular cluster M3 were used for spatial scale measurements. The pseudo-slit images were reconstructed into an image with the FoV of the IFU. Positions of stars in the reconstructed image were compared with those in a y-band image of Pan-STARRS29 to measure the slice width and the spatial scale along the slices (Fig. 15). The positions of the stars were measured with the IRAF imexam task, and the slice width and the spatial scales were computed by the IRAF ccmap task.

Table 6 shows the mean and standard deviation of measurements for each filter, as well as the designed values. Both the slice width and the spatial scale for each filter are almost consistent with the designed values. With the measured slice width and the spatial scale, the FoV of the SWIMS-IFU becomes 13.4×4.8  arcsec2.

Table 6

Measured and designed spatial scales of the IFU.

Slice width (arcsec)Spatial scale (arcsec pix−1)
Blue arm (J)0.403 ± 0.0090.0950 ± 0.0003
Red arm (KS)0.398 ± 0.0230.0962 ± 0.0002
Design0.3900.0960

6.2.

Wavelength Scales and Coverage

Wavelength calibration was carried out using the night sky OH emission lines30 in the images of the standard star. Pixel scales in the dispersion direction were measured to be 2.47  Åpix1 for the blue arm and 4.67  Åpix1 for the red arm. With the measured pixel scales, the observed spectra cover wavelength ranges of 0.85 to 1.48  μm for the blue arm and 1.42 to 2.56  μm for the red arm. It should be noted that a gap between the two detectors, as seen in Fig. 14, corresponds to a 320   width in the range of 9420 to 10070 Å for the blue arm and a 585   width in 15510 to 16770 Å for the red arm, affecting the observed spectra.

6.3.

Image Quality

We evaluated image quality using the size of the image of the standard star HIP51684 at the slice-mirror array and at the SWIMS detector. Due to time constraints of the engineering observations, the image quality measurements were made for some of the 12 available channels. However, the laboratory performance test in Sec. 5.2 shows the expected results for all channels, and it is expected that the other channels not evaluated in the engineering observations will show similar results.

The image quality at the slice-mirror array was evaluated by reconstructing a stellar profile at the slice-mirror array from signal counts of the spectra of the standard star as follows. For the obtained spectra, several wavelengths that were not affected by atmospheric absorption or night sky emissions were selected, and signal counts of the standard star at the wavelengths were extracted from the 5–6 channels. The extracted signal counts were arranged in the order of the slice mirrors to yield a stellar profile in the direction perpendicular to the slice mirrors. The stellar profile was fitted with a Gaussian function, and an FWHM of the Gaussian function was used to evaluate the size of the standard star image using the slice width of 0.4 arcsec. The evaluations were performed using data in which the standard star was introduced at the nine positions of the FoV of the IFU.

Figure 16 shows the FWHMs at the slice-mirror array as a function of wavelength. The resulting FWHM was 0.49 to 0.60 arcsec, which is consistent with an expected value of 0.50 to 0.53 arcsec calculated as a root sum square of the image quality from a ray-tracing simulation of 0.205 arcsec and the measured seeing sizes of 0.49 arcsec in the Y-band and 0.45 arcsec in the H-band.

Fig. 16

FWHM of the standard star at the slice mirror as a function of wavelength at the nine positions in the IFU FoV.

JATIS_10_1_015004_f016.png

The image quality at the detectors was evaluated by fitting the spatial extent of the standard star spectra observed in CH±1 and CH±4 along the slice length direction with a Gaussian function at wavelengths of 1.0  μm (the central wavelength of the Y-band), 1.2  μm (J-band), 1.6  μm (H-band), and 2.1 m (KS-band). The Gaussian fitting was done by the IRAF imexam task, and the pixel scales in Table 6 were used. Table 7 shows the measured FWHMs. They range 0.46 to 0.65 arcsec depending on the wavelength, which agrees with an expected value of 0.53 to 0.56 arcsec calculated as a root sum square of the image quality from a ray-tracing simulation of 0.268 arcsec and the measured seeing sizes of 0.49 arcsec in the Y-band and 0.45 arcsec in the H-band. There is no significant difference between the channels. It should be noted that a gap in the FWHM between the blue (Y- and J-band) and the red (H- and KS-band) arms is caused by SWIMS optics and is not due to the IFU.

Table 7

FWHM of the standard star in the spatial direction at the detectors.

CHFWHM Y (arcsec)FWHM J (arcsec)FWHM H (arcsec)FWHM KS (arcsec)
−10.6210.6100.4730.473
+10.6160.6050.4980.481
−40.6450.6300.4600.457
+40.6150.5980.4830.479

6.4.

Throughput

The throughput of the SWIMS-IFU was evaluated as a ratio of counts of dome flat data with the IFU to that without the IFU. The dome flat data were obtained in the Y-, J-, H-, and KS-band for the imaging mode and in the spectroscopy mode. For the spectroscopy mode, dome flat data without the IFU were obtained using a slit mask with slits in the same positions and widths as those of the IFU pseudo-slit images.

Figure 17 shows the results, which range from 50% to 75% depending on the wavelength and channel. The results are almost consistent between the imaging and the spectroscopy modes. In Fig. 17, we also show throughput models including scattering losses by the measured surface roughness of manufactured mirrors using Eq. (1), reflectivity of uncoated aluminum mirrors,31 internal transmittance and reflectivity of anti-reflection coating of the PO1, and losses by vignetting for CH±1 and CH±6. The throughput models for the other channels should be between the models of CH±1 and CH±6 because the loss by vignetting is between those of CH±1 and CH±6. We can see that the measured values are consistent with the throughput models. The throughput is dominated by the reflectivity of uncoated aluminum mirrors in shorter wavelengths and the reflectivity of the anti-reflection coating of the PO1 in longer wavelengths. The vignetting described in Sec. 5.1 also affects the throughput over the entire wavelength range.

Fig. 17

Throughput of the SWIMS-IFU. Open circles show the average throughput of twelve channels by Y-, J-, H-, and KS-band imaging. Error bars show the wavelength range of each filter and the range from minimum to maximum values for the 12 channels. Colored lines show the throughput from the spectroscopy data. Black lines show the throughput models calculated from the scattering loss by the measured mirror surface roughness, aluminum mirror reflectivity, internal transmittance and reflectivity of anti-reflection coating of the PO1 lens, and losses by vignetting for CH±1 and CH±6.

JATIS_10_1_015004_f017.png

6.5.

Stray Light

During the first engineering observation, we observed stray light affecting wavelengths longer than 2.1  μm [Fig. 18(a)]. Ray-tracing analysis revealed that the stray light was caused by thermal emission from a baffle around the Cassegrain hole of the Subaru telescope, which enters the IFU at different incident angles and paths from the designed light, is reflected by a slice mirror, and then enters a pupil mirror in an adjacent channel. Based on the results of the ray-tracing analysis, we narrowed the entrance aperture of the IFU in the slice length direction from 14.5 to 8.5 mm to suppress the stray light, as shown in Fig. 18(b).

Fig. 18

Stray light observed in the pseudo-slit image of CH-1 in the KS-band, (a) before modification of the entrance aperture of the SWIMS-IFU and (b) after modification. White arrows indicate the position of the stray light.

JATIS_10_1_015004_f018.png

Although we could reduce the thermal stray light, it persists in the current optical design of the SWIMS-IFU, and modification in the optical design is necessary to eliminate the stray light in the future upgrade for the TAO 6.5 m telescope. One solution is to narrow the FoV and thus the shape of the entrance aperture of the IFU to prevent the stray light from entering the IFU. Another is to change the arrangement of the pupil mirrors, especially in the vertical direction, to prevent the stray light from entering the adjacent channel, such as SPIFFI of VLT/SINFONI.32

7.

Conclusions

We have developed an image-slicer type IFU with a larger FoV for a near-infrared spectrograph, SWIMS. The FoV of 13.5×10.4  arcsec2, which is larger than existing near-infrared IFS instruments, was realized by matching the width of the slice mirrors to a seeing size at the Subaru telescope of 0.4 arcsec. We also realized a compact optical design to fit within a volume of 235×170×55  mm3 using off-axis ellipsoidal pupil mirrors for the outer channels that reduced aberrations.

Its complicated optical elements were fabricated by ultra-precision cutting; two types of flat mirrors in the PO0+slice-mirror array were simultaneously fabricated using dual-mounted tools, and free-form surfaces in the pupil-mirror array and the slit-mirror array were fabricated by ball end milling with a small-radius tool. The finished mirrors meet our requirements of RMS surface roughness of less than 10 nm and P-V shape error of less than 300 nm. We also used the ultra-precision cutting to machine mounting surfaces and custom-made shims for easy assembly. As a result, the assembly was completed with only one adjustment of the PO1 lens. The concept of the easy assembly can be applied to the development of other optical systems. Remaining technical issues are how to measure the shape of aspherical mirrors and how to assess their relative positions.

The main results of the performance evaluation of the completed IFU are as follows:

  • We confirmed the improvement of image quality by the off-axis ellipsoidal pupil mirrors through the laboratory test. The engineering observation at the Subaru telescope showed the FWHMs of 0.49 to 0.60 arcsec at the slice-mirror array and 0.46 to 0.65 arcsec at the detectors, which are consistent with their expected values.

  • The throughput of the IFU was 50% to 75% as designed. It is dominated by the reflectivity of uncoated aluminum mirrors in shorter wavelengths and the reflectivity of the anti-reflection coating of the PO1 in longer wavelengths.

  • Thermal emission from the baffle of the telescope was observed as stray light. It was confirmed to be suppressed by narrowing the entrance aperture of the IFU; however, modification of arrangement of the pupil mirrors is necessary to remove it completely.

The SWIMS-IFU will be upgraded for the observation at the TAO 6.5 m telescope. To improve the vignetting, we will remove the PO1 lens from the optical design and realize the required magnification by adopting a pick-off mirror with power. Additional aberration caused by the new pick-off mirror will be corrected using aspherical mirrors. This update will also improve the low throughput in the longer wavelengths. The low throughput in the shorter wavelengths can be improved by coating the mirrors with gold. To remove the thermal stray light, we have to create a more complicated slice-mirror array to modify the arrangement of the pupil images. The fabrication of the new slice-mirror array will be a new technical challenge.

Code and Data Availability

The data that support the findings of this article are not publicly available. They can be requested by contacting the corresponding author.

Acknowledgments

We appreciate the support of the Subaru telescope for the engineering observations. The development activities are supported by the Advanced Technology Center, NAOJ, and the Advanced Manufacturing Support Team, RIKEN. This work made use of iraf33,34 and pyraf.35 This work made use of Astropy,36 a community-developed core Python package, and an ecosystem of tools and resources for astronomy.3739 The Pan-STARRS1 Surveys (PS1) and the PS1 public science archive were made possible through contributions by the Institute for Astronomy, the University of Hawaii, the Pan-STARRS Project Office, the Max-Planck Society and its participating institutes, the Max Planck Institute for Astronomy, Heidelberg and the Max Planck Institute for Extraterrestrial Physics, Garching, Johns Hopkins University, Durham University, the University of Edinburgh, the Queen’s University Belfast, the Harvard-Smithsonian Center for Astrophysics, the Las Cumbres Observatory Global Telescope Network Incorporated, the National Central University of Taiwan, the Space Telescope Science Institute, the National Aeronautics and Space Administration (Grant No. NNX08AR22G) issued through the Planetary Science Division of the NASA Science Mission Directorate, the National Science Foundation (Grant No. AST-1238877), the University of Maryland, Eotvos Lorand University (ELTE), the Los Alamos National Laboratory, and the Gordon and Betty Moore Foundation. This work was supported by JSPS KAKENHI (Grant Nos. JP23540261, JP15H02062, JP20H00171, JP14J06780, and JP20J21493). This work was supported by the Research Coordination Committee, National Astronomical Observatory of Japan (NAOJ). KK was supported as a research fellow of the Japan Society for the Promotion of Science. This work is based on the following works: “Development status of a near-infrared integral field unit SWIMS-IFU,” Proceedings of the SPIE, 12188, 121882V (2022);40 “Diamond machining of two-in-one optical element including slice mirror array for near-infrared integral-field spectrograph,” Proceedings of the SPIE, 12188, 121882X (2022);26 and the Ph.D. thesis of KK of the University of Tokyo (2023).

References

1. 

R. Bacon et al., “The MUSE second-generation VLT instrument,” Proc. SPIE, 7735 773508 https://doi.org/10.1117/12.856027 PSISDG 0277-786X (2010). Google Scholar

2. 

P. Morrissey et al., “The keck cosmic web imager integral field spectrograph,” ApJ, 864 93 https://doi.org/10.3847/1538-4357/aad597 (2018). Google Scholar

3. 

J. Allington-Smith et al., “Integral field spectroscopy with the Gemini multiobject spectrograph. I. Design, construction, and testing,” Publ. Astron. Soc. Pac., 114 892 –912 https://doi.org/10.1086/341712 (2002). Google Scholar

4. 

S. Ozaki et al., “Integral field unit for the existing imaging and spectroscopy instrument, FOCAS,” Publ. Astron. Soc. Jpn., 72 97 https://doi.org/10.1093/pasj/psaa092 (2020). Google Scholar

5. 

F. Eisenhauer et al., “SINFONI - Integral field spectroscopy at 50 milli-arcsecond resolution with the ESO VLT,” Proc. SPIE, 4841 1548 –1561 https://doi.org/10.1117/12.459468 PSISDG 0277-786X (2003). Google Scholar

6. 

R. Sharples et al., “First light for the KMOS multi-object integral-field spectrometer,” Messenger, 151 21 –23 (2013). Google Scholar

7. 

J. Larkin et al., “OSIRIS: a diffraction limited integral field spectrograph for Keck,” Proc. SPIE, 6269 62691A https://doi.org/10.1117/12.672061 PSISDG 0277-786X (2006). Google Scholar

8. 

J. R. Allington-Smith et al., “New techniques for integral field spectroscopy - I. Design, construction and testing of the GNIRS IFU,” Mon. Not. R. Astron. Soc., 371 380 –394 https://doi.org/10.1111/j.1365-2966.2006.10666.x (2006). Google Scholar

9. 

P. J. McGregor et al., “Gemini near-infrared integral field spectrograph (NIFS),” Proc. SPIE, 4841 1581 –1591 https://doi.org/10.1117/12.459448 PSISDG 0277-786X (2003). Google Scholar

10. 

M. F. Closs et al., “The integral field unit on the James Webb Space Telescope’s near-infrared spectrograph,” Proc. SPIE, 7010 701011 https://doi.org/10.1117/12.788820 PSISDG 0277-786X (2008). Google Scholar

11. 

M. Konishi et al., “Design and status of a near-infrared multi-object spectrograph for the TAO 6.5-m Telescope,” Proc. SPIE, 7735 77356T https://doi.org/10.1117/12.856230 PSISDG 0277-786X (2010). Google Scholar

12. 

M. Konishi et al., “Design and development of SWIMS: a near-infrared multi-object spectrograph for the University of Tokyo Atacama Observatory,” Proc. SPIE, 8446 84467P https://doi.org/10.1117/12.924259 PSISDG 0277-786X (2012). Google Scholar

13. 

K. Motohara et al., “Development of a simultaneous two-color near-infrared multi-object spectrograph SWIMS for the TAO 6.5-m telescope,” Proc. SPIE, 9147 91476K https://doi.org/10.1117/12.2054861 PSISDG 0277-786X (2014). Google Scholar

14. 

K. Motohara et al., “NIR camera and spectrograph SWIMS for TAO 6.5m telescope: overview and development status,” Proc. SPIE, 9908 99083U https://doi.org/10.1117/12.2231386 PSISDG 0277-786X (2016). Google Scholar

15. 

M. Konishi et al., “Development status of the simultaneous two-color near-infrared multi-object spectrograph SWIMS for the TAO 6.5 m telescope,” Proc. SPIE, 10702 1070226 https://doi.org/10.1117/12.2310060 PSISDG 0277-786X (2018). Google Scholar

16. 

M. Konishi et al., “The University of Tokyo Atacama Observatory 6.5 m telescope: On-sky performance of the near-infrared instrument SWIMS on the Subaru telescope,” Proc. SPIE, 11447 114475H https://doi.org/10.1117/12.2560422 PSISDG 0277-786X (2020). Google Scholar

17. 

Y. Yoshii et al., “The University of Tokyo Atacama Observatory 6.5 m telescope project,” Proc. SPIE, 7733 773308 https://doi.org/10.1117/12.856680 PSISDG 0277-786X (2010). Google Scholar

18. 

Y. Yoshii et al., “Overview of University of Tokyo Atacama Observatory 6.5 m telescope project,” Proc. SPIE, 9145 914507 https://doi.org/10.1117/12.2054371 PSISDG 0277-786X (2014). Google Scholar

19. 

Y. Yoshii et al., “The University of Tokyo Atacama Observatory 6.5 m telescope: project overview and current status,” Proc. SPIE, 9906 99060R https://doi.org/10.1117/12.2231391 PSISDG 0277-786X (2016). Google Scholar

20. 

M. Doi et al., “The University of Tokyo Atacama Observatory 6.5 m telescope: project overview and current status,” Proc. SPIE, 10700 107000W https://doi.org/10.1117/12.2313099 PSISDG 0277-786X (2018). Google Scholar

21. 

Y. Yoshii et al., “The University of Tokyo Atacama Observatory 6.5 m Telescope: overview and construction status,” Proc. SPIE, 11445 1144514 https://doi.org/10.1117/12.2560238 PSISDG 0277-786X (2020). Google Scholar

22. 

T. Miyata et al., “The University of Tokyo Atacama Observatory 6.5 m telescope: project status 2022,” Proc. SPIE, 12182 1218211 https://doi.org/10.1117/12.2628146 PSISDG 0277-786X (2022). Google Scholar

23. 

H. Takahashi et al., “Development of multi-object spectroscopy unit for simultaneous-color wide-field infrared multi-object spectrograph,” Proc. SPIE, 9147 91476N https://doi.org/10.1117/12.2055409 PSISDG 0277-786X (2014). Google Scholar

24. 

R. Content, “New design for integral field spectroscopy with 8-m telescopes,” Proc. SPIE, 2871 1295 –1305 https://doi.org/10.1117/12.269020 PSISDG 0277-786X (1997). Google Scholar

25. 

G. P. H. Gubbels et al., “Rapidly solidified aluminium for optical applications,” Proc. SPIE, 7018 70183A https://doi.org/10.1117/12.788766 PSISDG 0277-786X (2008). Google Scholar

26. 

M. Takeda et al., “Diamond machining of two-in-one optical element including slice mirror array for near-infrared integral-field spectrograph,” Proc. SPIE, 12188 121882X https://doi.org/10.1117/12.2625703 PSISDG 0277-786X (2022). Google Scholar

27. 

D. J. Purll et al., “Flight model performance of the integral field unit for the James Webb Space Telescope’s near-infrared spectrograph,” Proc. SPIE, 7739 773917 https://doi.org/10.1117/12.857737 PSISDG 0277-786X (2010). Google Scholar

28. 

C. M. Dubbeldam et al., “The KMOS Integral Field System: fabrication, alignment, and test of 1000+ optical surfaces,” Proc. SPIE, 8450 84501M https://doi.org/10.1117/12.925134 PSISDG 0277-786X (2012). Google Scholar

29. 

K. C. Chambers et al., “The pan-STARRS1 surveys,” (2016). Google Scholar

30. 

P. Rousselot et al., “Night-sky spectral atlas of OH emission lines in the near-infrared,” Astron. Astrophys., 354 1134 –1150 AAEJAF 0004-6361 (2000). Google Scholar

31. 

National Astronomical Observatory of Japan, Chronological Scientific Tables 2022, Maruzen( (2021). Google Scholar

32. 

F. Eisenhauer et al., “Imaging the universe in 3D with the VLT: the next-generation field spectrometer SPIFFI,” Proc. SPIE, 4008 289 –297 https://doi.org/10.1117/12.395483 PSISDG 0277-786X (2000). Google Scholar

33. 

D. Tody, “The IRAF data reduction and analysis system,” Proc. SPIE, 0627 733 https://doi.org/10.1117/12.968154 PSISDG 0277-786X (1986). Google Scholar

34. 

D. Tody, “IRAF in the nineties,” Astronomical Data Analysis Software and Systems II, 52 173 (1993). Google Scholar

35. 

Science Software Branch at STScI, “PyRAF: Python alternative for IRAF,” (2012). Google Scholar

37. 

Astropy Collaboration, “Astropy: a community Python package for astronomy,” Astron. Astrophys., 558 A33 https://doi.org/10.1051/0004-6361/201322068 AAEJAF 0004-6361 (2013). Google Scholar

38. 

Astropy Collaboration et al., “The astropy project: building an open-science project and status of the v2.0 core package,” Astron. J., 156 123 https://doi.org/10.3847/1538-3881/aabc4f (2018). Google Scholar

39. 

Astropy Collaboration et al., “The Astropy project: sustaining and growing a community-oriented open-source project and the latest major release (v5.0) of the core package,” Astrophys. J., 935 167 https://doi.org/10.3847/1538-4357/ac7c74 (2022). Google Scholar

40. 

K. Kushibiki et al., “Development status of a near-infrared integral field unit SWIMS-IFU,” Proc. SPIE, 12188 121882V https://doi.org/10.1117/12.2627205 PSISDG 0277-786X (2022). Google Scholar

Biography

Kosuke Kushibiki is a project researcher at the Institute of Astronomy, the University of Tokyo. He received his PhD in astronomy from the University of Tokyo in 2023. His current research interests include nearby starburst galaxies, galaxy mergers, and quenching. He has been involved in the development of an integral field unit. He has also been a member of the TAO project since 2023.

Biographies of the other authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Kosuke Kushibiki, Shinobu Ozaki, Masahiro Takeda, Takuya Hosobata, Yutaka Yamagata, Shin-ya Morita, Toshihiro Tsuzuki, Keiichi Nakagawa, Takao Saiki, Yutaka Ohtake, Kenji Mitsui, Hirofumi Okita, Yutaro Kitagawa, Yukihiro Kono, Kentaro Motohara, Hidenori Takahashi, Masahiro Konishi, Natsuko M. Kato, Shuhei Koyama, and Nuo Chen "Development of a near-infrared wide-field integral field unit by ultra-precision diamond cutting," Journal of Astronomical Telescopes, Instruments, and Systems 10(1), 015004 (4 March 2024). https://doi.org/10.1117/1.JATIS.10.1.015004
Received: 2 October 2023; Accepted: 8 February 2024; Published: 4 March 2024
Advertisement
Advertisement
KEYWORDS
Mirrors

Image quality

Telescopes

Mirror surfaces

Vignetting

Surface roughness

Engineering

RELATED CONTENT

Webb Telescope imaging performance
Proceedings of SPIE (August 27 2022)
Imaka: a ground-layer adaptive optics system on Maunakea
Proceedings of SPIE (August 05 2016)
ASSIST development of a test infrastructure for the VLT...
Proceedings of SPIE (September 18 2007)
Adaptive system of the AST-25
Proceedings of SPIE (July 07 2000)
Low turbulence--high performance
Proceedings of SPIE (July 01 1990)
All-spherical wide-field camera
Proceedings of SPIE (November 19 2021)

Back to Top